Skip to content

BD Pyxis: MedStation ES, REF: 323 recallZ-1223-2024 · CareFusion 303, Inc. · initiated Feb 1, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.

FDA root cause category: Software design

What the company is doing

On 2/1/24, correction notices were mailed or emailed to customers and distributors who were told the following: Devices with a random application hang, freeze and/or crashing can be resolved with a reboot of the device. In downtime scenarios, medications can be obtained from other devices or from pharmacy. There are also keys that can be used to manually access contents within the device. 1) Ensure you continue to manually reboot your devices as a part of your maintenance schedule. 2. Disseminate this notification within your facility network to ensure they are aware of the issue. 3. If you experience this issue, please contact Technical Support. 4. Complete the Customer Response Form and return to [email protected] If you require further assistance, please contact the firm: Customer Support: 1-866-583-8783, 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday. Technical Support: bd.com/self-service, 1-800-727-6102, 24 hours, 7 days a week

Recall details

Recall number
Z-1223-2024
Status
Open, Classified
Initiated by the company
Feb 1, 2024
Posted by the FDA
Aug 20, 2024
Recalling firm
CareFusion 303, Inc.
Quantity
20,443
Product code
BRY
Distribution
Worldwide distribution - US Nationwide distribution including in the states of VA, SC, IL, AL, KS, LA, GA, FL, CA, PR, PA, WV, WA, HI, NC, MD, AR, NY, ND, TX, NM, GU, MS, AZ, DC, AK, NE, DE, CO, SD, MO, NJ, OK, NV, OH, KY, WY, MT, UT, ID, MA, MI, MP, NH, IA, TN, CT, RI, IN, WI, OR, MN, VT, ME and the countries of BH, DE, GB, IT, JP, ES, GU, KR, BE, AR, BR, BS, CL, CO, MX, AE, AU, CA, EG, FR, HK, MY, NZ, QA, SA, TH, NL, MC, CH, FI, JO, KW, LB, OM.

Product description

BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;

Code information (models, lots, serial numbers)

REF/UDI-DI: 323/10885403512667, 327/10885403477836, 352/10885403512674. All Using Software Version ES System Release 1.4.x, 1.5.x and 1.6.x

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates