Skip to content

D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid… recallZ-1279-2024 · Peter Lazic Gmbh · initiated Oct 6, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Peter Lazic Gmbh or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip.

FDA root cause category: Device Design

What the company is doing

Reportable Event - Removal letters dated October 6, 2023 were sent to customers. Customers were instructed to return all product subject to recall and to notify any of their customers as well. Please return items for the attention of one of the PRRCs provided to ensure quick processing and assignment to the case. If you have any questions, please contact Mr. Sven Lazic ([email protected] +49 (0) 7461 96643 - 20) or Ms. Kim Maier ([email protected] or +49 (0) 7461 96643 - 21).

Recall details

Recall number
Z-1279-2024
Status
Open, Classified
Initiated by the company
Oct 6, 2023
Posted by the FDA
Mar 11, 2024
Recalling firm
Peter Lazic Gmbh
Quantity
10 units
Product code
HCI
FDA submissions
K180757
Distribution
US Nationwide distribution in the state of ME.

Product description

D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid, double action.

Code information (models, lots, serial numbers)

Model number: 35.412; UDI/DI: 04250603739616; All Lot numbers.