ADVIA Centaur Erythropoietin (EPO) Assay (100 Test) recallZ-1281-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Jan 24, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Correction notification letter dated 01/24/24 was sent to customers. Actions to be Taken by the Customer: Discontinue use of and discard the EPO kit lots listed in Table 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued EPO patient testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-1281-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 24, 2024
- Posted by the FDA
- Mar 11, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 264 units (6 US, 258 OUS)
- Product code
- GGT
- FDA submissions
- K183088
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.
Product description
ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
Code information (models, lots, serial numbers)
Siemens Material Number: 10995096; UDI/DI: 00630414010380; Kit Lots: 53984040, 55443042.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA Centaur EPO 10995096Siemens Healthcare Diagnostics Inc. · ASSAY, ERYTHROPOIETIN1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K183088 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica IM EPO 10733006Siemens Healthcare Diagnostics Inc. · ASSAY, ERYTHROPOIETIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93939).
- Z-1280-2024 · Open, ClassifiedAtellica IM Erythropoietin (EPO) Assay (100 Test)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.