Why it was recalled

Wrong product coloring. The paddle clips should be blank, but they are colored blue.

FDA root cause category: Process control

What the company is doing

Scripted Phone calls were made to customers on July 14, 2023. The customers were provided information on how to identify affected product and to return all product subject to recall and to notify any of their customers as well. Instructions for return were provided. Updated 3/11/24: Customer notification letters sent out on March 5, 2024 to summarize the information from the phone calls. Customers were instructed to inspect their inventory and return all stock products to one of the PRRCs ([email protected]/+49 (0) 7461 96643-20 or [email protected]/+49 (0) 7461 96643-21).

Recall details

Recall number
Z-1288-2024
Status
Open, Classified
Initiated by the company
Jul 14, 2023
Posted by the FDA
Mar 12, 2024
Recalling firm
Peter Lazic Gmbh
Quantity
51 units
Product code
HCH
FDA submissions
K081489
Distribution
US Nationwide distribution in the state of ME.

Product description

L-Aneurysm Clip TITANIUM Paddle-Clip, permanent, 11 mm straight, Non-Sterile

Code information (models, lots, serial numbers)

Model Number: 45.782; UDI/DI: 04250603733799; Serial Numbers: LX61385, LW36020, LW36021, LW36022, LW36023, LW36024, LW36025, LW36026, LW36027;