Skip to content

CEREBASE recallZ-1284-2024 · Medos International Sarl · initiated Feb 2, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medos International Sarl or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.

FDA root cause category: Under Investigation by firm

What the company is doing

Johnson&Johnson MedTech/DePuy Synthes issued an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 02/02/2024 via letter delivered by FedEx. The notice explained the issue and the risk and requested the affected inventory be quarantined and the notice be disseminated to all appropriate personnel. Further instructions state: "Complete the Business Reply Form (BRF) confirming receipt of this notice and scan and email signed form to [email protected] or fax it to (305) 265-6889, Attention Line: Cerebase FA2350411. Please return the BRF even if you do not have product subject to this recall. Follow instructions in the letter and immediately call 1-844-483-3882 to return any inventory of CEREBASE DA Guide Sheath devices subject to this recall. To receive credit reimbursement, customers must return product subject to this recall. Attention Line: Cerebase FA2350411"

Recall details

Recall number
Z-1284-2024
Status
Open, Classified
Initiated by the company
Feb 2, 2024
Posted by the FDA
Mar 13, 2024
Recalling firm
Medos International Sarl
Quantity
1343 units
Product code
QJP
FDA submissions
K192804
Distribution
Worldwide distribution.

Product description

CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

a) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.