CEREBASE recallZ-1284-2024 · Medos International Sarl · initiated Feb 2, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medos International Sarl or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.
FDA root cause category: Under Investigation by firm
What the company is doing
Johnson&Johnson MedTech/DePuy Synthes issued an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 02/02/2024 via letter delivered by FedEx. The notice explained the issue and the risk and requested the affected inventory be quarantined and the notice be disseminated to all appropriate personnel. Further instructions state: "Complete the Business Reply Form (BRF) confirming receipt of this notice and scan and email signed form to [email protected] or fax it to (305) 265-6889, Attention Line: Cerebase FA2350411. Please return the BRF even if you do not have product subject to this recall. Follow instructions in the letter and immediately call 1-844-483-3882 to return any inventory of CEREBASE DA Guide Sheath devices subject to this recall. To receive credit reimbursement, customers must return product subject to this recall. Attention Line: Cerebase FA2350411"
Recall details
- Recall number
- Z-1284-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 2, 2024
- Posted by the FDA
- Mar 13, 2024
- Recalling firm
- Medos International Sarl
- Quantity
- 1343 units
- Product code
- QJP
- FDA submissions
- K192804
- Distribution
- Worldwide distribution.
Product description
CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
a) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- CEREBASE GS9090SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall2 deaths97 reports
- CEREBASE GS9095SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall1 death33 reports
- CEREBASE GS9080SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall13 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.