Skip to content

LINK Embrace Shoulder System Line Extension recallZ-1258-2024 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Jan 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The surgical techniques for the Embrace Shoulder Instruments -Drill Tower (Article REF 645-081/62 & /63) have mixed-up type descriptions ("standard" vs "wedged") in the overview of the instruments.

FDA root cause category: Under Investigation by firm

What the company is doing

LinkBio Corp issued Urgent: Medical Device Correction letter via email on 2/8/24 to Distributors. Letter states reason for recall, health risk and action to take: Please take note of the correction of the Surgical Techniques noted in section 2. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to follow the updated surgical technique. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Recall details

Recall number
Z-1258-2024
Status
Open, Classified
Initiated by the company
Jan 29, 2024
Posted by the FDA
Mar 6, 2024
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
6 units
Product code
HTW
Distribution
US Nationwide distribution in the states of AL and LA.

Product description

Embrace Drill Tower, Standard/Lateral (25mm)

Code information (models, lots, serial numbers)

Item Number: 645-081/63; UDI/DI: 04026575534722; All lots of the LINK Embrace Drill Tower surgical instruments.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93979).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page