Battery Mobile X-ray Unit EASY MOVING Model with telescopic… recallZ-1754-2024 · Sedecal SA · initiated Jan 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sedecal SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential interference with the anti-fall system.

FDA root cause category: Component design/selection

What the company is doing

The firm issued letters dated 1/22/2024 via email to notify their distributor customers of the potential problem that could occur and the availability of a service kit that corrects the anomaly. The letter informs the distributor that a service kit has been created to install a new cable without the interference point in all of the installed bases of the devices. This kit contains parts for updating the parachute safety system making it similar to the one in production for the current product release. The entire installed base of Mobile System 3.1 must be inspected using the provided instructions in SIN 23-08-22 and the kit installed using the instructions in SIN 23-04-10 immediately if the cable could interfere with the parachute system. If no damage is found during the inspection, the kit can be installed in the next cable replacement scheduled according to the Service Manual instruction. An Acknowledgment Form was enclosed for completion and return as soon as possible that the distributor understands the letter. An Acknowledgment of Action Implemented was also enclosed for completion and return as soon as possible.

Recall details

Recall number
Z-1754-2024
Status
Open, Classified
Initiated by the company
Jan 22, 2024
Posted by the FDA
May 2, 2024
Recalling firm
Sedecal SA
Quantity
636 devices
Product code
IZL
FDA submissions
K161345
Distribution
Worldwide distribution - US Nationwide in the states of CA, IL, NJ, and areas yet to be identified by Philips Medical Systems DMC GmbH, Hamburg, Germany.

Product description

Battery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1, model numbers: a. SM-20HF-Batt; b. SM-32HF-Batt; c. SM-40HF-Batt; d. SM-50HF-Batt; e. SM-40-HF-B-D-VIR; f. SM-20HF-B-D-KM (AeroDR X30); g. SM-32-HF-B-D-KM (AeroDR X30); h. SM-40HF-B-D-KM (AeroDR X30); i. SM-50HF-B-D-KM (AeroDR X30); j. MobileDiagnost wDR.

Code information (models, lots, serial numbers)

All serial numbers of model 3.1 Battery Mobile X-ray Unit EASY MOVING are affected. Following are the UDI numbers involved: a. SM-20HF-Batt - UDI-DI 08436046001497 b. SM-32HF-Batt - UDI-DI 08436046001503 c. SM-40HF-Batt - UDI-DI 08436046001510; d. SM-50HF-Batt - UDI-DI 08436046001527; e. SM-40-HF-B-D-VIR - UDI-DI 08436046002166; f. SM-20HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002579; g. SM-32-HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002586; h. SM-40HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002593; i. SM-50HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002609; j. MobileDiagnost wDR - UDI-DI 08436046002357.