Battery Mobile X-ray Unit EASY MOVING Model with telescopic… recallZ-1754-2024 · Sedecal SA · initiated Jan 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sedecal SA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential interference with the anti-fall system.
FDA root cause category: Component design/selection
What the company is doing
The firm issued letters dated 1/22/2024 via email to notify their distributor customers of the potential problem that could occur and the availability of a service kit that corrects the anomaly. The letter informs the distributor that a service kit has been created to install a new cable without the interference point in all of the installed bases of the devices. This kit contains parts for updating the parachute safety system making it similar to the one in production for the current product release. The entire installed base of Mobile System 3.1 must be inspected using the provided instructions in SIN 23-08-22 and the kit installed using the instructions in SIN 23-04-10 immediately if the cable could interfere with the parachute system. If no damage is found during the inspection, the kit can be installed in the next cable replacement scheduled according to the Service Manual instruction. An Acknowledgment Form was enclosed for completion and return as soon as possible that the distributor understands the letter. An Acknowledgment of Action Implemented was also enclosed for completion and return as soon as possible.
Recall details
- Recall number
- Z-1754-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 22, 2024
- Posted by the FDA
- May 2, 2024
- Recalling firm
- Sedecal SA
- Quantity
- 636 devices
- Product code
- IZL
- FDA submissions
- K161345
- Distribution
- Worldwide distribution - US Nationwide in the states of CA, IL, NJ, and areas yet to be identified by Philips Medical Systems DMC GmbH, Hamburg, Germany.
Product description
Battery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1, model numbers: a. SM-20HF-Batt; b. SM-32HF-Batt; c. SM-40HF-Batt; d. SM-50HF-Batt; e. SM-40-HF-B-D-VIR; f. SM-20HF-B-D-KM (AeroDR X30); g. SM-32-HF-B-D-KM (AeroDR X30); h. SM-40HF-B-D-KM (AeroDR X30); i. SM-50HF-B-D-KM (AeroDR X30); j. MobileDiagnost wDR.
Code information (models, lots, serial numbers)
All serial numbers of model 3.1 Battery Mobile X-ray Unit EASY MOVING are affected. Following are the UDI numbers involved: a. SM-20HF-Batt - UDI-DI 08436046001497 b. SM-32HF-Batt - UDI-DI 08436046001503 c. SM-40HF-Batt - UDI-DI 08436046001510; d. SM-50HF-Batt - UDI-DI 08436046001527; e. SM-40-HF-B-D-VIR - UDI-DI 08436046002166; f. SM-20HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002579; g. SM-32-HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002586; h. SM-40HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002593; i. SM-50HF-B-D-KM(AeroDR X30) - UDI-DI 08436046002609; j. MobileDiagnost wDR - UDI-DI 08436046002357.
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- SEDECAL SM-40HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)24 recalls1 report
- SEDECAL SM-40HF-B-D-VIRSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)12 recalls1 report
- Philips MobileDiagnost wDR 2.2Sedecal · System, X-Ray, Mobile3 recalls7 reports
- SEDECAL SM-50HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)2 recalls0 reports
- SEDECAL SM-40HF-B-D-KM (AeroDR X30)Sedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)2 recalls0 reports
- SEDECAL SM-20HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)1 recall0 reports
- SEDECAL SM-32HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)1 recall0 reports
- SEDECAL SM-20HF-B-D-KM (AeroDR X30)Sedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)1 recall0 reports
- SEDECAL SM-32HF-B-D-KM (AeroDR X30)Sedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)1 recall0 reports
- SEDECAL SM-50HF-B-D-KM (AeroDR X30)Sedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page