DCA Systems HbA1c Reagent Kit recallZ-1304-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Feb 7, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.

FDA root cause category: Under Investigation by firm

What the company is doing

An Urgent Medical Device Correction notification letter was sent to customers on 02/07/2024. Actions to be Taken by the Customer: Discontinue use of the reagent lots listed in Table 1. We are offering you the option to formally request a complimentary replacement product for discarded reagent kits through your local Siemens representative or authorized distributor by calling 877-885-4873. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. If the lot number you are using is not listed in the table, you may continue using it. If you are a distributor, please ensure your customers receive this UMDC letter. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-1304-2024
Status
Open, Classified
Initiated by the company
Feb 7, 2024
Posted by the FDA
Mar 19, 2024
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
74,211 Kits (Expanded 2/28/24: additional 72,892 kits)
Product code
LCP
FDA submissions
K951361
Distribution
Worldwide distribution - US Nationwide and the country of Taiwan.

Product description

DCA Systems HbA1c Reagent Kit

Code information (models, lots, serial numbers)

Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.