Why it was recalled

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

FDA root cause category: Storage

What the company is doing

McKesson notified customers by phone January 9, 2024 with a follow-up call January 17, 2024 instructing customers of three scenarios based upon the products they received and the investigation in conjunction with relevant manufacturer. The firm also sent Urgent Product Recall Notifications via email on February 8, 2024 instructing customers to immediately examine inventory and quarantine all affected devices, complete and return the provided response form to [email protected], and notify customers who you have further distributed recalled devices to of this recall and to segregate and quarantine any devices they have on hand. Replacement products will be provided and may have the same lot number(s) as the recalled items but are not affected by the recall and are acceptable for use. Customers with questions can contact McKesson directly by phone at (800) 688-8840. Representatives are available Monday to Friday, 8:00 am to 5:00 pm PST.

Recall details

Recall number
Z-1313-2024
Status
Open, Classified
Initiated by the company
Jan 9, 2024
Posted by the FDA
Mar 19, 2024
Recalling firm
Mckesson Medical-Surgical Inc. Corporate Office
Quantity
25 units
Product code
MNL
Distribution
US Nationwide distribution in the states of AZ, PA and TX.

Product description

Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"

Code information (models, lots, serial numbers)

Catalogue Number: LCC101; UDI/DI: 10724995220072

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94024).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page