Arctic Sun Temperature Management System, ArcticGel Pads recallZ-1381-2024 · Medivance Inc. · initiated Feb 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medivance Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Neonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are part of a temperature management system.
FDA root cause category: Process control
What the company is doing
On 2/16/24, BD mailed or emailed recall notices to customers who were asked to do the following: 1) Immediately Discontinue Use 2) Immediately quarantine affected pads, record your inventory totals on the Customer Response Form, discard all devices within your facility's control per your facility's procedures. 3) Share this notice with any users of the product within your facilities or with any facility affected product was distributed to. 4) Complete and return the attached Customer Response Form to the BD contact noted on the form. If you require further assistance contact BD Customer Support at 1-844-823-5433. Say "Recall" when prompted M-F 8am -5pm CT.
Recall details
- Recall number
- Z-1381-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 16, 2024
- Posted by the FDA
- Mar 26, 2024
- Recalling firm
- Medivance Inc.
- Quantity
- 1,654
- Product code
- DWJ
- FDA submissions
- K142702
- Distribution
- Worldwide - US Nationwide distribution including in the states of IL, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Belgium, Australia, Singapore, South Korea, Taiwan, Japan, China, Thailand.
Product description
Medivance Neonatal ArcticGel Pads, REF: 31802, 3180202, 31802C
Code information (models, lots, serial numbers)
REF/UDI-DI/Lot(Expiration Date): 31802/00801741132131/NGGX5390(31Oct2024), NGGX1179(31Oct2024), NGGX1156(31Oct2024), NGHP2415(28Feb2025), NGHP2411(28Feb2025), NGHT1139(31May2025), NGHTY603(31May2025), NGHV4383(31Jul2025), NGHW1507(31Aug2025); 3180202/10801741132138/NGGZ2622(31Dec2024), NGGV1829(31Aug2024), NGGW3254(30Sep2024), NGHQ2940(31Mar2025); 31802C/00801741223457/NGGV1784(31Aug2024)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Arctic Sun® Temperature Management System, ArcticGel™ Pads 31802Medivance, Inc. · System, thermal regulating1 recall385 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page