Aptima recallZ-1371-2024 · Hologic, Inc. · initiated Sep 14, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Use of assay along with plasma specimen may result in error code which would invalidate sample.

FDA root cause category: Use error

What the company is doing

On February 23, 2024, Hologic issued a "Urgent Medical Device Recall Update" via FedEx. The update notified consignees of a sample dilution workflow for specific plasma specimens. On September 14, 2023 Hologic issued a "Urgent Medical Device Recall Notification" to affected consignees, via Fed Ex. Hologic provided the following information to customers: 1. Continue to use the Aptima CMV Quant assay. 2.When obtaining an ML2 error, please run the Mag Wash Clean Procedure and Contact Hologic Technical Support 3. With the Aptima CMV Quant assay, it is not recommended to retest a sample that was previously invalidated due to an ML2 error on the Panther instrument. 4. Complete Customer Response Form On (date) the firm issued a Customer Technical Bulletin, to consignees describing a sample dilution workflow for plasma specimens.

Recall details

Recall number
Z-1371-2024
Status
Open, Classified
Initiated by the company
Sep 14, 2023
Posted by the FDA
Mar 22, 2024
Recalling firm
Hologic, Inc.
Quantity
684 kits
Product code
PAB
Distribution
Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL, MD, MI, MO, TN, TX, WI and the countries of Australia, Germany, Denmark, Spain, France, Great Britain, Italy, Norway, Qatar, Sweden.

Product description

Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter Reagent, Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Promoter Reconstruction solution, Target capture Reagent, Reconstitution Collars.

Code information (models, lots, serial numbers)

Lot #: 402941/ UDI:15420045514553

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.