Why it was recalled
Boston Scientific is conducting a removal of specific batches of the iSLEEVETM 14F Expandable Introducer Set due to complaints associated with a hemostatic valve leak.
FDA root cause category: Process control
What the company is doing
Boston Scientific issued an Urgent Field Safety Notice - Medical Device Product Removal to its consignees on 02/20/2027 via FedEx. The notice explained the problem with the device, potential risk, and requested the following: 1. Immediately discontinue use and segregate the affected product. 2. Complete and return the RVTF to get a Return Goods Authorization (RGA) number. 3. Once Boston Scientific receives your completed RVTF, you will be contacted within 2 business days and provided an RGA number for product return. For questions, contact your local Boston Scientific representative.
Recall details
- Recall number
- Z-1390-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 20, 2024
- Posted by the FDA
- Mar 28, 2024
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 4060 devices
- Product code
- DYB
- FDA submissions
- K191871, K230051, K233503, K180785
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada.
Product description
Boston Scientific iSLEEVE EXPANDABLE Introducer Set REF H74939349140, facilitates femoral access to the vascular system
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
GTIN 08714729950660, Batch Numbers: 31368041, 31390084, 31390088, 31390089, 31501866, 31501867, 31501868, 31501869, 31520897, 31520898, 31604492, 31616143, 31633635, 31650457, 31664216, 31664217, 31671140, 31671143, 31693638, 31693639, 31724952, 31732930, 31732931, 31740806, 31740807, 31740808, 31751118, 31777134, 31786582, 31786583, 31818290, 31818299, 31818341, 31818342, 31818343, 31818344, 31818345, 31818346, 31818347, 31838684, 31838685, 31838686, 31838687, 31838688, 31838689, 31877302, 31877304, 31877305, 31877306, 31904748, 31904749, 31923755, 31923756, 31953345, 31953348, 32008170, 32031191, 32031192, 32031193, 32031196, 32059171, 32059386, 32059387, 32059388, 32059392, 32059393, 32059394, 32059396, 32059397, 32212151, 32212153, 32230544, 32230545, 32230546, 32230547, 32230548, 32230549, 32230550, 32230551, 32284196, 32284198, 32284200, 32304326, 32304327, 32304328, 32313757, 32313758, 32313759, 32322405, 32322406, 32322407, 32331162, 32331163, 32331164, 32331165, 32427040, 32452794, 32461848, 32461849, 32461850, 32470848, 32470849, 32480089, 32480090, 32481344, 32481345, 32522178, 32530450, 32530451, 32602892, 32602893, 32611557, 32611558, 32611559, 32620708, 32620709, 32620710, 32629426, 32642560, 32642561
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- iSLEEVE™ H74939349140Boston Scientific Corporation · INTRODUCER, CATHETER1 recall12 deaths243 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page