Mini Stick MAX recallZ-1347-2024 · Angiodynamics, Inc. · initiated Jan 30, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Angiodynamics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
FDA root cause category: Process control
What the company is doing
AngioDynamics initiated Urgent Voluntary Medical Device Recall Notification packets by Federal Express to consignees on 1/30/24. Letter states reason for recall, health risk and action to take: IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). Mark affected items with text QUARANTINE to segregate from usable stock. o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. o Notify Healthcare Providers of this possible issue that could occur during use of the Mini Stick Max device. 2. Complete and return the Reply Verification Tracking Form. " If you have any questions regarding this notice and/or if affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. " Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return), following the directions on this page and the Reply Verification Tracking Form. Email Reply Verification Tracking Form (preferred): [email protected] Fax Reply Verification Tracking Form: Attn: Mini Stick Max Recall Coordinator Fax number 1-855-273-0519 3. Package and Return the Recalled Product. " Package any product that is being returned in an appropriate shipping box. " Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. " Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: Mini Stick Max Recall Coordinator
Recall details
- Recall number
- Z-1347-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 30, 2024
- Posted by the FDA
- Mar 21, 2024
- Recalling firm
- Angiodynamics, Inc.
- Quantity
- 40 units
- Product code
- DRE
- FDA submissions
- K132713
- Distribution
- Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Product description
MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-776
Code information (models, lots, serial numbers)
UDI: 15051684023191 UPN: H965457761 Lot Number: 5787840 5803520 5805069
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Mini Stick MAX H965457760Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
Product lines
Other devices under the same FDA submission
23 devices cleared under K132713 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Mini Stick MAX H965457490Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
- Mini Stick MAX H965457510Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
- Mini Stick MAX H965457500Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457520Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457540Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457560Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457580Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457590Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457740Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457750Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- Mini Stick MAX H965457810Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall0 reports
- AngioDynamics 45759Navilyst Medical, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION24 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94035).
- Z-1334-2024 · Open, ClassifiedMINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-749
- Z-1335-2024 · Open, ClassifiedMINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750
- Z-1336-2024 · Open, ClassifiedMINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-751
- Z-1337-2024 · Open, ClassifiedMINI STICK MAX 4F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-752
- Z-1338-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG -Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-754
- Z-1339-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756
- Z-1340-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-758
- Z-1341-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-759
- Z-1342-2024 · Open, ClassifiedMINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-762
- Z-1343-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-765
- Z-1344-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-767
- Z-1345-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-774
- Z-1346-2024 · Open, ClassifiedMINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-775
- Z-1348-2024 · Open, ClassifiedMINI STICK MAX 4F X 10CM STIFF .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. - Catalog Number: 45-781
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page