SPII Model Lubinus recallZ-1296-2024 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Feb 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to two complaints, has come to our attention that additional guidance is required on the correct interpretation of the carton label Size and Type columns to avoid misinterpretation during surgery.
FDA root cause category: Under Investigation by firm
What the company is doing
Urgent Medical Device Correction (UMDC) Letters were sent via email to consignees on February 16, 2024. Actions to be taken by the customer/user: Please take note of the clarification of the carton label noted in section 2. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to correctly interpret the Size and Type columns on the label. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.
Recall details
- Recall number
- Z-1296-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 6, 2024
- Posted by the FDA
- Mar 14, 2024
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 382 units
- Product code
- JDG
- FDA submissions
- K953653
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, NJ, NV, PA, TN, TX, VA, WA.
Product description
SPII Model Lubinus, Hip Prosthesis Stem Standard Neck, cemented
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Product Code (UDI/DI): 127-610/17 (04026575387168) , 127-610/26 (04026575387304) , 127-610/35 (04026575387441), 127-612/17 (04026575387175) , 127-612/26 (04026575387311) , 127-612/35 (04026575387458) , 127-614/17 (04026575387182) , 127-614/26 (04026575387328) , 127-614/35 (04026575387465) , 127-616/17 (04026575387199) , 127-616/26 (04026575387335), 127-616/35 (04026575387472) , 127-618/17 (04026575387205) , 127-618/26 (04026575387342) , 127-618/35 (04026575387489) , 127-620/17 (04026575387212) , 127-620/26 (04026575387359) , 127-620/35 (04026575387496) , 127-622/17 (04026575387229) , 127-622/26 (04026575387366) , 127-622/35 (04026575387502) , 127-624/17 (04026575387236) , 127-624/26 (04026575387373) , 127-624/35 (04026575387519) , 127-626/17 (04026575387243) , 127-626/26 (04026575387380) , 127-626/35 (04026575387526) , 127-628/17 (04026575387250) , 127-628/26 (04026575387397) , 127-628/35 (04026575387533) , 127-630/17 (04026575387267) , 127-630/26 (04026575387403) , 127-630/35 (04026575387540) , 127-632/17 (04026575387274) , 127-632/26 (04026575387410) , 127-632/35 (04026575387557) , 127-634/17 (04026575387281) , 127-634/26 (04026575387427) , 127-634/35 (04026575387564) , 127-636/17 (04026575387298) , 127-636/26 (04026575387434) , 127-636/35 (04026575387571), 127-710/17 (04026575232154) , 127-710/26 (04026575232161) , 127-710/35 (04026575232178) , 127-711/17 (04026575232291) , 127-711/26 (04026575232307) , 127-711/35 (04026575232314) , 127-712/17 (04026575232147) , 127-712/26 (04026575024032) , 127-712/35 (04026575024049) , 127-713/17 (04026575232246) , 127-713/26 (04026575024056) , 127-713/35 (04026575024063) , 127-714/17 (04026575232192) , 127-714/26 (04026575024070) , 127-714/35 (04026575024087) , 127-715/17 (04026575230150) , 127-715/26 (04026575024094) , 127-715/35 (04026575024100) , 127-716/17 (04026575232208) , 127-716/26 (04026575024117) , 127-716/35 (04026575024124) , 127-717/17 (04026575232253) , 127-717/26 (04026575024131) , 127-717/35 (04026575024148) , 127-718/17 (04026575232215) , 127-718/26 (04026575024155) , 127-718/35 (04026575024162) , 127-719/17 (04026575232260) , 127-719/26 (04026575024179) , 127-719/35 (04026575024186) , 127-720/17 (04026575229628) , 127-720/26 (04026575225422) , 127-720/35 (04026575225439) , 127-721/17 (04026575229697) , 127-721/26 (04026575225507) , 127-721/35 (04026575225514) , 127-722/17 (04026575229635) , 127-722/26 (04026575024193) , 127-722/35 (04026575024209) , 127-723/17 (04026575229703) , 127-723/26 (04026575024216) , 127-723/35 (04026575024223), 127-724/17 (04026575229642) , 127-724/26 (04026575024230) , 127-724/35 (04026575024247) , 127-725/17 (04026575229710) , 127-725/26 (04026575024254) , 127-725/35 (04026575024261) , 127-726/17 (04026575229659) , 127-726/26 (04026575024278) , 127-726/35 (04026575024285) , 127-727/17 (04026575229727) , 127-727/26 (04026575024292) , 127-727/35 (04026575024308) , 127-728/17 (04026575229666) , 127-728/26 (04026575024315), 127-728/35 (04026575024322) , 127-729/17 (04026575229734) , 127-729/26 (04026575024339) , 127-729/35 (04026575024346) , 127-730/17 (04026575232239) , 127-730/26 (04026575024353) , 127-730/35 (04026575024360) , 127-731/17 (04026575232284) , 127-731/26 (04026575024377) , 127-731/35 (04026575024384) , 127-732/17 (04026575229680) , 127-732/26 (04026575024391) , 127-732/35 (04026575024407) , 127-733/17 (04026575229758) , 127-733/26 (04026575024414) , 127-733/35 (04026575024421) , 127-736/17 (04026575232222) , 127-736/26 (04026575225408) , 127-736/35 (04026575225415) , 127-737/17 (04026575232277) , 127-737/26 (04026575225460) , 127-737/35 (04026575225477) , 127-738/17 (04026575229673) , 127-738/26 (04026575225446) , 127-738/35 (04026575225453) , 127-739/17 (04026575229741) , 127-739/26 (04026575225484) , 127-739/35 (04026575225491);
Devices named in this recall
99 devices in the FDA device registry whose identifier appears in the recall's code information.
- SPII Model Lubinus 127-720/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall2 reports
- SPII Model Lubinus 127-728/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall1 report
- SPII Model Lubinus 127-721/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall1 report
- SPII Model Lubinus 127-712/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-712/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-713/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-713/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-714/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-714/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-715/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-715/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-716/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-716/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-717/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-717/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-718/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-718/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-719/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-719/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-722/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-722/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-723/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-723/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-724/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-724/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-725/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-725/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-726/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-726/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-727/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-727/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-728/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-729/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-729/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-730/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-730/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-731/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-731/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-732/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-732/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-733/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-733/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-736/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-736/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-720/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-738/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-738/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-737/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-737/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-739/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-739/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-721/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-725/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-710/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-710/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-711/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-711/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-610/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-612/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-614/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-616/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-618/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-620/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-622/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-624/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-626/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-628/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-630/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-632/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-634/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-636/17WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-610/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-612/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-614/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-616/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-618/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-620/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-622/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-624/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-626/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-628/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-630/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-632/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-634/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-636/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-610/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-612/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-614/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-616/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-618/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-620/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-622/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-624/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-626/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-628/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-630/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-632/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-634/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-636/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
Product lines
Other devices under the same FDA submission
77 devices cleared under K953653 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SPII Model Lubinus 127-922/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall1 report
- SPII Model Lubinus 127-767/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall1 report
- SPII Model Lubinus 127-910/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-910/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-911/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-911/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-912/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-912/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-913/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-913/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-914/26WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
- SPII Model Lubinus 127-914/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Femoral Component, Cemented, Metal1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94078).
- Z-1297-2024 · Open, ClassifiedSPII Model Lubinus, Hip Prosthesis Stem XL Neck, cemented
- Z-1298-2024 · Open, ClassifiedSPII Model Lubinus, Long Stem Prosthesis Standard Neck, cemented
- Z-1299-2024 · Open, ClassifiedSPII Model Lubinus, Long Stem Prosthesis XL Neck, cemented
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page