Cystatin C Calibrator - In vitro diagnostic (IVD) use in… recallZ-1369-2024 · Randox Laboratories Ltd. · initiated Feb 14, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods
FDA root cause category: Under Investigation by firm
What the company is doing
Randox Laboratories Ltd, UK ( manufacturer) notified the US Distributor Randox Laboratories USA . The distributor initial contact and the follow up contact via email on 2/20/24. Letter states reason for recall, health risk and action to take: We have realigned the target values in the above lot numbers to the IFCC Reference Standard. Please refer to the table below for the updated calibrator targets. You may experience a positive shift in Quality Control and patient sample recovery. If you are using CVS 2699 batch 600957 lots 1378CY-1382CY please provide proof of scrappage and a replacement calibrator lot will be provided. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com. If further information is required, please contact [email protected] . In line with the realignment to the Cystatin C Calibrator Series, we are updating the target and range for the Cystatin C controls, CYS5019 and CYS5020. The updated target and range can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
Recall details
- Recall number
- Z-1369-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 14, 2024
- Posted by the FDA
- Mar 22, 2024
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- 666 units
- Product code
- NDY
- Distribution
- US Nationwide distribution in the states of CA, IN.
Product description
Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog Number: CVS 2699
Code information (models, lots, serial numbers)
GTIN: 05055273207439 (1) Batch 600957 Lot 1378CY - 1382CY (2) Batch 621809, 621815 Lot 1403CY - 1407CY (3) Batch 650533 Lot 1414CY - 1418CY
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CYSC CAL CYS2699Randox Laboratories Limited · Calibrator, Secondary1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page