Why it was recalled

Endotracheal Tube Holder included in certain kits is not compatible with a nasal intubation procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

FDA root cause category: Component design/selection

What the company is doing

Bound Tree issued URGENT NOTICE OF RECALL letters on February 12, 2024 via USPS Letter to Medical Inventory Controls/Supplies Coordinator/Buyer. Letter states reason for recall, health risk and actin to take: Should you determine your inventory contains product covered by this recall, please immediately quarantine your affected product and contact Bound Tree Medical Customer Care at 800.533.0523 to arrange for a return of the product in exchange for a credit

Recall details

Recall number
Z-1307-2024
Status
Open, Classified
Initiated by the company
Feb 12, 2024
Posted by the FDA
Mar 19, 2024
Recalling firm
Sarnova HC, Llc
Quantity
540 kits
Product code
LNZ
Distribution
US Nationwide distribution.

Product description

Curaplex Nasal Intubation Kit with 6mm Endotrol Tube-To perform a nasal intubation to establish an airway in an emergency. Item Number: 023060S

Code information (models, lots, serial numbers)

UDI/GTIN: 00815277020008 each, 00810071639630 case Lot Numbers: ASM0038077 ASM0038129 ASM0038334 ASM0038598