Nimbus recallZ-1285-2024 · InfuTronix LLC · initiated Feb 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from InfuTronix LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
InfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing. The device will not be available or supported after June 20, 2024.
FDA root cause category: Device Design
What the company is doing
On February 21, 2024, the firm began notifying customers via URGENT MEDICAL DEVICE REMOVAL letters. Customers were informed that the identified Nimbus Infusion Pumps are being removed from the market. The products will no longer be supported by InfuTronix for the Nimbus Infusion Pumps or related infusion sets beyond June 20, 2024. Customers of InfuTronix who have Nimbus Infusion Pumps and associated infusion sets in their possession should contact InfuTronix customer service at [email protected], where they will be instructed on the Return Material Authorization (RMA) process to be followed for returning the both the Nimbus Infusion Pumps and associated infusion sets. As the device will not be available or supported after June 20, 2024, InfuTronix encourages the healthcare provider to seek alternative methods of drug product infusion appropriate for their patient s needs, based on their medical expertise, at the earliest possible opportunity. ***Updated 4/12/24*** The firm issued updated letters on 4/9/24 to reflect that the recall had been classified Class I by the FDA. Effective immediately, Nimbus Pumps will no longer be shipped. However, administration sets will continue to be shipped through June 20, 2024.
Recall details
- Recall number
- Z-1285-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 21, 2024
- Posted by the FDA
- Mar 28, 2024
- Recalling firm
- InfuTronix LLC
- Quantity
- 52,328 total
- Product code
- FRN
- FDA submissions
- K140783, K153193
- Distribution
- Domestic US distribution nationwide. No international distribution.
Product description
Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog Number IT1025. Ambulatory Infusion Pump and components
Code information (models, lots, serial numbers)
UDI-DI 00817170020000 Lots A100-150205200 A100-150415015 A100-150430270 A100-150611300 A100-150716620 A100D-1705191000 AZ060-171201564 AZ060-1801021020 AZ060-180609264 AZ060-180626552 AZ060-190104792 AZ060-1902151200 A100F-160707073 A100F-160908250 A100F-161114700 All units recalled
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NimbusInfutronix LLC · Pump, Infusion, Pca1 recall4 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K140783, K153193 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Nimbus II PlusInfutronix LLC · Pump, Infusion, Pca1 recall1 death800 reports
- Nimbus II PainPROInfutronix LLC · Pump, Infusion, Pca2 recalls186 reports
- Nimbus II FlexInfutronix LLC · Pump, Infusion, Pca1 recall1 death131 reports
- Nimbus HS-002Infutronix LLC · Set, Administration, Intravascular1 recall100 reports
- Nimbus Nimbus II EMSInfutronix LLC · Pump, Infusion, Pca1 recall0 reports
- Nimbus II EpiDInfutronix LLC · Pump, Infusion1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94124).
- Z-1286-2024 · Open, ClassifiedNimbus II Infusion System; Nimbus II PainPRO IT1053; Nimbus II Flex IT1062; Nimbus II Plus IT1078; Nimbus II EpiD IT1124; Nimbus II EMS IT1063; Ambulatory Infusion Pump and…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.