Auto Suture recallZ-1470-2024 · Covidien, LP · initiated Feb 28, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for trocar seal disengagement when using mesh products incorrectly with the device.
FDA root cause category: Labeling design
What the company is doing
On February 28, 2024, the firm notified customers of the issue via URGENT: MEDICAL DEVICE RECALL letters. Actions to be taken by the customer: 1. Inform all surgeons and clinicians who handle the preparation and/ or placement of a Mesh Device that utilize the balloon and blunt tip trocar devices. 2. Prior to using any Mesh device in conjunction with the following Trocars: Covidien Auto Suture" Structural Balloon Trocar and Auto Suture" Blunt Tip Trocar, please carefully review and adhere to the mesh manufacturer s Instructions for Use (IFU) on proper insertion techniques. 3. Please complete and return the customer confirmation form enclosed with this letter acknowledging receipt of this information via email to [email protected]. 4. Please transfer this notice to other organizations on which this action has an impact and maintain a copy of this notice in your records. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) and Medtronic Vascular Customer Service at 800-716-6700. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Representative.
Recall details
- Recall number
- Z-1470-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 28, 2024
- Posted by the FDA
- Apr 3, 2024
- Recalling firm
- Covidien, LP
- Quantity
- 165,356 units
- Product code
- GCJ
- FDA submissions
- K924011
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
Product description
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BT
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Product Number/CFN: OMS-T10BT; UDI/DI: 10884521080874, 20884521080871; Lot Serial Number: P1D1353, P1D1353, P1D1355, P1D1409, P1D1410, P1D1506, P1D1506, P1D1507, P1D1507, P1E1402, P1E1402, P1E1403, P1E1403, P1E1505, P1E1505, P1E1506, P1E1506, P1F0061, P1F0061, P1F0062, P1F0062, P1F0063, P1F1018, P1F1018, P1F1019, P1F1388, P1F1388, P1F1546, P1G0140, P1G0140, P1G1117, P1G1117, P1G1119, P1G1119, P1G1121, P1G1121, P1G1364, P1G1364, P1H0049, P1H0049, P1H0421, P1H0527Q, P1H0527Q, P1H0859, P1H0859, P1H0860, P1H0860, P1J1272, P1J1321, P1J1321, P1J1339, P1J1339, P1J1342, P1J1342, P1K0179, P1K0179, P1K0180, P1K0180, P1K0182, P1K0182, P1K0353, P1K0353, P1K0560, P1K0560, P1K0561, P1K0561, P1K0589, P1K0589, P1L0490, P1L0490, P1L0681, P1L0681, P1L0705, P1L0705, P1L0976, P1L0976, P1M0021, P1M0021, P1M0022, P1M0744, P1M0744, P1M0745, P1M0970, P1M0971, P1M1151, P1M1152, P2A0118, P2A0118, P2B0028Q, P2B0137, P2B0137, P2B0139, P2B0139, P2B0308, P2B0745, P2B0745, P2B0912, P2B0912, P2B0913, P2B0913, P2C0175, P2C0175, P2C0178, P2C0643, P2C0643, P2C0758, P2C0758, P2C0764, P2C0765, P2C0765, P2C0877, P2C0877, P2C0878, P2C0878, P2D0231, P2D0231, P2D0232, P2D0232, P2D0373, P2D0373, P2D0374, P2D0374, P2D0415, P2D0415, P2D0539, P2D0539, P2D0540, P2D0540, P2E0244, P2E0244, P2E0245, P2E0371, P2E0371, P2F0041, P2F0041, P2F0042, P2F0133, P2F0133, P2F0134, P2F0134, P2F0135, P2F0135, P2F0449, P2F0449, P2F0450, P2F0450, P2F0451, P2F0451, P2F0452, P2F0452, P2F0453, P2F0453, P2F0454, P2F0454, P2G0141Q, P2G0533, P2G0533, P2H0085, P2H0085, P2H0142, P2H0142, P2H0419, P2H0419, P2H0570, P2H0570, P2J0081, P2J0081, P2J0362, P2J0362, P2J0442, P2J0442, P2J0443, P2J0443, P2J0460, P2J0460, P2K0080, P2K0080, P2K0123, P2K0123, P2K0432, P2K0432, P2K0565, P2K0565, P2L0554, P2L0554, P2L0555, P2L0555, P2L0556, P2L0556, P2M0032, P2M0408, P2M0408, P2M0409, P2M0410, P2M0410, P2M0411, P2M0411, P3A0084, P3A0084, P3A0085, P3A0086, P3A0086, P3A0087, P3A0271, P3A0271, P3A0272, P3A0272, P3A0273, P3A0273, P3A0277, P3A0277, P3A0278, P3A0278, P3A0279, P3A0279, P3B0075, P3B0075, P3B0147, P3B0147, P3B0158, P3B0158, P3B0162Q, P3D0121, P3D0121, P3D0122, P3D0122, P3D0150, P3D0150, P3D0151, P3D0151, P3D0152, P3D0152, P3D0399, P3D0400, P3D0400, P3D0401, P3D0401, P3E0294, P3E0294, P3E0295, P3E0295, P3E0315, P3E0315, P3E0369, P3E0369, P3F0461, P3F0461, P3F0462, P3F0462, P3H1033, P3H1033, P3H1034, P3H1193, P3H1193, P3H1194, P3H1194, P3H1195, P3H1195, P3J0855, P3J0855, P3J0856, P3J0857, P3J0857, P3J1100, P3J1100, P3J1101, P3J1101, P3J1102, P3J1103, P3J1103, P3K0951, P3K1067, P3K1068, P3K1069, P3K1243, P3K1244, P3K1339, P3K1340, P3L0889, P3L0890;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K924011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Auto Suture OMS-T12BTCovidien LP · Laparoscope, general & plastic surgery1 recall475 reports
- Auto Suture OMS-T10BTNLCovidien LP · Laparoscope, general & plastic surgery1 recall218 reports
- Auto Suture OMST12BTNLCovidien LP · Laparoscope, general & plastic surgery1 recall60 reports
- Auto Suture OMS-T10BTSCovidien LP · Laparoscope, general & plastic surgery1 recall35 reports
- Auto Suture OMST10BTSNLCovidien LP · Laparoscope, general & plastic surgery1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94131).
- Z-1465-2024 · Open, ClassifiedCovidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB
- Z-1466-2024 · Open, ClassifiedCovidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL
- Z-1467-2024 · Open, ClassifiedCovidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL
- Z-1468-2024 · Open, ClassifiedCovidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BT
- Z-1469-2024 · Open, ClassifiedCovidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS
- Z-1471-2024 · Open, ClassifiedCovidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BTNL
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page