Skip to content

Centurion Hemostat kits labeled as recallZ-1438-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 23, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

FDA root cause category: Process control

What the company is doing

Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.

Recall details

Recall number
Z-1438-2024
Status
Open, Classified
Initiated by the company
Feb 23, 2024
Posted by the FDA
Mar 29, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
80903 units
Product code
HRQ
Distribution
Worldwide distribution - US Nationwide and the countries of Panama, Canada.

Product description

Centurion Hemostat kits labeled as: a) STERILE CURV KELLY HEMOSTAT, Product Code 65340; b) STERILE CURV MOSQUITO HEMOSTAT (MHS103), Product Code 66145; c) STERILE STRT KELLY HEMOSTAT (SK8039), Product Code 66170; d) STERILE CURV KELLY HEMOSTAT (CK8040), Product Code 66175; e) STERILE CURV SATIN HEMOSTAT (MHS103S), Product Code 66445; f) STERILE STRT SATIN HEMOSTAT (MHS102S), Product Code 66485; g) STERILE STRT KELLY HEMOSTAT (SK8039S), Product Code 66695; h) STERILE STRAIGHT KELLY HEMOSTAT, Product Code 67140; i) ST. STR MOSQUITO HEMOSTAT 3 1/2"(MHB700), Product Code 67325; j) STERILE CVD MOSQT HEMOST 3 1/2"(MHB710), Product Code 67330; k) STERILE 8" CRVD KELLY HEMOSTAT (CK1008), Product Code 67655; l) STERILE 8" STR KELLY HEMOSTAT (SK1008), Product Code 67660; m) ST MOSQUITO HEMO CURVED FINE PT (P/S), Product Code I68100; n) ST CURVED KELLY HEMOSTAT-TAMPA GEN, Product Code I68310; o) STERILE STRT KELLY HEMOSTAT (ZM-09417), Product Code P-65240; p) STERILE STRT MOSQUITO HEMOSTAT (MHS102), Product Code TRI66140; q) STERILE CURVED KELLY HEMO (ST153), Product Code TRI66765

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

a) 65340, UDI/DI 10653160038023 (case) 00653160038026 (each), Lot Numbers: 2023041090, 2023061290, 2023091190; b) 66145, UDI/DI 70653160008783 (case) 00653160008784 (each), Lot Numbers: 2023042690, 2023042790, 2023072790; c) 66170, UDI/DI 70653160024035 (case) 00653160024036 (each), Lot Numbers: 2023031090, 2023041790; d) 66175, UDI/DI 60653160039575 (case) 00653160039573 (each), Lot Numbers: 2023031390, 2023050990, 2023082490, 2023103090; e) 66445, UDI/DI 70653160023779 (case) 00653160023770 (each), Lot Numbers: 2023070390, 2023071090; f) 66485, UDI/DI 60653160029200 (case) 00653160029208 (each), Lot Numbers: 2023042690, 2023090790; g) 66695, UDI/DI 60653160024861 (case) 00653160024869 (each), Lot Numbers: 2023042790; h) 67140, UDI/DI 60653160049871 (case) 00653160049879 (each), Lot Numbers: 2023112990; i) 67325, UDI/DI 10653160110675 (case) 00653160110678 (each), Lot Numbers: 2023042490; j) 67330, UDI/DI 30653160110730 (case) 00653160110739 (each), Lot Numbers: 2023042890; k) 67655, UDI/DI 10653160196006 (case) 00653160196009 (each), Lot Numbers: 2023030290; l) 67660, UDI/DI 10653160196013 (case) 00653160196016 (each), Lot Numbers: 2023031390, 2023101990; m) I68100, UDI/DI 10653160235811 (case) 00653160235814 (each), Lot Numbers: 2023041190, 2023060790; n) I68310, UDI/DI 00653160278835 (case) 10653160278832 (each), Lot Numbers: 2023022190; o) P-65240, UDI/DI 10653160359234 (case) 20653160359231 (each), Lot Numbers: 2023120590; p) TRI66140, UDI/DI 00653160317855 (case) 10653160317852 (each), Lot Numbers: 2023032890, 2023061690; q) TRI66765, UDI/DI 00653160317947 (case) 10653160317944 (each), Lot Numbers: 2023030990

Devices named in this recall

17 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 94136).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page