Medline Sub-G Endotracheal Tube with Subglottic Suctioning recallZ-1876-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued letters via email and first class mail on 2/26/2024. The recall letter informed the customer the recall was being conducted for specific items and lots of SubG Endotracheal Tube with Subglottic Suction due to receiving complaints that the inflation tube detached and/or tore from the main tube. There had also been reports indicating the suction pump is difficult to connect or detaches during use. The potential product impact was also provided. Affected product could be destroyed for credit and no product was to be returned. Required actions were provided: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. The link to the recall portal was provided. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form. The letter provided a link to their recall portal to access the response form for completion and view the specific products distributed to the respective customers. If the consignee is a distributor or has resold or transferred the product to another company or individual, they are to contact them and notify them of this recall communication and request they document and destroy any affected product and include the customer quantities in the distributor response form. The recalling firm issued a second recall letter dated 3/19/2024. The letter informed the consignee that the recalling firm was issuing a recall for specific items and lots of SubG Endotracheal Tube with Subglottic Suction with Medline Kits. If the consignee has the affected kits, they are to indicate the quantity
Recall details
- Recall number
- Z-1876-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 26, 2024
- Posted by the FDA
- May 28, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 168,632 tubes and 13,092 kits
- Product code
- BTR
- FDA submissions
- K110269
- Distribution
- Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.
Product description
Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5 mm ID; (3) REF DYNJ18870, Polyurethane Cuff 7.0 mm ID; (4) REF DYNJ18875, Polyurethane Cuff 7.5 mm ID; (5) REF DYNJ18880, Polyurethane Cuff 8.0 mm ID; and (6) REF DYNJ18885, Polyurethane Cuff 8.5 mm ID. Medline SubG Endotracheal Tube with Subglottic Suctioning packaged into the following convenience kits: (1) Medline Intubation Kit - Drawer 3, Reorder No. ACC010502, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (2) Medline Intubation Tube - Drawer 3, Reorder No. ACC010527, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (3) Medline Adult Intubation, Reorder No. ACC010540, containing 8.5 size subglottic cuff, 3 kits/case; (4) Medline Adult Intubation, Reorder No. ACC010540A, containing 8.5 size subglottic cuff, 3 kits/case; (5) Medline Airway Kit - Drawer 3, Reorder No. ACC010717, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 3 kits/case; (6) Medline Airway Kit, Reorder No. DYNDJ1132, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 26 kits/case; (7) Medline Adult Intubation W Evac, Reorder No. DYNDJ1133, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 10 kits/case; (8) Medline Backup Kit, Reorder No. DYNJ909501, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (9) Medline Backup Kit, Reorder No. DYNJ909501A, containing 6.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (10) Medline Adult Intubation Tray, Reorder No. DYNJAA269, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 10 kits/case.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI-DI numbers for the individual tubes: (1) DYNJ18860 - UDI-DI 20888277652672; (2) DYNJ18865 - UDI-DI 20888277652689; (3) DYNJ18870 - UDI-DI 20888277652696; (4) DYNJ18875 - UDI-DI 20888277652702; (5) DYNJ18880 - UDI-DI 20888277652719; (6) DYNJ18885 - UDI-DI 20888277657318. Kit code information: (1) Kit ACC010502 - Kit UDI-DI: 10193489846171; Case UDI-DI: 40193489846172; lot numbers: 23FDC233, 23FDB211, 23EDC153, 23EDB775, 23DDB280, 23CDB943, 23CDA556, 22LDA029, 22IDA550, 22HDA813, 22FDC033, 22CDC284, 21JDC054, 21IDB345, 21HDB405, and 21DDB305; (2) Kit ACC010527 - Kit UDI-DI: 10193489884333; Case UDI-DI: 40193489884334; lot numbers: 23JDA443, 23IDA363, 23IDA232, 23HDC331, 23HDA117, 23GDB600, 23FDC372, 23FDB913, 23FDA966, 23FDA310, 23EDC371, 23EDC154, 23EDC094, 23EDB732, 23EDA387, 23CDB838, 23CDA557, 23BDC008, 23BDA945, 23ADB791, 22LDA005, 22JDB800, 22HDB976, 22HDA451, 22GDB041, 21LDB937, 21LDB416, 21JDB262, 21IDB491, 21IDA193, 21GDB207, and 21FDC754; (3) Kit ACC10540 - Kit UDI-DI: 10193489977547; Case UDI-DI 40193489977548; lot number: 23ADB410; (4) Kit ACC10540A - Kit UDI-DI 10195327330088; Case UDI-DI 40195327330089; lot numbers: 23HDA588, 23GDB107, and 23BDA684; (5) Kit ACC010717 - Kit UDI-DI 10195327458997; Case UDI-DI 40195327458998; lot numbers: 24ADA577, 23LDA731, and 23IDA629; (6) Kit DYNDJ1132 - Kit UDI-DI 10195327454821; Case UDI-DI 40195327454822; lot numbers: 23LDA733, 23LDA617, 23JDC021, and 23HDA569; (7) Kit DYNDJ1133 - Kit UDI-DI 10195327428532; Case UDI-DI 40195327428533; lot numbers: 24BMB124, 23LMB682, 23HMF259, and 23HMD765; (8) Kit DYNJ909501 - Kit UDI-DI 10195327384852; Case UDI-DI 40195327384853; lot number: 23GBI504; (9) Kit DYNJ909501A - Kit UDI-DI 10195327549817; Case UDI-DI 40195327549818; lot numbers: 24BBD129, 24ABK281, and 23KBR773; (10) Kit DYNJAA269 - Kit UDI-DI 10195327209049; Case UDI-DI 40195327209040; lot numbers: 24BBE224, 24ABM718, 24ABD835, 23LBB055, 23JBP694, 23JBQ783, 23IBU821, 23IBF348, 23HBP188, 23CBX318, 23CBX319, 23BBR494, 23ABI514,22LBI988, 22KBH009, and 22JBH478.
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. ACC010540Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline DYNJ18875Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall6 reports
- Medline DYNJ18870Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall4 reports
- Medline DYNJ18880Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall2 reports
- Medline DYNJ18860Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall1 report
- Medline Industries, Inc. ACC010527Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJAA269Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. ACC010540AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJ909501Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNDJ1133Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNDJ1132Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. ACC010717Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJ909501AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline DYNJ18865Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall0 reports
- Medline DYNJ18885Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page