VITROS recallZ-1373-2024 · Ortho-Clinical Diagnostics · initiated Feb 27, 2024

Class II Open, Classified All recalls posted in March 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

FDA root cause category: Under Investigation by firm

What the company is doing

Consignees were sent an URGENT PRODUCT CORRECTION NOTIFICATION dated 2/27/24 informing them that for product Lot Q1174, diluent Lot M9815 was found in some sales units instead of the expected diluent Lot Q1175, and for product Lot R1176, diluent Lot P9967 was found in some sales units instead of the expected diluent Lot R1177. Consignees were asked to inspect their inventory to see if any mismatched product from affected lots was present and to discard any affected units. The provided Confirmation of Receipt form is to be returned by 3/8/24. Results received using affected devices are to be reviewed for any unexpected QC shift. The recall notification is to be forwarded if any devices were further distributed and the notification maintained in the User Documentation. Customers with any questions can call Global Services at 1-800-421-3311.

Recall details

Recall number
Z-1373-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 27, 2024
Posted by the FDA
Mar 22, 2024
Recalling firm
Ortho-Clinical Diagnostics
Quantity
11,224 units
Product code
JJY
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS Performance Verifier II

Code information (models, lots, serial numbers)

Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94174).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page