Skip to content

X-Guide Handpiece Adaptor Sleeve 1, Model P010701 recallZ-1392-2024 · X-NAV Technologies, LLC · initiated Mar 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from X-NAV Technologies, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Handpiece Adaptor may have a manufacturing defect causing incorrect geometry. The incorrect geometry in turn causes the Handpiece Adaptor to not fit onto the doctor's dental handpiece.

FDA root cause category: Use error

What the company is doing

An URGENT MEDICAL DEVICE RECALL dated 3/4/24 was sent to customers. Action To Be Taken by the User Identify if you possess any of the affected devices. The lot numbers of the affected devices are the following: 2308010012 2306020012 2310010012 If the lot number on your device is not one of the above, then that Adaptor Sleeve is not part of this Recall. If you have a defective device, quarantine the device so that it is prevented from being used. Finally, destroy the device or return to X-Nav Technologies, LLC. Return the response form. X-Nav will ship all affected users a new Handpiece Adaptor Sleeve to replace their defective unit free of charge. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organizations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback.

Recall details

Recall number
Z-1392-2024
Status
Open, Classified
Initiated by the company
Mar 4, 2024
Posted by the FDA
Mar 29, 2024
Recalling firm
X-NAV Technologies, LLC
Quantity
88 units
Product code
QRY
FDA submissions
K232148
Distribution
Worldwide distribution - US Nationwide and the countries of China, Europe, Japan.

Product description

X-Guide Handpiece Adaptor Sleeve 1, Model P010701

Code information (models, lots, serial numbers)

UDI-DI: 00817421021480; Lot: 2308010012

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 94175).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page