RAD-G recallZ-1538-2024 · Masimo Corporation · initiated Feb 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Masimo Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is a potential that Oximeter may automatically power off and on resulting in loss of monitoring.
FDA root cause category: Material/Component Contamination
What the company is doing
On February 14, 2024 Masimo issued a "Urgent: Medical Device Recall Notice" via FedEx and US Mail to affected consignees. On February 29, 2024, Masimo issued an updated communication to provide additional information. Masimo's asked consignees to take the following actions: " Please assess your inventory to confirm if you have Rad-G devices listed in Attachment 2-part number(s) and serial number(s). " If you identify affected Rad-G devices, please promptly remove them from use and follow the steps below. " Contact Masimo Technical Services to request return of the affected part and serial number (s) for repair or replacement. " Masimo s Technical Services is available Monday through Friday 6:00am -5:00pm PST for live assistance or 24 hours a day to leave a message and someone will contact you. Local contact information can be found at http://service.masimo.com. Complete Attachment 1, then email it to [email protected], or Fax it to 1-949-297-7575 by MM DD, 2023. Starting on 8/13/24, additional recall notices were sent identifying additional affected devices. Distributors were asked to identify their direct customers and provide them with the recall notice.
Recall details
- Recall number
- Z-1538-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 15, 2024
- Posted by the FDA
- Apr 12, 2024
- Recalling firm
- Masimo Corporation
- Quantity
- 4978 units
- Product code
- DQA
- FDA submissions
- K201770
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Belize, Bolivia, Burkina Faso, Burundi, Cambodia, Canada, Chile, Colombia, Congo, The Democratic Republic of, Costa Rica, Cyprus, Czech Republic, Ethiopia, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Lesotho, Mozambique, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Poland, Puerto Rico, Qatar, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Tanzania, United Republic Of, Thailand, Trinidad And Tobago, Uganda, United Kingdom, Uruguay, Yemen, Zimbabwe.
Product description
Masimo Rad-G, Pulse Oximeter (W/Patient Cable), REF:9849, Rx Only,
Code information (models, lots, serial numbers)
Show the full code information (55k characters)
UDI:(01)00843997000666/Serial Number: 6000063773 6000063918 6000064156 6000064251 6000064298 6000065602 6000066119 6000066282 6000066408 6000066411 6000066442 6000066476 6000066577 6000066623 6000066856 6000067119 6000067120 6000067122 6000067134 6000067140 6000067142 6000067145 6000067150 6000067151 6000067162 6000067163 6000067167 6000067168 6000067170 6000067173 6000067175 6000067176 6000067177 6000067178 6000067179 6000067180 6000067182 6000067183 6000067185 6000067186 6000067188 6000067192 6000067193 6000067194 6000067196 6000067198 6000067201 6000067202 6000067207 6000067211 6000067224 6000067225 6000067235 6000067241 6000067245 6000067246 6000067247 6000067252 6000067256 6000067269 6000067272 6000067278 6000067293 6000067300 6000067309 6000067311 6000067312 6000067314 6000067345 6000067356 6000067357 6000067360 6000067366 6000067367 6000067418 6000067443 6000067446 6000067472 6000067475 6000067482 6000067484 6000067486 6000067496 6000067515 6000067552 6000067563 6000067584 6000067609 6000067625 6000067626 6000067629 6000067642 6000067658 6000067660 6000067662 6000067663 6000067673 6000067679 6000067680 6000067686 6000067689 6000067695 6000067701 6000067702 6000067710 6000067711 6000067718 6000067721 6000067729 6000067730 6000067732 6000067736 6000067751 6000067752 6000067757 6000067760 6000067764 6000067780 6000067783 6000067790 6000067807 6000067824 6000067832 6000067833 6000067838 6000067839 6000067849 6000067853 6000067864 6000067902 6000067906 6000067921 6000067922 6000067925 6000067926 6000067944 6000067959 6000067963 6000067993 6000068018 6000068042 6000068074 6000068120 6000068140 6000068226 6000068259 6000068262 6000068273 6000068274 6000068295 6000068301 6000068307 6000068311 6000068325 6000068327 6000068332 6000068333 6000068385 6000068386 6000068390 6000068391 6000068404 6000068406 6000068422 6000068437 6000068438 6000068440 6000068451 6000068452 6000068477 6000068489 6000068493 6000068494 6000068498 6000068503 6000068516 6000068522 6000068524 6000068569 6000068582 6000068592 6000068593 6000068638 6000068644 6000068648 6000068691 6000068725 6000068733 6000068736 6000068743 6000068749 6000068768 6000068785 6000068830 6000068835 6000068836 6000068838 6000068840 6000068841 6000068842 6000068843 6000068844 6000068845 6000068846 6000068849 6000068850 6000068851 6000068852 6000068857 6000068862 6000068867 6000068871 6000068874 6000068875 6000068877 6000068878 6000068880 6000068882 6000068888 6000068890 6000068892 6000068893 6000068896 6000068898 6000068900 6000068902 6000068903 6000068904 6000069017 6000069018 6000069020 6000069023 6000069029 6000069031 6000069037 6000069050 6000069056 6000069094 6000069105 6000069106 6000069109 6000069112 6000069113 6000069114 6000069115 6000069117 6000069119 6000069120 6000069121 6000069122 6000069126 6000069134 6000069153 6000069159 6000069160 6000069162 6000069164 6000069165 6000069173 6000069176 6000069177 6000069178 6000069181 6000069182 6000069185 6000069186 6000069187 6000069189 6000069192 6000069193 6000069194 6000069208 6000069211 6000069218 6000069219 6000069222 6000069249 6000069252 6000069253 6000069255 6000069262 6000069272 6000069273 6000069284 6000069294 6000069302 6000069304 6000069305 6000069306 6000069308 6000069309 6000069310 6000069312 6000069313 6000069314 6000069315 6000069316 6000069317 6000069318 6000069319 6000069320 6000069321 6000069322 6000069325 6000069327 6000069329 6000069330 6000069332 6000069335 6000069337 6000069338 6000069341 6000069348 6000069357 6000069358 6000069360 6000069361 6000069362 6000069364 6000069365 6000069366 6000069368 6000069371 6000069380 6000069395 6000069408 6000069409 6000069429 6000069447 6000069449 6000069450 6000069452 6000069457 6000069458 6000069462 6000069464 6000069465 6000069467 6000069468 6000069469 6000069470 6000069471 6000069472 6000069473 6000069475 6000069476 6000069477 6000069480 6000069481 6000069482 6000069528 6000069532 6000069536 6000069539 6000069556 6000069558 6000069561 6000069566 6000069568 6000069570 6000069575 6000069578 6000069585 6000069594 6000069612 6000069614 6000069625 6000069631 6000069655 6000069664 6000069687 6000069696 6000069711 6000069725 6000069727 6000069753 6000069761 6000069763 6000069768 6000069771 6000069772 6000069774 6000069781 6000069836 6000069844 6000069847 6000069850 6000069852 6000069856 6000069858 6000069862 6000069863 6000069864 6000069866 6000069867 6000069869 6000069871 6000069873 6000069874 6000069875 6000069876 6000069877 6000069881 6000069891 6000069898 6000069903 6000069904 6000069915 6000069916 6000069917 6000069918 6000069920 6000069922 6000069923 6000069924 6000069925 6000069927 6000069928 6000069929 6000069931 6000069932 6000069933 6000069934 6000069946 6000069951 6000069957 6000069958 6000069959 6000069961 6000069963 6000069964 6000069966 6000069967 6000069978 6000069979 6000069981 6000069982 6000070004 6000070018 6000070021 6000070024 6000070034 6000070048 6000070056 6000070081 6000070095 6000070105 6000070108 6000070109 6000070113 6000070115 6000070116 6000070117 6000070118 6000070120 6000070138 6000070142 6000070152 6000070162 6000070168 6000070185 6000070195 6000070199 6000070229 6000070234 6000070240 6000070248 6000070249 6000070252 6000070280 6000070346 6000070377 6000070380 6000070436 6000070437 6000070440 6000070453 6000070503 6000070621 6000070638 6000070681 6000070740 6000070804 6000070811 6000070831 6000070860 6000070867 6000070889 6000070890 6000070904 6000070911 6000070912 6000070916 6000070920 6000070921 6000070925 6000070928 6000070939 6000070957 6000070969 6000070982 6000071006 6000071037 6000071041 6000071042 6000071043 6000071044 6000071045 6000071046 6000071049 6000071050 6000071052 6000071054 6000071055 6000071056 6000071058 6000071059 6000071060 6000071062 6000071063 6000071064 6000071066 6000071068 6000071070 6000071071 6000071072 6000071073 6000071075 6000071127 6000071128 6000071129 6000071131 6000071132 6000071133 6000071135 6000071136 6000071140 6000071142 6000071144 6000071145 6000071148 6000071149 6000071151 6000071152 6000071153 6000071154 6000071156 6000071157 6000071160 6000071181 6000071196 6000071223 6000071225 6000071235 6000071238 6000071244 6000071246 6000071247 6000071249 6000071250 6000071251 6000071253 6000071257 6000071259 6000071260 6000071261 6000071262 6000071273 6000071275 6000071276 6000071279 6000071280 6000071283 6000071288 6000071291 6000071295 6000071296 6000071302 6000071304 6000071328 6000071350 6000071369 6000071381 6000071382 6000071386 6000071387 6000071391 6000071398 6000071400 6000071405 6000071408 6000071410 6000071411 6000071412 6000071413 6000071415 6000071418 6000071419 6000071429 6000071436 6000071438 6000071442 6000071445 6000071456 6000071458 6000071460 6000071467 6000071474 6000071476 6000071489 6000071505 6000071508 6000071509 6000071513 6000071515 6000071520 6000071523 6000071527 6000071531 6000071546 6000071548 6000071557 6000071561 6000071566 6000071570 6000071571 6000071572 6000071574 6000071578 6000071580 6000071584 6000071598 6000071599 6000071604 6000071606 6000071613 6000071651 6000071665 6000071671 6000071672 6000071676 6000071677 6000071678 6000071681 6000071682 6000071692 6000071693 6000071698 6000071703 6000071705 6000071708 6000071710 6000071711 6000071712 6000071713 6000071715 6000071716 6000071717 6000071719 6000071742 6000071749 6000071753 6000071754 6000071756 6000071757 6000071765 6000071766 6000071769 6000071771 6000071778 6000071785 6000071787 6000071792 6000071793 6000071799 6000071800 6000071804 6000071806 6000071808 6000071809 6000071810 6000071812 6000071815 6000071819 6000071820 6000071822 6000071823 6000071824 6000071827 6000071830 6000071832 6000071833 6000071836 6000071842 6000071844 6000071852 6000071856 6000071858 6000071862 6000071863 6000071867 6000071870 6000071875 6000071876 6000071884 6000071888 6000071890 6000071891 6000071901 6000071905 6000071908 6000071909 6000071910 6000071911 6000071913 6000071914 6000071915 6000071917 6000071922 6000071928 6000071931 6000071932 6000071934 6000071940 6000071944 6000071945 6000071946 6000071950 6000071970 6000071974 6000071975 6000071977 6000071980 6000071981 6000071984 6000071987 6000071988 6000071996 6000071997 6000071998 6000072002 6000072011 6000072014 6000072015 6000072016 6000072017 6000072019 6000072021 6000072024 6000072025 6000072027 6000072034 6000072049 6000072060 6000072063 6000072070 6000072074 6000072083 6000072085 6000072090 6000072092 6000072100 6000072101 6000072102 6000072106 6000072112 6000072114 6000072116 6000072118 6000072121 6000072122 6000072126 6000072130 6000072131 6000072134 6000072136 6000072139 6000072143 6000072144 6000072146 6000072147 6000072151 6000072154 6000072160 6000072163 6000072164 6000072166 6000072167 6000072172 6000072173 6000072175 6000072176 6000072177 6000072178 6000072179 6000072180 6000072181 6000072182 6000072184 6000072187 6000072197 6000072200 6000072207 6000072210 6000072212 6000072213 6000072215 6000072216 6000072217 6000072218 6000072220 6000072223 6000072224 6000072225 6000072227 6000072231 6000072234 6000072235 6000072237 6000072238 6000072239 6000072242 6000072244 6000072245 6000072246 6000072247 6000072248 6000072249 6000072252 6000072254 6000072256 6000072259 6000072262 6000072263 6000072264 6000072265 6000072268 6000072271 6000072272 6000072273 6000072274 6000072275 6000072280 6000072282 6000072283 6000072284 6000072285 6000072290 6000072291 6000072292 6000072294 6000072298 6000072299 6000072302 6000072303 6000072306 6000072310 6000072311 6000072313 6000072314 6000072315 6000072316 6000072321 6000072322 6000072323 6000072326 6000072327 6000072329 6000072330 6000072333 6000072335 6000072338 6000072339 6000072341 6000072345 6000072349 6000072352 6000072353 6000072354 6000072356 6000072357 6000072359 6000072369 6000072371 6000072373 6000072375 6000072378 6000072380 6000072389 6000072395 6000072398 6000072400 6000072401 6000072402 6000072403 6000072404 6000072406 6000072412 6000072416 6000072417 6000072420 6000072422 6000072431 6000072433 6000072437 6000072438 6000072446 6000072461 6000072465 6000072467 6000072468 6000072471 6000072473 6000072478 6000072479 6000072484 6000072485 6000072486 6000072492 6000072495 6000072520 6000072524 6000072529 6000072533 6000072538 6000072539 6000072544 6000072546 6000072547 6000072560 6000072569 6000072570 6000072582 6000072584 6000072588 6000072589 6000072590 6000072595 6000072596 6000072598 6000072600 6000072601 6000072602 6000072603 6000072604 6000072605 6000072607 6000072608 6000072611 6000072613 6000072614 6000072616 6000072619 6000072623 6000072625 6000072626 6000072627 6000072630 6000072631 6000072632 6000072633 6000072639 6000072644 6000072646 6000072655 6000072670 6000072671 6000072672 6000072673 6000072676 6000072677 6000072680 6000072681 6000072690 6000072693 6000072699 6000072706 6000072707 6000072718 6000072719 6000072721 6000072722 6000072724 6000072726 6000072731 6000072734 6000072735 6000072739 6000072742 6000072743 6000072745 6000072746 6000072749 6000072750 6000072751 6000072757 6000072758 6000072759 6000072761 6000072762 6000072763 6000072766 6000072771 6000072777 6000072782 6000072786 6000072803 6000072804 6000072809 6000072824 6000072825 6000072827 6000072829 6000072830 6000072831 6000072832 6000072834 6000072841 6000072843 6000072851 6000072854 6000072862 6000072863 6000072866 6000072867 6000072868 6000072869 6000072870 6000072883 6000072888 6000072895 6000072898 6000072899 6000072910 6000072916 6000072917 6000072921 6000072923 6000072933 6000072938 6000072943 6000072948 6000072958 6000072960 6000072963 6000072972 6000072987 6000072988 6000072992 6000072996 6000072999 6000073003 6000073008 6000073009 6000073014 6000073016 6000073021 6000073022 6000073028 6000073030 6000073038 6000073041 6000073047 6000073048 6000073049 6000073050 6000073054 6000073055 6000073063 6000073065 6000073068 6000073073 6000073080 6000073081 6000073083 6000073085 6000073087 6000073090 6000073092 6000073094 6000073095 6000073099 6000073100 6000073103 6000073104 6000073106 6000073112 6000073116 6000073117 6000073118 6000073125 6000073135 6000073136 6000073141 6000073149 6000073154 6000073160 6000073162 6000073163 6000073183 6000073184 6000073200 6000073209 6000073214 6000073216 6000073224 6000073227 6000073232 6000073238 6000073240 6000073262 6000073263 6000073264 6000073270 6000073274 6000073275 6000073276 6000073277 6000073279 6000073291 6000073293 6000073295 6000073296 6000073300 6000073304 6000073306 6000073309 6000073313 6000073314 6000073315 6000073325 6000073330 6000073332 6000073333 6000073345 6000073351 6000073354 6000073363 6000073376 6000073380 6000073387 6000073389 6000073390 6000073391 6000073392 6000073395 6000073401 6000073405 6000073408 6000073410 6000073413 6000073420 6000073423 6000073433 6000073436 6000073437 6000073439 6000073440 6000073444 6000073446 6000073447 6000073452 6000073464 6000073476 6000073481 6000073485 6000073490 6000073506 6000073517 6000073520 6000073523 6000073524 6000073529 6000073535 6000073536 6000073541 6000073544 6000073545 6000073550 6000073557 6000073562 6000073567 6000073568 6000073585 6000073589 6000073599 6000073601 6000073608 6000073614 6000073617 6000073630 6000073631 6000073632 6000073634 6000073635 6000073639 6000073643 6000073646 6000073648 6000073649 6000073661 6000073667 6000073669 6000073675 6000073684 6000073698 6000073707 6000073712 6000073721 6000073722 6000073723 6000073726 6000073728 6000073730 6000073752 6000073758 6000073763 6000073764 6000073770 6000073771 6000073778 6000073780 6000073782 6000073789 6000073794 6000073796 6000073802 6000073807 6000073821 6000073826 6000073829 6000073844 6000073847 6000073854 6000073859 6000073860 6000073861 6000073863 6000073868 6000073874 6000073882 6000073887 6000073899 6000073901 6000073912 6000073913 6000073951 6000073977 6000073982 6000073997 6000074001 6000074011 6000074013 6000074015 6000074016 6000074018 6000074020 6000074025 6000074027 6000074030 6000074034 6000074040 6000074042 6000074046 6000074047 6000074048 6000074049 6000074056 6000074069 6000074074 6000074076 6000074094 6000074095 6000074101 6000074109 6000074114 6000074116 6000074117 6000074118 6000074119 6000074121 6000074122 6000074123 6000074124 6000074125 6000074129 6000074130 6000074131 6000074133 6000074134 6000074136 6000074138 6000074144 6000074145 6000074146 6000074150 6000074151 6000074154 6000074155 6000074158 6000074159 6000074161 6000074162 6000074163 6000074164 6000074165 6000074168 6000074172 6000074173 6000074176 6000074177 6000074178 6000074179 6000074180 6000074181 6000074182 6000074185 6000074188 6000074190 6000074193 6000074194 6000074196 6000074201 6000074202 6000074204 6000074209 6000074212 6000074214 6000074215 6000074217 6000074222 6000074224 6000074226 6000074228 6000074229 6000074230 6000074231 6000074232 6000074236 6000074243 6000074245 6000074251 6000074252 6000074259 6000074261 6000074262 6000074263 6000074264 6000074266 6000074267 6000074268 6000074272 6000074273 6000074275 6000074276 6000074280 6000074289 6000074290 6000074298 6000074301 6000074304 6000074305 6000074307 6000074309 6000074313 6000074320 6000074321 6000074324 6000074325 6000074326 6000074327 6000074329 6000074333 6000074335 6000074336 6000074338 6000074339 6000074341 6000074343 6000074344 6000074345 6000074348 6000074349 6000074351 6000074353 6000074355 6000074358 6000074360 6000074361 6000074363 6000074364 6000074367 6000074370 6000074371 6000074376 6000074378 6000074381 6000074383 6000074388 6000074389 6000074394 6000074397 6000074400 6000074403 6000074408 6000074409 6000074410 6000074413 6000074416 6000074418 6000074419 6000074421 6000074422 6000074423 6000074432 6000074433 6000074435 6000074436 6000074438 6000074441 6000074442 6000074443 6000074445 6000074446 6000074447 6000074449 6000074453 6000074456 6000074459 6000074461 6000074463 6000074464 6000074465 6000074466 6000074468 6000074469 6000074470 6000074471 6000074472 6000074474 6000074475 6000074476 6000074478 6000074480 6000074482 6000074484 6000074485 6000074487 6000074488 6000074493 6000074501 6000074503 6000074504 6000074512 6000074513 6000074515 6000074516 6000074517 6000074519 6000074522 6000074523 6000074527 6000074529 6000074530 6000074536 6000074545 6000074547 6000074549 6000074550 6000074553 6000074558 6000074560 6000074561 6000074562 6000074563 6000074567 6000074568 6000074569 6000074570 6000074571 6000074572 6000074574 6000074576 6000074581 6000074587 6000074593 6000074597 6000074603 6000074607 6000074609 6000074610 6000074613 6000074618 6000074619 6000074620 6000074625 6000074627 6000074628 6000074630 6000074632 6000074633 6000074635 6000074636 6000074637 6000074639 6000074677 6000074678 6000074679 6000074682 6000074685 6000074689 6000074691 6000074693 6000074694 6000074698 6000074706 6000074707 6000074710 6000074712 6000074714 6000074720 6000074723 6000074728 6000074748 6000074750 6000074759 6000074763 6000074765 6000074767 6000074768 6000074773 6000074774 6000074780 6000074786 6000074787 6000074791 6000074792 6000074794 6000074795 6000074801 6000074803 6000074806 6000074807 6000074808 6000074809 6000074811 6000074813 6000074819 6000074820 6000074886 6000074894 6000074896 6000074900 6000074902 6000074905 6000074908 6000074915 6000074918 6000074935 6000074945 6000074948 6000074949 6000074950 6000074955 6000074961 6000074970 6000074971 6000074974 6000074975 6000074980 6000074981 6000074985 6000074986 6000074992 6000074993 6000074995 6000075003 6000075005 6000075011 6000075012 6000075013 6000075015 6000075017 6000075018 6000075019 6000075020 6000075025 6000075028 6000075030 6000075031 6000075033 6000075034 6000075035 6000075036 6000075038 6000075040 6000075046 6000075049 6000075050 6000075052 6000075054 6000075058 6000075060 6000075062 6000075065 6000075066 6000075068 6000075069 6000075072 6000075074 6000075075 6000075076 6000075077 6000075080 6000075083 6000075085 6000075090 6000075093 6000075097 6000075099 6000075102 6000075103 6000075104 6000075108 6000075109 6000075111 6000075115 6000075116 6000075117 6000075118 6000075119 6000075120 6000075121 6000075123 6000075125 6000075126 6000075127 6000075130 6000075131 6000075146 6000075147 6000075149 6000075150 6000075152 6000075153 6000075155 6000075156 6000075158 6000075160 6000075162 6000075163 6000075164 6000075167 6000075170 6000075171 6000075173 6000075174 6000075175 6000075176 6000075178 6000075179 6000075180 6000075181 6000075183 6000075187 6000075190 6000075191 6000075193 6000075195 6000075199 6000075200 6000075202 6000075204 6000075205 6000075206 6000075209 6000075210 6000075212 6000075213 6000075215 6000075216 6000075220 6000075222 6000075225 6000075226 6000075229 6000075230 6000075231 6000075232 6000075233 6000075235 6000075236 6000075237 6000075238 6000075239 6000075247 6000075250 6000075252 6000075254 6000075256 6000075257 6000075258 6000075261 6000075264 6000075265 6000075267 6000075271 6000075272 6000075274 6000075277 6000075278 6000075279 6000075280 6000075282 6000075284 6000075285 6000075286 6000075288 6000075292 6000075293 6000075294 6000075296 6000075298 6000075300 6000075303 6000075304 6000075305 6000075307 6000075308 6000075310 6000075311 6000075312 6000075313 6000075314 6000075315 6000075317 6000075318 6000075321 6000075322 6000075323 6000075325 6000075326 6000075329 6000075330 6000075331 6000075333 6000075335 6000075337 6000075339 6000075344 6000075348 6000075352 6000075354 6000075355 6000075356 6000075361 6000075362 6000075363 6000075364 6000075366 6000075367 6000075371 6000075372 6000075375 6000075376 6000075377 6000075399 6000075402 6000075403 6000075404 6000075407 6000075416 6000075418 6000075419 6000075420 6000075421 6000075430 6000075438 6000075442 6000075443 6000075445 6000075448 6000075454 6000075455 6000075458 6000075459 6000075466 6000075472 6000075475 6000075476 6000075477 6000075478 6000075481 6000075484 6000075486 6000075491 6000075492 6000075494 6000075496 6000075498 6000075502 6000075504 6000075508 6000075513 6000075518 6000075520 6000075522 6000075527 6000075528 6000075534 6000075535 6000075538 6000075543 6000075544 6000075545 6000075549 6000075554 6000075555 6000075556 6000075557 6000075559 6000075561 6000075563 6000075564 6000075569 6000075570 6000075571 6000075572 6000075574 6000075576 6000075582 6000075583 6000075585 6000075586 6000075589 6000075591 6000075592 6000075594 6000075595 6000075597 6000075598 6000075600 6000075601 6000075602 6000075603 6000075605 6000075606 6000075607 6000075610 6000075611 6000075612 6000075613 6000075614 6000075615 6000075616 6000075620 6000075622 6000075623 6000075625 6000075626 6000075628 6000075630 6000075633 6000075634 6000075636 6000075641 6000075642 6000075644 6000075647 6000075651 6000075652 6000075655 6000075660 6000075665 6000075666 6000075667 6000075669 6000075677 6000075679 6000075680 6000075685 6000075686 6000075688 6000075691 6000075694 6000075698 6000075701 6000075704 6000075705 6000075706 6000075708 6000075712 6000075716 6000075719 6000075722 6000075723 6000075724 6000075729 6000075731 6000075739 6000075741 6000075742 6000075743 6000075745 6000075746 6000075748 6000075755 6000075756 6000075759 6000075765 6000075766 6000075771 6000075775 6000075788 6000075794 6000075797 6000075798 6000075800 6000075801 6000075807 6000075810 6000075812 6000075815 6000075816 6000075817 6000075818 6000075819 6000075821 6000075822 6000075823 6000075828 6000075831 6000075835 6000075836 6000075839 6000075840 6000075843 6000075845 6000075849 6000075854 6000075858 6000075860 6000075861 6000075862 6000075865 6000075867 6000075868 6000075869 6000075870 6000075872 6000075874 6000075875 6000075878 6000075884 6000075886 6000075887 6000075888 6000075890 6000075891 6000075896 6000075898 6000075901 6000075904 6000075905 6000075906 6000075907 6000075908 6000075909 6000075911 6000075913 6000075916 6000075917 6000075918 6000075922 6000075923 6000075924 6000075927 6000075931 6000075933 6000075934 6000075939 6000075940 6000075941 6000075943 6000075946 6000075947 6000075948 6000075949 6000075950 6000075967 6000075968 6000075985 6000075986 6000075988 6000075989 6000075990 6000075991 6000075992 6000075993 6000075996 6000075997 6000075998 6000075999 6000076004 6000076008 6000076009 6000076011 6000076014 6000076015 6000076016 6000076017 6000076020 6000076022 6000076023 6000076024 6000076025 6000076027 6000076032 6000076035 6000076038 6000076045 6000076049 6000076058 6000076059 6000076061 6000076062 6000076071 6000076072 6000076073 6000076076 6000076077 6000076083 6000076090 6000076091 6000076092 6000076095 6000076098 6000076101 6000076102 6000076103 6000076105 6000076107 6000076108 6000076110 6000076112 6000076113 6000076124 6000076127 6000076131 6000076132 6000076137 6000076138 6000076143 6000076144 6000076147 6000076148 6000076149 6000076151 6000076152 6000076153 6000076154 6000076155 6000076156 6000076158 6000076161 6000076165 6000076168 6000076171 6000076173 6000076175 6000076178 6000076182 6000076183 6000076191 6000076192 6000076194 6000076195 6000076199 6000076200 6000076201 6000076204 6000076205 6000076208 6000076209 6000076210 6000076211 6000076212 6000076213 6000076215 6000076216 6000076217 6000076218 6000076221 6000076223 6000076224 6000076226 6000076233 6000076234 6000076237 6000076239 6000076240 6000076241 6000076252 6000076256 6000076265 6000076269 6000076272 6000076276 6000076304 6000076309 6000076313 6000076314 6000076317 6000076318 6000076320 6000076321 6000076323 6000076324 6000076326 6000076327 6000076328 6000076330 6000076332 6000076335 6000076336 6000076339 6000076345 6000076353 6000076354 6000076356 6000076360 6000076362 6000076369 6000076373 6000076379 6000076383 6000076384 6000076385 6000076416 6000076442 6000076443 6000076448 6000076461 6000076472 6000076475 6000076479 6000076482 6000076489 6000076499 6000076500 6000076503 6000076504 6000076508 6000076509 6000076537 6000076543 6000076548 6000076607 6000076609 6000076611 6000076615 6000076621 6000076622 6000076624 6000076648 6000076673 6000076677 6000076678 6000076685 6000076686 6000076688 6000076695 6000076697 6000076698 6000076699 6000076701 6000076710 6000076711 6000076717 6000076718 6000076721 6000076722 6000076727 6000076730 6000076731 6000076732 6000076733 6000076735 6000076736 6000076737 6000076739 6000076743 6000076745 6000076746 6000076747 6000076758 6000076759 6000076769 6000076772 6000076775 6000076776 6000076777 6000076781 6000076785 6000076788 6000076789 6000076790 6000076791 6000076792 6000076795 6000076796 6000076797 6000076798 6000076799 6000076803 6000076806 6000076808 6000076811 6000076813 6000076814 6000076815 6000076818 6000076823 6000076825 6000076827 6000076830 6000076834 6000076835 6000076837 6000076839 6000076840 6000076842 6000076844 6000076846 6000076847 6000076853 6000076855 6000076860 6000076862 6000076864 6000076873 6000076877 6000076878 6000076880 6000076883 6000076887 6000076889 6000076892 6000076895 6000076896 6000076897 6000076903 6000076904 6000076911 6000076912 6000076913 6000076914 6000076918 6000076920 6000076922 6000076927 6000076928 6000076930 6000076932 6000076933 6000076935 6000076937 6000076938 6000076940 6000076942 6000076945 6000076947 6000076948 6000076949 6000076952 6000076954 6000076956 6000076959 6000076964 6000076971 6000076973 6000076974 6000076979 6000076981 6000076984 6000076985 6000076986 6000076992 6000076994 6000076996 6000076997 6000076998 6000077003 6000077014 6000077016 6000077021 6000077022 6000077024 6000077025 6000077029 6000077033 6000077034 6000077035 6000077036 6000077040 6000077041 6000077043 6000077045 6000077047 6000077050 6000077055 6000077057 6000077060 6000077062 6000077064 6000077065 6000077067 6000077068 6000077069 6000077073 6000077075 6000077078 6000077081 6000077085 6000077086 6000077091 6000077093 6000077094 6000077095 6000077096 6000077097 6000077098 6000077102 6000077104 6000077109 6000077110 6000077112 6000077113 6000077114 6000077115 6000077116 6000077117 6000077119 6000077121 6000077123 6000077125 6000077133 6000077140 6000077142 6000077144 6000077147 6000077148 6000077149 6000077150 6000077151 6000077153 6000077155 6000077156 6000077160 6000077163 6000077165 6000077173 6000077175 6000077176 6000077178 6000077180 6000077181 6000077182 6000077188 6000077191 6000077193 6000077196 6000077199 6000077202 6000077204 6000077206 6000077211 6000077215 6000077220 6000077221 6000077222 6000077224 6000077226 6000077228 6000077230 6000077232 6000077233 6000077238 6000077240 6000077245 6000077246 6000077247 6000077254 6000077260 6000077263 6000077268 6000077269 6000077272 6000077273 6000077274 6000077276 6000077277 6000077278 6000077279 6000077283 6000077286 6000077287 6000077288 6000077290 6000077293 6000077297 6000077303 6000077307 6000077309 6000077311 6000077312 6000077314 6000077316 6000077318 6000077319 6000077328 6000077329 6000077330 6000077332 6000077333 6000077334 6000077335 6000077336 6000077338 6000077342 6000077345 6000077346 6000077347 6000077356 6000077357 6000077359 6000077367 6000077369 6000077375 6000077377 6000077379 6000077385 6000077394 6000077396 6000077398 6000077399 6000077406 6000077410 6000077411 6000077413 6000077418 6000077420 6000077425 6000077429 6000077430 6000077431 6000077435 6000077436 6000077437 6000077438 6000077446 6000077447 6000077448 6000077452 6000077454 6000077455 6000077457 6000077458 6000077459 6000077462 6000077463 6000077464 6000077465 6000077470 6000077472 6000077474 6000077479 6000077483 6000077486 6000077487 6000077491 6000077492 6000077498 6000077499 6000077501 6000077503 6000077504 6000077505 6000077506 6000077509 6000077510 6000077511 6000077512 6000077514 6000077515 6000077517 6000077520 6000077521 6000077529 6000077530 6000077531 6000077532 6000077534 6000077535 6000077537 6000077540 6000077546 6000077548 6000077550 6000077554 6000077556 6000077565 6000077566 6000077567 6000077569 6000077570 6000077571 6000077574 6000077577 6000077582 6000077583 6000077585 6000077589 6000077592 6000077593 6000077594 6000077598 6000077602 6000077611 6000077613 6000077614 6000077621 6000077624 6000077626 6000077627 6000077630 6000077634 6000077635 6000077640 6000077645 6000077654 6000077657 6000077663 6000077664 6000077667 6000077670 6000077677 6000077679 6000077681 6000077683 6000077684 6000077687 6000077689 6000077692 6000077694 6000077705 6000077710 6000077711 6000077717 6000077722 6000077729 6000077730 6000077731 6000077733 6000077735 6000077736 6000077737 6000077738 6000077740 6000077742 6000077744 6000077746 6000077747 6000077751 6000077753 6000077754 6000077755 6000077756 6000077759 6000077760 6000077763 6000077764 6000077765 6000077766 6000077767 6000077768 6000077772 6000077773 6000077774 6000077778 6000077780 6000077782 6000077784 6000077786 6000077788 6000077793 6000077794 6000077795 6000077797 6000077799 6000077802 6000077803 6000077804 6000077805 6000077807 6000077808 6000077809 6000077810 6000077811 6000077812 6000077813 6000077814 6000077815 6000077816 6000077819 6000077823 6000077825 6000077828 6000077829 6000077831 6000077832 6000077833 6000077834 6000077835 6000077836 6000077837 6000077849 6000077850 6000077855 6000077856 6000077858 6000077860 6000077867 6000077868 6000077873 6000077875 6000077876 6000077879 6000077880 6000077882 6000077883 6000077888 6000077889 6000077892 6000077898 6000077899 6000077902 6000077903 6000077905 6000077908 6000077910 6000077914 6000077917 6000077925 6000077928 6000077931 6000077933 6000077935 6000077940 6000077941 6000077943 6000077946 6000077948 6000077950 6000077951 6000077956 6000077959 6000077961 6000077962 6000077964 6000077965 6000077966 6000077967 6000077970 6000077971 6000077985 6000077986 6000077988 6000077994 6000077997 6000077998 6000078000 6000078001 6000078002 6000078003 6000078008 6000078010 6000078013 6000078020 6000078021 6000078024 6000078027 6000078029 6000078032 6000078037 6000078039 6000078040 6000078044 6000078045 6000078049 6000078050 6000078051 6000078055 6000078056 6000078059 6000078060 6000078064 6000078065 6000078066 6000078067 6000078070 6000078073 6000078074 6000078077 6000078080 6000078081 6000078085 6000078086 6000078089 6000078090 6000078091 6000078092 6000078094 6000078095 6000078098 6000078100 6000078104 6000078108 6000078109 6000078111 6000078113 6000078128 6000078136 6000078141 6000078155 6000078162 6000078164 6000078166 6000078176 6000078184 6000078193 6000078204 6000078206 6000078212 6000078216 6000078225 6000078231 6000078232 6000078235 6000078236 6000078240 6000078247 6000078252 6000078253 6000078257 6000078258 6000078260 6000078279 6000078303 6000078304 6000078313 6000078316 6000078320 6000078323 6000078326 6000078328 6000078341 6000078346 6000078349 6000078362 6000078363 6000078367 6000078384 6000078390 6000078391 6000078392 6000078395 6000078396 6000078411 6000078417 6000078418 6000078422 6000078434 6000078448 6000078458 6000078475 6000078491 6000078492 6000078503 6000078515 6000078531 6000078534 6000078536 6000078538 6000078542 6000078545 6000078548 6000078551 6000078553 6000078554 6000078556 6000078559 6000078560 6000078561 6000078562 6000078563 6000078565 6000078566 6000078568 6000078570 6000078572 6000078573 6000078574 6000078575 6000078578 6000078585 6000078586 6000078595 6000078596 6000078599 6000078600 6000078601 6000078602 6000078604 6000078606 6000078609 6000078610 6000078612 6000078617 6000078618 6000078619 6000078621 6000078622 6000078623 6000078625 6000078626 6000078627 6000078628 6000078629 6000078631 6000078632 6000078634 6000078636 6000078640 6000078641 6000078642 6000078643 6000078644 6000078646 6000078647 6000078651 6000078653 6000078655 6000078658 6000078660 6000078666 6000078667 6000078669 6000078670 6000078673 6000078674 6000078675 6000078695 6000078699 6000078700 6000078705 6000078711 6000078715 6000078718 6000078724 6000078731 6000078735 6000078736 6000078737 6000078740 6000078741 6000078750 6000078756 6000078760 6000078767 6000078769 6000078770 6000078773 6000078774 6000078775 6000078778 6000078779 6000078781 6000078783 6000078784 6000078819 6000078833 6000078837 6000078840 6000078842 6000078844 6000078851 6000078852 6000078854 6000078857 6000078858 6000078876 6000078883 6000078884 6000078887 6000078888 6000078889 6000078894 6000078897 6000078903 6000078906 6000078907 6000078915 6000078920 6000078930 6000078931 6000078934 6000078935 6000078939 6000078940 6000078941 6000078945 6000078946 6000078948 6000078958 6000078968 6000078970 6000078973 6000078974 6000078977 6000078980 6000078982 6000078999 6000079005 6000079006 6000079007 6000079012 6000079014 6000079015 6000079016 6000079017 6000079020 6000079023 6000079024 6000079031 6000079039 6000079040 6000079046 6000079047 6000079049 6000079051 6000079052 6000079055 6000079056 6000079057 6000079058 6000079064 6000079066 6000079067 6000079068 6000079069 6000079075 6000079078 6000079080 6000079083 6000079086 6000079092 6000079098 6000079101 6000079102 6000079104 6000079112 6000079113 6000079115 6000079116 6000079118 6000079119 6000079123 6000079124 6000079126 6000079129 6000079130 6000079147 6000079150 6000079152 6000079153 6000079154 6000079155 6000079156 6000079157 6000079162 6000079163 6000079164 6000079167 6000079168 6000079170 6000079171 6000079172 6000079173 6000079178 6000079179 6000079181 6000079182 6000079183 6000079186 6000079187 6000079190 6000079193 6000079194 6000079195 6000079196 6000079200 6000079202 6000079203 6000079206 6000079207 6000079210 6000079211 6000079215 6000079217 6000079218 6000079219 6000079220 6000079221 6000079222 6000079230 6000079231 6000079233 6000079234 6000079235 6000079236 6000079239 6000079240 6000079241 6000079242 6000079244 6000079245 6000079247 6000079248 6000079249 6000079250 6000079251 6000079252 6000079257 6000079260 6000079261 6000079265 6000079274 6000079276 6000079290 6000079295 6000079301 6000079302 6000079306 6000079307 6000079313 6000079314 6000079319 6000079325 6000079328 6000079332 6000079333 6000079335 6000079336 6000079349 6000079354 6000079355 6000079357 6000079359 6000079360 6000079363 6000079379 6000079381 6000079385 6000079388 6000079389 6000079392 6000079395 6000079397 6000079400 6000079402 6000079403 6000079405 6000079407 6000079409 6000079413 6000079416 6000079417 6000079420 6000079425 6000079427 6000079428 6000079434 6000079440 6000079446 6000079447 6000079456 6000079465 6000079467 6000079475 6000079476 6000079484 6000079487 6000079492 6000079495 6000079497 6000079500 6000079502 6000079508 6000079515 6000079520 6000079526 6000079527 6000079530 6000079535 6000079540 6000079543 6000079553 6000079573 6000079578 6000079581 6000079582 6000079583 6000079590 6000079594 6000079595 6000079604 6000079616 6000079618 6000079621 6000079622 6000079625 6000079626 6000079628 6000079634 6000079635 6000079636 6000079639 6000079640 6000079648 6000079650 6000079653 6000079654 6000079656 6000079657 6000079659 6000079660 6000079662 6000079664 6000079667 6000079668 6000079670 6000079672 6000079677 6000079678 6000079679 6000079684 6000079686 6000079688 6000079693 6000079695 6000079699 6000079701 6000079706 6000079710 6000079717 6000079723 6000079725 6000079726 6000079728 6000079736 6000079746 6000079752 6000079753 6000079758 6000079760 6000079761 6000079762 6000079763 6000079764 6000079765 6000079770 6000079774 6000079781 6000079786 6000079790 6000079794 6000079796 6000079799 6000079801 6000079805 6000079807 6000079812 6000079813 6000079815 6000079825 6000079826 6000079833 6000079835 6000079836 6000079837 6000079841 6000079843 6000079845 6000079852 6000079858 6000079859 6000079863 6000079870 6000079881 6000079882 6000079886 6000079890 6000079893 6000079898 6000079904 6000079906 6000079912 6000079920 6000079922 6000079925 6000079937 6000079944 6000079948 6000079966 6000079974 6000079975 6000079979 6000079981 6000079983 6000080000 6000080013 6000080015 6000080017 6000080018 6000080019 6000080045 6000080047 6000080052 6000080053 6000080055 6000080058 6000080062 6000080065 6000080069 6000080073 6000080078 6000080080 6000080085 6000080086 6000080089 6000080092 6000080094 6000080096 6000080098 6000080110 6000080115 6000080116 6000080121 6000080124 6000080125 6000080127 6000080130 6000080136 6000080141 6000080142 6000080144 6000080147 6000080149 6000080151 6000080155 6000080158 6000080160 6000080161 6000080163 6000080164 6000080165 6000080166 6000080167 6000080168 6000080169 6000080170 6000080172 6000080173 6000080174 6000080176 6000080179 6000080182 6000080183 6000080184 6000080185 6000080188 6000080189 6000080191 6000080192 6000080194 6000080198 6000080199 6000080201 6000080205 6000080207 6000080208 6000080211 6000080212 6000080215 6000080216 6000080219 6000080220 6000080221 6000080222 6000080224 6000080225 6000080226 6000080231 6000080232 6000080233 6000080234 6000080235 6000080236 6000080237 6000080238 6000080243 6000080246 6000080249 6000080250 6000080253 6000080255 6000080258 6000080260 6000080261 6000080263 6000080264 6000080265 6000080267 6000080269 6000080273 6000080276 6000080277 6000080278 6000080279 6000080280 6000080283 6000080284 6000080285 6000080286 6000080288 6000080291 6000080292 6000080294 6000080295 6000080296 6000080297 6000080298 6000080303 6000080304 6000080305 6000080306 6000080309 6000080310 6000080311 6000080312 6000080313 6000080314 6000080319 6000080320 6000080321 6000080323 6000080324 6000080326 6000080327 6000080329 6000080330 6000080331 6000080332 6000080337 6000080338 6000080340 6000080341 6000080342 6000080345 6000080346 6000080348 6000080349 6000080351 6000080353 6000080357 6000080358 6000080359 6000080360 6000080361 6000080362 6000080366 6000080368 6000080371 6000080372 6000080373 6000080374 6000080375 6000080377 6000080378 6000080379 6000080380 6000080381 6000080386 6000080387 6000080388 6000080389 6000080391 6000080395 6000080396 6000080399 6000080400 6000080403 6000080404 6000080405 6000080409 6000080411 6000080412 6000080413 6000080414 6000080416 6000080419 6000080421 6000080425 6000080427 6000080428 6000080429 6000080430 6000080432 6000080433 6000080434 6000080435 6000080437 6000080438 6000080439 6000080441 6000080443 6000080446 6000080447 6000080448 6000080449 6000080450 6000080451 6000080452 6000080453 6000080454 6000080455 6000080456 6000080459 6000080463 6000080465 6000080466 6000080467 6000080468 6000080469 6000080470 6000080472 6000080473 6000080474 6000080476 6000080477 6000080478 6000080479 6000080481 6000080482 6000080483 6000080485 6000080486 6000080489 6000080498 6000080501 6000080502 6000080507 6000080508 6000080509 6000080511 6000080513 6000080514 6000080515 6000080516 6000080521 6000080525 6000080526 6000080529 6000080531 6000080536 6000080537 6000080539 6000080540 6000080541 6000080542 6000080543 6000080544 6000080545 6000080546 6000080547 6000080549 6000080552 6000080553 6000080554 6000080556 6000080558 6000080562 6000080563 6000080564 6000080566 6000080568 6000080572 6000080575 6000080577 6000080580 6000080581 6000080582 6000080583 6000080587 6000080588 6000080591 6000080593 6000080594 6000080598 6000080600 6000080601 6000080612 6000080614 6000080616 6000080617 6000080618 6000080622 6000080623 6000080624 6000080625 6000080626 6000080627 6000080631 6000080632 6000080635 6000080636 6000080637 6000080639 6000080640 6000080641 6000080642 6000080643 6000080644 6000080646 6000080647 6000080650 6000080651 6000080652 6000080654 6000080657 6000080659 6000080663 6000080666 6000080668 6000080670 6000080671 6000080673 6000080676 6000080677 6000080682 6000080687 6000080688 6000080690 6000080691 6000080694 6000080697 6000080699 6000080701 6000080704 6000080705 6000080706 6000080707 6000080709 6000080714 6000080718 6000080719 6000080720 6000080722 6000080727 6000080728 6000080729 6000080730 6000080732 6000080733 6000080734 6000080735 6000080736 6000080739 6000080740 6000080741 6000080743 6000080744 6000080745 6000080746 6000080749 6000080751 6000080754 6000080758 6000080759 6000080760 6000080761 6000080764 6000080766 6000080767 6000080768 6000080769 6000080771 6000080772 6000080777 6000080781 6000080783 6000080785 6000080786 6000080793 6000080797 6000080798 6000080799 6000080803 6000080807 6000080814 6000080820 6000080822 6000080823 6000080824 6000080825 6000080826 6000080827 6000080828 6000080837 6000080839 6000080841 6000080842 6000080843 6000080844 6000080846 6000080847 6000080848 6000080850 6000080851 6000080853 6000080855 6000080857 6000080863 6000080864 6000080866 6000080867 6000080868 6000080869 6000080871 6000080873 6000080874 6000080875 6000080878 6000080880 6000080881 6000080884 6000080885 6000080887 6000080888 6000080896 6000080898 6000080900 6000080901 6000080908 6000080909 6000080912 6000080913 6000080914 6000080915 6000080916 6000080917 6000080918 6000080919 6000080920 6000080922 6000080924 6000080925 6000080926 6000080930 6000080936 6000080940 6000080941 6000080946 6000080949 6000080953 6000080954 6000080955 6000080957 6000080958 6000080962 6000080965 6000080970 6000080972 6000080976 6000080977 6000080979 6000080986 6000080988 6000080989 6000080991 6000080995 6000080997 6000080998 6000081004 6000081007 6000081008 6000081011 6000081013 6000081014 6000081015 6000081017 6000081018 6000081019 6000081020 6000081021 6000081022 6000081023 6000081025 6000081026 6000081027 6000081028 6000081029 6000081030 6000081031 6000081032 6000081033 6000081034 6000081035 6000081037 6000081038 6000081039 6000081040 6000081041 6000081042 6000081043 6000081044 6000081046 6000081047 6000081049 6000081050 6000081051 6000081052 6000081053 6000081054 6000081055 6000081056 6000081058 6000081059 6000081060 6000081062 6000081063 6000081064 6000081066 6000081067 6000081068 6000081069 6000081071 6000081073 6000081074 6000081076 6000081077 6000081078 6000081080 6000081081 6000081082 6000081083 6000081085 6000081086 6000081087 6000081088 6000081089 6000081090 6000081091 6000081092 6000081093 6000081094 6000081095 6000081096 6000081097 6000081098 6000081099 6000081100 6000081101 6000081102 6000081103 6000081105 6000081106 6000081107 6000081108 6000081109 6000081110 6000081111 6000081113 6000081114 6000081115 6000081116 6000081117 6000081118 6000081120 6000081121 6000081122 6000081123 6000081124 6000081125 6000081126 6000081127 6000081128 6000081129 6000081130 6000081131 6000081133 6000081134 6000081135 6000081137 6000081139 6000081140 6000081143 6000081144 6000081146 6000081148 6000081149 6000081152 6000081153 6000081154 6000081156 6000081157 6000081158 6000081159 6000081160 6000081162 6000081163 6000081165 6000081166 6000081167 6000081169 6000081170 6000081171 6000081173 6000081174 6000081177 6000081179 6000081180 6000081181 6000081182 6000081183 6000081185 6000081189 6000081190 6000081192 6000081193 6000081194 6000081195 6000081196 6000081198 6000081199 6000081200 6000081202 6000081203 6000081205 6000081206 6000081207 6000081208 6000081209 6000081210 6000081211 6000081212 6000081213 6000081214 6000081215 6000081216 6000081219 6000081220 6000081221 6000081223 6000081224 6000081225 6000081226 6000081227 6000081229 6000081230 6000081231 6000081232 6000081233 6000081238 6000081239 6000081241 6000081244 6000081248 6000081249 6000081250 6000081252 6000081254 6000081255 6000081256 6000081258 6000081259 6000081261 6000081262 6000081263 6000081264 6000081265 6000081266 6000081267 6000081269 6000081271 6000081272 6000081273 6000081274 6000081276 6000081277 6000081278 6000081279 6000081281 6000081282 6000081283 6000081284 6000081285 6000081286 6000081288 6000081292 6000081296 6000081297 6000081299 6000081300 6000081302 6000081304 6000081310 6000081311 6000081318 6000081322 6000081329 6000081330 6000081331 6000081332 6000081334 6000081344 6000081346 6000081348 6000081349 6000081350 6000081351 6000081352 6000081353 6000081356 6000081364 6000081365 6000081370 6000081372 6000081374 6000081375 6000081377 6000081383 6000081384 6000081387 6000081390 6000081392 6000081396 6000081398 6000081399 6000081402 6000081405 6000081406 6000081408 6000081409 6000081412 6000081416 6000081418 6000081419 6000081420 6000081421 6000081423 6000081424 6000081425 6000081426 6000081427 6000081428 6000081429 6000081430 6000081432 6000081433 6000081435 6000081438 6000081439 6000081441 6000081442 6000081444 6000081447 6000081450 6000081451 6000081455 6000081456 6000081458 6000081463 6000081465 6000081466 6000081468 6000081469 6000081471 6000081472 6000081473 6000081474 6000081477 6000081481 6000081484 6000081485 6000081486 6000081489 6000081491 6000081492 6000081493 6000081494 6000081496 6000081497 6000081499 6000081500 6000081501 6000081502 6000081503 6000081504 6000081510 6000081512 6000081519 6000081523 6000081525 6000081531 6000081535 6000081540 6000081541 6000081544 6000081546 6000081549 6000081550 6000081556 6000081559 6000081560 6000081563 6000081564 6000081570 6000081574 6000081575 6000081577 6000081580 6000081581 6000081582 6000081599 6000081602 6000081604 6000081605 6000081616 6000081618 6000081619 6000081625 6000081629 6000081632 6000081633 6000081634 6000081636 6000081640 6000081641 6000081650 6000081651 6000081658 6000081675 6000081681 6000081696 6000081701 6000081704 6000081707 6000081708 6000081709 6000081711 6000081714 6000081715 6000081716 6000081717 6000081718 6000081719 6000081720 6000081726 6000081728 6000081730 6000081733 6000081734 6000081736 6000081737 6000081738 6000081739 6000081741 6000081743 6000081745 6000081746 6000081747 6000081748 6000081749 6000081750 6000081751 6000081752 6000081753 6000081754 6000081755 6000081757 6000081758 6000081760 6000081761 6000081762 6000081763 6000081764 6000081766 6000081768 6000081774 6000081776 6000081778 6000081780 6000081782 6000081784 6000081787 6000081790 6000081791 6000081792 6000081793 6000081794 6000081795 6000081797 6000081799 6000081801 6000081803 6000081804 6000081806 6000081807 6000081808 6000081810 6000081812 6000081814 6000081818 6000081819 6000081822 6000081823 6000081824 6000081826 6000081828 6000081829 6000081830 6000081831 6000081832 6000081833 6000081835 6000081836 6000081838 6000081839 6000081842 6000081845 6000081846 6000081847 6000081848 6000081849 6000081851 6000081853 6000081855 6000081856 6000081859 6000081860 6000081863 6000081865 6000081866 6000081868 6000081871 6000081873 6000081875 6000081877 6000081879 6000081880 6000081882 6000081884 6000081905 6000081911 6000081912 6000081914 6000081915 6000081916 6000081919 6000081923 6000081925 6000081926 6000081929 6000081931 6000081932 6000081936 6000081941 6000081942 6000081944 6000081946 6000081947 6000081948 6000081950 6000081952 6000081953 6000081956 6000081958 6000081959 6000081962 6000081964 6000081971 6000081993 6000081994 6000081996 6000082002 6000082010 6000082012 6000082014 6000082018 6000082022 6000082023 6000082108 6000082111 6000082112 6000082116 6000082118 6000082122 6000082132 6000082139 6000082141 6000082142 6000082145 6000082166 6000082237 6000082252 6000082277 6000082283 6000082300 6000082311 6000082349 6000082352 6000082366 6000082370 6000082371 6000082410 6000082413 6000082418 6000082422 6000082426 6000082437 6000082464 6000082479 6000082483 6000082487 6000082491 6000082496 6000082500 6000082509 6000082512 6000082515 6000082534 6000082537 6000082540 6000082549 6000082559 6000082588 6000082596 6000082624 6000082626 6000082644 6000082664 6000082697 6000082698 6000082750 6000082772 6000082778 6000082779 6000082784 6000082819 6000082847 6000082869 6000082899 6000082903 6000082920 6000082928 6000082935 6000082938 6000082970 6000082971 6000082975 6000082978 6000082979 6000082986 6000082996 6000083001 6000083005 6000083008 6000083119 6000083143 6000083161 6000083167 6000083176 6000083209 6000083219 6000083223 6000083233 6000083234 6000083238 6000083242 6000083245 6000083247 6000083249 6000083253 6000083262 6000083264 6000083327 6000083338 6000083349 6000083403 6000083449 6000083474 6000083490 6000083494 6000083500 6000083501 6000083502 6000083514 6000083531 6000083538 6000083545 6000083549 6000083552 6000083564 6000083565 6000083571 6000083584 6000083593 6000083618 6000083620 6000083670 6000083707 6000083722 6000083730 6000083734 6000083737 6000083755 6000083762 6000083766 6000083768 6000083771 6000083773 6000083774 6000083777 6000083781 6000083784 6000083799 6000083810 6000083816 6000083821 6000083826 6000083843 6000084068 6000084502 6000085924 6000085929 6000085932 6000085952 6000085958 6000085962 6000085964 6000085989 6000086012 6000086034 6000086054 6000086056 6000086073 6000086080 6000086081 6000086085 6000086105 6000086119 6000086135 6000086142 6000086148 6000086168 6000086175 6000086234 6000086236 6000086250 6000086253 6000086272 6000086276 6000086301 6000086312 6000086326 6000086329 6000086335 6000086340 6000086393 6000086412 6000086439 6000086444 6000086446 6000086449 6000086469 6000086479 6000086483 6000086487 6000086496 6000086502 6000086513 6000086522 6000086525 6000086539 6000086540 6000086550 6000086625 6000086627 6000086648 6000086669 6000086671 6000086741 6000086744 6000086751 6000086759 6000086778 6000086860 6000086894 6000086912 6000086917 6000086981 6000086991 6000086998 6000086999 6000087025 6000087080 6000087108 6000087114 6000087115 6000087120 6000087122 6000087124 6000087190 6000087195 6000087215 6000087218 6000087220 6000087221 6000087222 6000087223 6000087224 6000087228 6000087229 6000087230 6000087231 6000087326 6000087328 6000087336 6000087338 6000087340 6000087342 6000087346 6000087354 6000087355 6000087356 6000087358 6000087396 6000087397 6000087399 6000087401 6000087402 6000087403 6000087405 6000087406 6000087408 6000087409 6000087411 6000087412 6000087416 6000087419 6000087420 6000087421 6000087423 6000087424 6000087426 6000087429 6000087430 6000087469 6000087470 6000087471 6000087472 6000087477 6000087481 6000087482 6000087484 6000087485 6000087488 6000087490 6000087500 6000087528 6000087531 6000087533 6000087534 6000087537 6000087548 6000087575 6000087576 6000087577 6000087578 6000087579 6000087580 6000087583 6000087584 6000087585 6000087586 6000087588 6000087590 6000087592 6000087618 6000087629 6000087630 6000087631 6000087632 6000087634 6000087635 6000087637 6000087638 6000087639 6000087642 6000087643 6000087644 6000087645 6000087669 6000087677 6000087679 6000087680 6000087740 6000087741 6000087748 6000087749 6000087754 6000087786 6000087791 6000087824 6000087829 6000087834 6000087840 6000087841 6000087843 6000087844 6000087849 6000087851 6000087891 6000087893 6000087895 6000087915 6000087944 6000087945 6000087958 6000087973 6000087979 6000087984 6000087998 6000088001 6000088032 6000088077 6000088137 6000088143 6000088148 6000088154 6000088315 6000088322 6000088329 6000088340 6000088345 6000088346 6000088349 6000088357 6000088364 6000088390 6000088391 6000088398 6000088399 6000088401 6000088446 6000088449 6000088451 6000088452 6000088455 6000088456 6000088457 6000088458 6000088459 6000088460 6000088479 6000088480 6000088483 6000088484 6000088487 6000088489 6000088490 6000088491 6000088492 6000088494 6000088496 6000088534 6000088536 6000088541 6000088569 6000088571 6000088573 6000088574 6000088576 6000088584 6000088585 6000088586 6000088641 6000088642 6000088643 6000088644 6000088645 6000088646 6000088649 6000088652 6000088654 6000088658 6000088660 6000088662 6000088663 6000088665 6000088666 6000088667 6000088668 6000088670 6000088671 6000088674 6000088675 6000088680 6000088689 6000088694 6000088718 6000088720 6000088732 6000088734 6000088735 6000088738 6000088743 6000088747 6000088772 6000088786 6000088799 6000088800 6000088807 6000088836 6000088856 6000088919 6000088930 6000088935 6000088962 6000088987 6000089004 6000089027 6000089033 6000089068 6000089074 6000089087 6000089089 6000089098 6000089107 6000089132 6000089133 6000089140 6000089144 6000089157 6000089158 6000089240 6000089259 6000089262 6000089271 6000089274 6000089281 6000089282 6000089304 6000089314 6000089349 6000089351 6000089355 6000089367 6000089382 6000089389 6000089390 6000089391 6000089399 6000089408 6000089411 6000089438 6000089440 6000089448 6000089450 6000089487 6000089499 6000089500 6000089510 6000089513 6000089520 6000089522 6000089524 6000089529 6000089535 6000089536 6000089570 6000089573 6000089582 6000089613 6000089614 6000089616 6000089619 6000089621 6000089622 6000089625 6000089632 6000089635 6000089660 6000089662 6000089663 6000089665 6000089677 6000089697 6000089700 6000089739 6000089741 6000089747 6000089752 6000089761 6000089774 6000089783 6000089784 6000089786 6000089863 6000089892 6000089893 6000089903 6000089946 6000090011 6000090021 6000090067 6000090071 6000090074 6000090077 6000090079 6000090263 6000091372 6000091476 6000091511 6000092022 6000092213 6000092214 6000092451 6000092810 6000092812 6000092819 6000092964 6000092969 6000093231 6000093447 6000093655 6000093819 6000093862 6000093884 6000093905 6000093921 6000093989 6000093995 6000094031 6000094038 6000094045 6000094058 6000094093 6000094106 6000094131 6000094144 6000094165 6000094196 6000094231 6000094273 6000094282 6000094288 6000094308 6000094335 6000094372 6000094399 6000094419 6000094677 6000094735 6000094740 6000094823 6000094836 6000094841 6000094944 6000094946 6000096949 6000096950 6000096951 6000096952 6000096953 6000096954 6000096955 6000096956 6000096957 6000096958 6000096959 6000096960 6000096961 6000096962 6000096963 6000096964 6000096965 6000096966 6000096967 6000096968 6000096969 6000096970 6000096971 6000096972 6000096973 6000096974 6000096975 6000096976 6000096977 6000096978 6000078924 6000081201 6000081268 6000080245 6000080344 6000080444 6000080684 6000080907 6000080937 6000081176 6000081431 6000081608 6000081756 6000080621 6000080457 6000081825 6000080431 6000080239 6000080423 6000080424 6000080475 6000080992 6000081607 6000081623 6000080410 6000081057 6000080685 6000080518 6000080178 6000081104 6000081119 6000080289 6000081175 6000081188 6000081789 6000081065 6000080210 6000080417 6000081260 6000081290 6000080229 6000080350 6000080484 6000080849 6000081257 6000080203 6000080774 6000080244 6000081075 6000080275 6000080254 6000080460 6000080317 6000080527 6000081735 6000081222 6000080653 6000080442 6000081132 6000081024 6000080865 6000080384 6000080680 6000081723 6000080490 6000080555 6000080638 6000080725 6000081765 6000069634 6000069642 6000071745 6000072317 6000072183 6000073434 6000071896 6000072642 6000076957 6000080156 6000080708 6000084129 6000084264 6000076282 6000084174 6000084466 6000084069 6000084180 6000069583 6000069976 6000069564 6000068889 6000071061 6000070070 6000071877 6000072920 6000069818 6000072775 6000072856 6000072865 6000072873 6000072946 6000073062 6000073066 6000073074 6000073089 6000073451 6000073556 6000073611 6000073746 6000073774 6000073907 6000073960 6000073981 6000074009 6000074012 6000074023 6000071274 6000073889 6000067144 6000067189 6000069744 6000071794 6000073939 6000071138 6000073628 6000073307 6000073815 6000076826 6000074405 6000075621 6000074170 6000074210 6000074520 6000074521 6000076539 6000076829 6000077732 6000077750 6000077840 6000078909 6000078652 6000089649 6000088344 6000074292 6000089515 6000077644 6000080872 6000081609 6000081660 6000082059 6000082455 6000082683 6000083066 6000086132 6000087530 6000093676
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Rad-G 9849Masimo Corporation · Oximeter1 recall150 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94170).
- Z-1536-2024 · Open, ClassifiedMasimo Rad-G, Pulse Oximeter with temperature (W/Sensor), REF:9210, Rx Only,
- Z-1537-2024 · Open, ClassifiedMasimo Rad-G, Pulse Oximeter (W/Sensor), REF:9847, Rx Only,
- Z-1539-2024 · Open, ClassifiedMasimo Rad-G, Pulse Oximeter (W/Patient Cable), REF:9895 Rx Only,
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page