Expo recallZ-1658-2024 · Boston Scientific Corporation · initiated Mar 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
FDA root cause category: Under Investigation by firm
What the company is doing
Boston Scientific issued an Urgent Medical Device Removal notice to its consignees on 03/21/2024 via overnight delivery. The notice explained the issue, the potential risk, and requested the removal of the affected products. Distributors were directed to notify their consignees.
Recall details
- Recall number
- Z-1658-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 21, 2024
- Posted by the FDA
- May 10, 2024
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 4715 catheters
- Product code
- DQO
- FDA submissions
- K152605
- Distribution
- Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.
Product description
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4.5 (5PK), REF H74908526232; cardiac catheter
Code information (models, lots, serial numbers)
UDI/DI 08714729187837, Batch Numbers: 60443928, 60457123, 60464496, 60473188, 60504888, 60522597, 60477305, 60517337, 60533589
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Expo™ H74908526231Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94240).
- Z-1638-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR3.5 (5PK), REF H74908526012; cardiac catheter
- Z-1639-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR4 (5PK), REF H74908526022; cardiac catheter
- Z-1640-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR5 (SGL), REF H74908526031; cardiac catheter
- Z-1641-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR5 (5PK), REF H74908526032; cardiac catheter
- Z-1642-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR6 (5PK), REF H74908526042; cardiac catheter
- Z-1643-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR4 MOD (5PK), REF H74908526052; cardiac catheter
- Z-1644-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AL3 (5PK), REF H749085261002; cardiac catheter
- Z-1645-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CAS1 (5PK), REF H749085261712; cardiac catheter
- Z-1646-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO KIMNY (5PK), REF H749085261802; cardiac catheter
- Z-1647-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO RCB (5PK), REF H749085261902; cardiac catheter
- Z-1648-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO LCB (5PK), REF H749085261952; cardiac catheter
- Z-1649-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO IM (5PK), REF H749085262012; cardiac catheter
- Z-1650-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO IMT (5PK), REF H749085262022; cardiac catheter
- Z-1651-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO IMC (5PK), REF H749085262032; cardiac catheter
- Z-1652-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL3 (5PK)REF H7490852620A2; cardiac catheter
- Z-1653-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL3.5 (5PK), REF H74908526212; cardiac catheter
- Z-1654-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4 125CM (5PK), REF H749085262202; cardiac catheter
- Z-1655-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4 (5PK), REF H74908526222; cardiac catheter
- Z-1656-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR4 125CM (5PK), REF H749085262252; cardiac catheter
- Z-1657-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR5 125CM (5PK), REF H749085262302; cardiac catheter
- Z-1659-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL5 125CM (5PK), REF H749085262352; cardiac catheter
- Z-1660-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO IM 125CM (5PK), REF H749085262402; cardiac catheter
- Z-1661-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO TIG4 (5PK), REF H749085262502; cardiac catheter
- Z-1662-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL6 (5PK), REF H74908526252; cardiac catheter
- Z-1663-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO JIG3.5 (5PK), REF H749085262522; cardiac catheter
- Z-1664-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO JIG4 (5PK), REF H749085262532; cardiac catheter
- Z-1665-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK (SGL), REF H749085263001; cardiac catheter
- Z-1666-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK (5PKX3), REF H749085263002; cardiac catheter
- Z-1667-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK 145 (SGL), REF H749085263011; cardiac catheter
- Z-1668-2024 · Open, ClassifiedExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK 145 (5PKX3), REF H749085263012; cardiac catheter
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page