Skip to content

RENASYS EDGE recallZ-1597-2024 · Smith & Nephew Medical, Ltd. · initiated Mar 11, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew Medical, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump

FDA root cause category: Under Investigation by firm

What the company is doing

Smith & Nephew Medical Ltd. issued Urgent Medical Device Recall Notice R-2024-02 on March 11, 2024 via overnight mail. Letter states reason for recall, health risk and action to take: Based on the regulatory classification and degree of hazard, this field action extends to the retail level. Therefore, it is not required to communicate this notice to the patient at the consumer level. If your customer contacts you regarding a pump failure: 1. Contact S+N Clinical Hotline (800-876-1261) with any questions. 2. Please ensure you provide a replacement as soon as possible. Replacement Stock. Smith+Nephew is supplying replacement stock quantities in parallel with this notice. Please contact [email protected] if you have any questions. If you have any questions or concerns regarding this recall please contact [email protected].

Recall details

Recall number
Z-1597-2024
Status
Open, Classified
Initiated by the company
Mar 11, 2024
Posted by the FDA
Apr 18, 2024
Recalling firm
Smith & Nephew Medical, Ltd.
Quantity
675 units
Product code
OMP
FDA submissions
K223041
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Chile.

Product description

RENASYS EDGE 800ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing Product Number: 66803140

Code information (models, lots, serial numbers)

GTIN: (01)05000223501082 Lot Numbers: 1016051, 1017549

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K223041 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94248).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page