Skip to content

Vyaire recallZ-1730-2024 · Vyaire Medical, Inc. · initiated Apr 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vyaire Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential of the nozzle separating during patient use.

FDA root cause category: Process control

What the company is doing

Vyaire Medical issued an URGENT Field Safety Notice to its consignees on 04/08/2024 via email. The notice explained the issue, requested a pull out test for all affected devices to verify the adhesive is adequate prior to use. If the product was further distributed, the party distributing device was directed to notify those to whom the device was sent. For any additional questions or concerns, please contact Vyaire at GMBEMEA- [email protected].

Recall details

Recall number
Z-1730-2024
Status
Open, Classified
Initiated by the company
Apr 8, 2024
Posted by the FDA
May 8, 2024
Recalling firm
Vyaire Medical, Inc.
Quantity
649 devices
Product code
BZC
FDA submissions
K150810
Distribution
US nationwide distribution.

Product description

Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX. The TwinTube sample line 240cm is connected to the adapter for DVT. Twin Tube Probenschlauch (707004), which is sold outside the US only. Used in breath analysis for metabolic response study.

Code information (models, lots, serial numbers)

UDI/DI 14250892903603, All lots prior to June 2023 or with a serial number before 04I00122 are potentially affected.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K150810 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page