ZPLP (Zimmer Periarticular Locking Plate System) Distal… recallZ-1709-2024 · Zimmer, Inc. · initiated Mar 20, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
FDA root cause category: Under Investigation by firm
What the company is doing
Zimmer issued Urgent Medical Device Recall Letter to Distributors Risk Managers and Surgeons on March 20, 2024 via mail/email. Letter states reason for recall, health risk and action to take: Our records indicate that you may have received one or more of the affected products. The affected units were distributed between March 2019 and January 2024. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals or surgeons and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgeme
Recall details
- Recall number
- Z-1709-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 20, 2024
- Posted by the FDA
- Apr 30, 2024
- Recalling firm
- Zimmer, Inc.
- Product code
- HWC
- FDA submissions
- K070906
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Product description
ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-017-04
Code information (models, lots, serial numbers)
UDI: (01)00889024088498(17)321205(10)65620069 (01)00889024088498(17)321205(10)65635922 (01)00889024088498(17)321205(10)65635923 (01)00889024088498(17)330111(10)65674071 (01)00889024088498(17)330115(10)65674072 (01)00889024088498(17)330428(10)66010704 (01)00889024088498(17)330730(10)66166096 (01)00889024088498(17)330822(10)66216633 Lot Numbers: 65620069 65635922 65635923 65674071 65674072 66010704 66166096 66216633
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 47-2357-017-04Zimmer, Inc. · PLATE, FIXATION, BONE1 recall1 report
Other devices under the same FDA submission
21 devices cleared under K070906 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 47-2357-017-06Zimmer, Inc. · Screw, fixation, bone1 recall2 reports
- NA 00-2357-018-06Zimmer, Inc. · Screw, fixation, bone1 recall1 report
- NA 00-2357-017-04Zimmer, Inc. · PLATE, FIXATION, BONE1 recall0 reports
- NA 00-2357-017-06Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- NA 00-2357-017-08Zimmer, Inc. · PLATE, FIXATION, BONE1 recall0 reports
- NA 00-2357-017-10Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- NA 00-2357-018-04Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- NA 00-2357-018-08Zimmer, Inc. · PLATE, FIXATION, BONE1 recall0 reports
- NA 00-2357-018-10Zimmer, Inc. · PLATE, FIXATION, BONE1 recall0 reports
- NA 47-2357-017-08Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- NA 47-2357-017-10Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- NA 47-2357-017-12Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94276).
- Z-1710-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste…
- Z-1711-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteo…
- Z-1712-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste…
- Z-1713-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste…
- Z-1714-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16 Holes, 236 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste…
- Z-1715-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-0…
- Z-1716-2024 · Open, ClassifiedZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteot…
- Z-1717-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteot…
- Z-1718-2024 · Open, ClassifiedZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteo…
- Z-1719-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste…
- Z-1720-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 14 Holes, 210 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-01…
- Z-1721-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 16 Holes, 236 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteo…
- Z-1722-2024 · Open, ClassifiedZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of os…
- Z-1723-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357…
- Z-1724-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357…
- Z-1725-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-235…
- Z-1726-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-0…
- Z-1727-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-…
- Z-1728-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Nonsterile-Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-0…
- Z-1729-2024 · Open, ClassifiedZPLP Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Nonsterile-Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page