ADVANCED RESEARCH MEDICAL Lumbar Interbody Fusion System… recallZ-1698-2024 · Advanced Research Medical, LLC · initiated Mar 14, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Advanced Research Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential shearing of the disposable blade may occur and be contained within the of Flexible Curette Shaft Assembly during surgery.

FDA root cause category: Device Design

What the company is doing

The sole distributor was notified by emailed letter of updates to the Surgical Technique Guide and IFU on 14 March 2023. The IFU has been updated to address a few small details including the direct listing of single use disposable components and the definition of single use being for one level only. The surgical technique guide has improved graphics, more thorough descriptions for some of the instrument assembly steps that your scrub techs may find useful and a statement regarding the appropriate use of the Flexible Curette Blade to prevent surgeons from using it in a way that may overstress the blade and cause a break. This information was shared with the surgeons through the sole distributor.

Recall details

Recall number
Z-1698-2024
Status
Completed
Initiated by the company
Mar 14, 2023
Posted by the FDA
Apr 29, 2024
Recalling firm
Advanced Research Medical, LLC
Quantity
1150 units
Product code
MAX
FDA submissions
K173947
Distribution
US Nationwide distribution in the state of Minnesota.

Product description

ADVANCED RESEARCH MEDICAL Lumbar Interbody Fusion System (OLLIF), Flexible Curette Loop Blade, Part # 01-06-4; treatment of degenerative disc disease

Code information (models, lots, serial numbers)

UDI/DI 00850014575311, Lot Numbers: 87458, BB059AA, BB140AA, BB140AB, BB307AA, BB350AA, BC061AA, BC130AA, BC130AA-2, BC274AA, BD083AA, BD083AB, BD083AC

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates