Keller Funnel 2 recallZ-1694-2024 · Allergan Sales, LLC · initiated Mar 14, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Allergan Sales, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is a potential that breast implants lack the lubricious coating that allows them to slide through the funnels during the implantation procedure.
FDA root cause category: Process design
What the company is doing
Between the dates of March 28, 2024 - April 1, 2024, Allergan Aesthetics (Allergan Sales LLC) issued a "Urgent Medical Device Recall" notification to affected consignees via FedEx. Allergan asked consignees to take the following actions: 1. If you have the affected product, discontinue further use and quarantine products prior to return. 2. Conduct a physical count of the affected product in your possession and record the count on the enclosed Business Response Form. 3. Send the Business Response Form to Qualanex LLC, AbbVie s third-party recall processor, via email within five (5) business days of receipt. To ensure we are able to account for all recalled product, it is imperative that you return the form. Please return the Business Response Form even if you have no recalled product to return. 4. Upon receipt of your Business Response Form, Qualanex LLC will issue a Return Authorization Form. When returning the recalled products, include the Return Authorization form and attach the prepaid Authorized Return shipping label to the outside of the return carton. Return the recalled product to: Qualanex, LLC 1410 Harris Rd. Libertyville, IL 60048 5. Please do not return any products that are not the subject of this recall. On 7/15/2025, updated recall notices were emailed to customers informing them of an additional impacted lot.
Recall details
- Recall number
- Z-1694-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 14, 2024
- Posted by the FDA
- Apr 29, 2024
- Recalling firm
- Allergan Sales, LLC
- Quantity
- 10837
- Product code
- KDD
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Product description
Keller Funnel 2, REF: HA-001, STERILEEO, Rx Only,
Code information (models, lots, serial numbers)
UDI: 10888628043725 / Lot #: 23B35C, 23B41C, 23B42C, 23C05C, 23C06C, 23D01C, 24C12C
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Keller Funnel 2 HA-005Allergan, Inc. · Kit, Surgical Instrument, Disposable1 recall39 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page