Skip to content

AIR-FLOW recallZ-1688-2024 · Electro Medical Systems SA · initiated Mar 13, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Electro Medical Systems SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).

FDA root cause category: Process control

What the company is doing

On 3/13/2024, the firm notified, via email, its distributors and asked them to extend the recall to end users. Currently the Customer Notification Letter to end users is being drafted (date of distribution pending). The Customer Notification Letter will information customers that the firm is recalling WATER, CLEANER and PIEZON Bottles due to a manufacturing issue that may result with these affected bottles to leak and/or break and in unfavorable cases may lead to injuries. Customers are instructed to: 1. Cease use of the affected products. 2. Return the affected products to their distributors For questions or assistance, contact EMS Dallas at 1-800-367-0367 or email to [email protected] Monday thru Friday 8:00 a.m. to 5:00 p.m. CST.

Recall details

Recall number
Z-1688-2024
Status
Open, Classified
Initiated by the company
Mar 13, 2024
Posted by the FDA
Apr 24, 2024
Recalling firm
Electro Medical Systems SA
Quantity
54
Product code
ELC
FDA submissions
K190124
Distribution
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA REPUBLIC OF, KOSOVO, KUWAIT, LUXEMBOURG, MALAYSIA, MEXICO, MOLDOVA REPUBLIC OF, NETHERLANDS, NEW ZEALAND, OMAN, PALESTINE, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, Thailand, UKRAINE,UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM.

Product description

Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Product Description: These bottles are sold empty and are filled by the user with water - in some cases small quantities of disinfectant solutions - for dental treatments. The bottle is pressurized on the device during use. Component: YES. AIRFLOW PROPHYLAXIS MASTER and AIRFLOW PROPHYLAXIS ONE

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Lot Code: Model number: EG-111, UDI: 07613353180249 Lot number: 16442-190, Expiry date: 2026-04 Model number: EG-111, UDI: 07613353180249 Lot number: 16694-30 Expiry date: 2026-07 AIRFLOW Prophylaxis Master UDI-DI codes: 07613353213183 and 07613353184148 AIRFLOW ONE UDI-DI code: 07613353184124 Product Serial Numbers: KU23667 KU23689 KU23700 KU23701 KU23703 KU23722 KU24328 KU24330 KU24331 KU24333 KU24334 KU24335 KU24337 KU24339 KU24340 KU24341 KU23719 KU24336 KU24338 KU24342 KU24344 KU24346 KU24349 KU24354 KU24356 KU23666 KU23672 KU23678 KU23686 KU23688 KU23706 KU23716 KU24343 KU24345 KU24347 KU24348 KU24350 KU24351 KU24353 KU24355 KU24357 KU24363 KU24366 KU24367 KU24368 KU24369 KU24370 KU24371 KU24372 KU24373 KU24359 KU24360 KU24361 KU24365

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94293).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page