INNOVA recallZ-1777-2024 · GE Medical Systems, SCS · initiated Mar 18, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, SCS or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
FDA root cause category: Process control
What the company is doing
GE Healthcare issued an UGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/18/2024 via letter. The notice explained the issue, hazard, and requested the following actions be taken: Every day prior to first use of the device, perform a functional check of the gantry movements and images of your device to look for the following behaviors: 1. Abnormal noise during gantry movement. 2. Image misalignment. If any of the above is observed, DO NOT use the device and call GE HealthCare Service Immediately. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Post this letter in a visible location beside the product. GE HealthCare will schedule a service appointment to correct this potential issue in a timely manner. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-1777-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 18, 2024
- Posted by the FDA
- May 6, 2024
- Recalling firm
- GE Medical Systems, SCS
- Quantity
- 21 units
- Product code
- OWB
- FDA submissions
- K181403
- Distribution
- Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
Product description
GE Healthcare Innova IGS 5, computed tomography x-ray system
Code information (models, lots, serial numbers)
UDI/DI 00840682124621, Serial Numbers: M2-23-042, M2-23-046, M2-23-048, M2-23-051, M2-24-001, M3-23-097, M3-23-098, M3-23-106, M3-23-108, M3-23-109, M3-23-111, M3-23-112, M3-23-114, M3-23-116, M3-23-123, M4-23-046, M4-23-053, M4-23-054, M4-23-055, M4-23-056, M4-24-001
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INNOVA IGS 5 001Ge Medical Systems · System, x-ray, fluoroscopic, image-intensified1 recall0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K181403 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Allia IGS 7 001Ge Medical Systems · System, x-ray, fluoroscopic, image-intensified1 recall0 reports
- Allia IGS 7 OR 001Ge Medical Systems · Interventional fluoroscopic x-ray system1 recall0 reports
- INNOVA IGS 6 001Ge Medical Systems · Interventional Fluoroscopic X-Ray System1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94305).
- Z-1778-2024 · Open, ClassifiedGE Healthcare Innova IGS 6, Interventional fluoroscopic x-ray systems
- Z-1779-2024 · Open, ClassifiedGE Healthcare Interact Discovery RT, Interventional fluoroscopic x-ray systems
- Z-1780-2024 · Open, ClassifiedGE Healthcare Allia IGS 7, Interventional fluoroscopic x-ray systems
- Z-1781-2024 · Open, ClassifiedGE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems
- Z-1782-2024 · Open, ClassifiedGE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systems
- Z-1783-2024 · Open, ClassifiedGE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systems
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.