Skip to content

Prelude Roadster recallZ-1875-2024 · Merit Medical Systems, Inc. · initiated Mar 14, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Their is a potential that sheaths labeled with .018 dilators may contain a .038 dilator.

FDA root cause category: Employee error

What the company is doing

On March 18, 2024 Merit Medical issued a "Urgent Medical Device Recall Notice to affected consignees via E-Mail. Merit Medical asked consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed Customer Response Form to Customer Service at [email protected] within 7 days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.

Recall details

Recall number
Z-1875-2024
Status
Open, Classified
Initiated by the company
Mar 14, 2024
Posted by the FDA
May 20, 2024
Recalling firm
Merit Medical Systems, Inc.
Quantity
27 units
Product code
DYB
FDA submissions
K211405
Distribution
Worldwide - US Nationwide distribution in the state of CA and the countries of New Zealand and Canada.

Product description

Prelude Guide Sheath Roadster, REF PG6F45S018, 6F 0.087" (2.2 mm), 45 cm, STERILE EO, RX ONLY

Code information (models, lots, serial numbers)

Lot # H2817178/ UDI: 00884450770771

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94304).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page