Skip to content

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical… recallZ-1696-2024 · Advanced Instruments, LLC · initiated Mar 11, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Advanced Instruments, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.

FDA root cause category: Under Investigation by firm

What the company is doing

Advanced Instruments issued Urgent Medical Device Recall Letter on 3/11/24 via email. Letter states reason for recall, health risk and action to take: 1. Discontinue use of your device. 2. Advanced Instruments will contact you regarding the return and exchange of your Device. Several solutions are available as needed to ensure continuity of testing. 3. For those users whose lab protocol requires the review of previous data, Advanced Instruments is available to review the data. 4. Complete and return the attached acknowledgement form. Advanced Instruments is implementing permanent corrective actions to address the two issues described and as part of the process above you will be provided with information on the next steps to implement a solution for impacted devices. If you need any further information or support concerning this issue, please contact [email protected]. To better assist you, please have your instrument serial number ready. 1-800-225-4034 (Toll Free US & Canada) Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.

Recall details

Recall number
Z-1696-2024
Status
Open, Classified
Initiated by the company
Mar 11, 2024
Posted by the FDA
Apr 29, 2024
Recalling firm
Advanced Instruments, LLC
Quantity
58 units
Product code
JJM
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, France, Germany, Iceland, Netherlands, South Korea, Spain, Switzerland, Taiwan.

Product description

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page