Rheo Knee recallZ-1788-2024 · Ossur H / F · initiated Mar 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ossur H / F or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
FDA root cause category: Software design
What the company is doing
On April 4, 2024, Ossur, issued a Urgent Medical Device Correction via E-Mail to affected consignees. Ossur asked consignees to take the following actions: REMOTELY UPDATE FIRMWARE ON AFFECTED DEVICES 1a. Check the serial number of devices against the list of affected devices in this letter to confirm the affected units. 2b. Update the firmware on the device via the Rheo Knee Update application This application can only be used to update the affected devices: i. A computer running Windows Operating System 10 or 11 is required to run the Rheo Knee Update application. It is not compatible with Android or iOS and cannot be completed via mobile phones. ii. Access the Rheo Knee Update application via this link: htps://www.ossur.com/global/prismic/generic/rheo-knee-update-language iii. Follow the steps outlined in the application to complete the firmware update. iv. If there are any issues with completing the firmware update successfully, please contact ssur technical support at 1-800-233-6263 ext. 10. 2. PLEASE PASS THIS NOTICE to those who need to be aware within your organization. If you have further distributed this product, please identify your customers, and notify them at once of this product alert. We recommend that you include a copy of this notice. 3. PLEASE MAINTAIN AWARENESS of this notice and required actions for an appropriate period to ensure effectiveness of the corrective action.
Recall details
- Recall number
- Z-1788-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 25, 2024
- Posted by the FDA
- May 10, 2024
- Recalling firm
- Ossur H / F
- Quantity
- 143 units
- Product code
- ISW
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Product description
RKN130003, RHEO KNEE 3 YR WARRANTY, MD, CE, External lower limb prosthetic component
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI: 05690967595418/ Serial Numbers: HF383471 HF383327 HF383283 HF366857 HF383410 HF375146 HF365047 HF375188 HF383312 HF375583 HF383377 HF376371 HF383431 HF376550 HF383493 HF376791 HF383276 HF377332 HF383294 HF377766 HF383319 HF378019 HF383346 HF378063 HF383387 HF378251 HF383419 HF378272 HF383437 HF378504 HF383485 HF378508 HF383506 HF378541 HF383273 HF378957 HF383281 HF379043 HF383291 HF379062 HF383310 HF379277 HF383315 HF379328 HF383321 HF379484 HF383342 HF380010 HF383355 HF380230 HF383384 HF380398 HF383395 HF380403 HF383412 HF380450 HF383426 HF380510 HF383434 HF381405 HF383445 HF381419 HF383482 HF381429 HF383488 HF381443 HF383503 HF381512 HF383509 HF381970 HF383268 HF382159 HF383274 HF382314 HF383279 HF382346 HF383282 HF382371 HF383285 HF382395 HF383292 HF382481 HF383309 HF382878 HF383311 HF383192 HF383314 HF383193 HF383316 HF383194 HF383320 HF383195 HF383322 HF383196 HF383328 HF383198 HF383344 HF383200 HF383352 HF383210 HF383376 HF383216 HF383383 HF383218 HF383385 HF383219 HF383388 HF383220 HF383396 HF383221 HF383411 HF383222 HF383413 HF383224 HF383421 HF383230 HF383427 HF383234 HF383433 HF383237 HF383435 HF383238 HF383444 HF383239 HF383454 HF383240 HF383472 HF383241 HF383483 HF383243 HF383486 HF383244 HF383491 HF383245 HF383494 HF383246 HF383504 HF383262 HF383507 HF383264 HF383532 HF383266 HF383267
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Rheo Knee RKN130003Ossur Americas, Inc. · Joint, Knee, External Limb Component1 recall5 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94332).
- Z-1787-2024 · Open, ClassifiedRKN130002, RHEO KNEE 2 YR WARRANTY, MD, CE, External lower limb prosthetic component
- Z-1789-2024 · Open, ClassifiedRKNXC0003, RHEO KNEE XC 3 YR WARRANTY, MD, CE, External lower limb prosthetic component
- Z-1790-2024 · Open, ClassifiedRKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page