Sofsilk recallZ-1697-2024 · Covidien, LP · initiated Jan 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
FDA root cause category: Process control
What the company is doing
On January 25, 2024, Medtronic issued an expansion to their earlier recall for sutures (RES 93768) concerning gamma sterilization. As part of that expansion, there was an additional lot of product which was exposed to more than the approved number of EO sterilization cycles, which is covered in this recall, RES 94351. The issue was communicated to affected consignees via letters. Consignees are asked to return any affected product they may have, forward this notification to those who need to be aware within the organization and to any location which the devices have been transferred.
Recall details
- Recall number
- Z-1697-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 25, 2024
- Posted by the FDA
- Apr 29, 2024
- Recalling firm
- Covidien, LP
- Quantity
- 36 total, OUS distribution only
- Product code
- GAP
- FDA submissions
- K981128
- Distribution
- International distribution in the countries of Portugal and Italy.
Product description
Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.
Code information (models, lots, serial numbers)
UDI-DI 20884521086422 Lot D2F2101RY
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sofsilk S-2782KCovidien LP · SUTURE, NONABSORBABLE, SILK3 recalls0 reports
Product lines
Other devices under the same FDA submission
116 devices cleared under K981128 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Sofsilk S-1768KCovidien LP · SUTURE, NONABSORBABLE, SILK2 recalls5 reports
- Sofsilk S-304Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls2 reports
- Sofsilk S-316Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls2 reports
- Sofsilk S-187Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls1 report
- Sofsilk S-605Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls1 report
- McKesson S683GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Silk2 recalls0 reports
- Sofsilk CS-211Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-1172Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-1750KCovidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-176Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-182Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-184Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page