LOSPA Tibial Insert recallZ-1934-2024 · Corentec CO., Ltd · initiated Mar 29, 2024
Class II Open, Classified All recalls posted in May 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Corentec CO., Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to unsupported 10 year expiration date.
FDA root cause category: Labeling Change Control
What the company is doing
On 03/29/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that on 7/1/2020 the firm changed the labeling of their polyethylene (PE) and patella inserts from a 5 year shelf life to a 10 year shelf life. This change was made without adequate documentation evidence to support the increase in shelf life. Customers are instructed to: 1. Inform individuals within your organization who need to be aware of this device recall. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices. 4. Maintain awareness of this notice internally until all required actions have been completed within your facility. 5. Inform Corentec if any of the subject devices have been distributed to other organizations. a. Please provide Corentec contact details so that Corentec can inform the recipients appropriately. 6. Please complete the attached customer response form (acknowledgement form). It is possible that you no longer have any physical inventory of the impacted product on site. Completing this form will allow us to update our records and will also negate the need for us to send any further communications on this matter. Therefore, please complete the customer response form even if you no longer have any of the subject devices in your physical inventory. 7. Return the completed form to Corentec. For any questions or assistance contact Bobby Pham at 310-488-2886 or via email [email protected].
Recall details
- Recall number
- Z-1934-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Mar 29, 2024
- Posted by the FDA
- May 31, 2024
- Recalling firm
- Corentec CO., Ltd
- Quantity
- 397
- Product code
- JWH
- FDA submissions
- K110404, K130673, K160157
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, MI, PA, TX.
Product description
LOSPA Tibial Insert Model/Catalog Number: See below Software Version: N/A Product Description: Tibial Insert
Code information (models, lots, serial numbers)
Show the full code information (48k characters)
Lot Code: LOSPA Tibial Insert (CR) Model No: 01.10.1XX LOSPA Tibial Insert (PS) Model No: 01.10.2XX LOSPA Fixed Constrained Tibial Insert Model No: K2.IBN.4061-XXX, K2.IBN.4264-XXX, K2.IBN.4467-XXX, K2.IBN.4672-XXX, K2.IBN.4874-XXX, K2.IBN.5076-XXX, K2.IBN.5281-XXX, K2.IBN.5684-XXX UDI/Lot Numbers: See attached Lospa recall Attachment 1 Lospa Knee System 01.10.132 Lospa Tibial Insert (CR) - Size 3 / 10mm 0108806373820953172605281070AFDM33 70AFDM33 5/28/2026 Lospa Knee System 01.10.132 Lospa Tibial Insert (CR) - Size 3 / 10mm 0108806373820953172703281070AFL130 70AFL130 3/28/2027 Lospa Knee System 01.10.134 Lospa Tibial Insert (CR) - Size 3 / 12mm 0108806373820960172605281070AFDM34 70AFDM34 5/28/2026 Lospa Knee System 01.10.134 Lospa Tibial Insert (CR) - Size 3 / 12mm 0108806373820960172712281070AFL131 70AFL131 12/28/2027 Lospa Knee System 01.10.136 Lospa Tibial Insert (CR) - Size 3 / 14mm 0108806373820977172510281070AEK548 70AEK548 10/28/2025 Lospa Knee System 01.10.136 Lospa Tibial Insert (CR) - Size 3 / 14mm 0108806373820977172601281070AEK548 70AEK548 1/28/2026 Lospa Knee System 01.10.138 Lospa Tibial Insert (CR) - Size 3 / 16mm 0108806373820984172602281070AFAR33 70AFAR33 2/28/2026 Lospa Knee System 01.10.138 Lospa Tibial Insert (CR) - Size 3 / 16mm 0108806373820984172710281070AFL133 70AFL133 10/28/2027 Lospa Knee System 01.10.138 Lospa Tibial Insert (CR) - Size 3 / 16mm 0108806373820984172712281070AFL133 70AFL133 12/28/2027 Lospa Knee System 01.10.138 Lospa Tibial Insert (CR) - Size 3 / 16mm 0108806373820984172801281070AFL133 70AFL133 1/28/2028 Lospa Knee System 01.10.13A Lospa Tibial Insert (CR) - Size 3 / 18mm 0108806373820991172601281070AEKE31 70AEKE31 1/28/2026 Lospa Knee System 01.10.142 Lospa Tibial Insert (CR) - Size 4 / 10mm 0108806373821158172811281070AHME32 70AHME32 11/28/2028 Lospa Knee System 01.10.142 Lospa Tibial Insert (CR) - Size 4 / 10mm 0108806373821158172812281070AJA734 70AJA734 12/28/2028 Lospa Knee System 01.10.146 Lospa Tibial Insert (CR) - Size 4 / 14mm 0108806373821172172509281070AEK632 70AEK632 9/28/2025 Lospa Knee System 01.10.146 Lospa Tibial Insert (CR) - Size 4 / 14mm 0108806373821172172602281070AFAR35 70AFAR35 2/28/2026 Lospa Knee System 01.10.146 Lospa Tibial Insert (CR) - Size 4 / 14mm 0108806373821172172803281070AHBR39 70AHBR39 3/28/2028 Lospa Knee System 01.10.148 Lospa Tibial Insert (CR) - Size 4 / 16mm 0108806373821189172512281070AELH38 70AELH38 12/28/2025 Lospa Knee System 01.10.14A Lospa Tibial Insert (CR) - Size 4 / 18mm 0108806373821196172602281070AFAR36 70AFAR36 2/28/2026 Lospa Knee System 01.10.156 Lospa Tibial Insert (CR) - Size 5 / 14mm 0108806373821271172804281060AHAA36 60AHAA36 4/28/2028 Lospa Knee System 01.10.15A Lospa Tibial Insert (CR) - Size 5 / 18mm 0108806373821295172612281060AFL143 60AFL143 12/28/2026 Lospa Knee System 01.10.178 Lospa Tibial Insert (CR) - Size 7 / 16mm 0108806373821486172812281060AHLF32 60AHLF32 12/28/2028 Lospa Knee System 01.10.17A Lospa Tibial Insert (CR) - Size 7 / 18mm 0108806373821493172710281060AFL148 60AFL148 10/28/2027 Lospa Knee System 01.10.17A Lospa Tibial Insert (CR) - Size 7 / 18mm 0108806373821493172902281060AHAJ33 60AHAJ33 2/28/2029 Lospa Knee System 01.10.17C Lospa Tibial Insert (CR) - Size 7 / 20mm 0108806373821509172804281060AHDB36 60AHDB36 4/28/2028 Lospa Knee System 01.10.184 Lospa Tibial Insert (CR) - Size 8 / 12mm 0108806373821561172807281070AHFX35 70AHFX35 7/28/2028 Lospa Knee System 01.10.194 Lospa Tibial Insert (CR) - Size 9 / 12mm 0108806373821660172602281080AFAR40 80AFAR40 2/28/2026 Lospa Knee System 01.10.196 Lospa Tibial Insert (CR) - Size 9 / 14mm 0108806373821677172605281080AFCR45 80AFCR45 5/28/2026 Lospa Knee System 01.10.196 Lospa Tibial Insert (CR) - Size 9 / 14mm 0108806373821677172902281080AJBF34 80AJBF34 2/28/2029 Lospa Knee System 01.10.198 Lospa Tibial Insert (CR) - Size 9 / 16mm 0108806373821684172902281080AJC239 80AJC239 2/28/2029 Lospa Knee System 01.10.19A Lospa Tibial Insert (CR) - Size 9 / 18mm 0108806373821691172902281080AJC330 80AJC330 2/28/2029 Lospa Knee System 01.10.1B2 Lospa Tibial Insert (CR) - Size 11 / 10mm 0108806373821851172902281030AJC331 30AJC331 2/28/2029 Lospa Knee System 01.10.1B4 Lospa Tibial Insert (CR) - Size 11 / 12mm 0108806373821868172902281030AJC332 30AJC332 2/28/2029 Lospa Knee System 01.10.1C6 Lospa Tibial Insert (CR) - Size 12 / 14mm 0108806373821974172303281060accs32 60accs32 3/28/2023 Lospa Knee System 01.10.1C6 Lospa Tibial Insert (CR) - Size 12 / 14mm 0108806373821974172902281060AHH936 60AHH936 2/28/2029 Lospa Knee System 01.10.1C6 Lospa Tibial Insert (CR) - Size 12 / 14mm 0108806373821974172909281060AJK237 60AJK237 9/28/2029 Lospa Knee System 01.10.1C8 Lospa Tibial Insert (CR) - Size 12 / 16mm 0108806373821981172902281060AHA133 60AHA133 2/28/2029 Lospa Knee System 01.10.232 Lospa Tibial Insert (PS) - Size 3 / 10mm 0108806373822353172511281070AEK536 70AEK536 11/28/2025 Lospa Knee System 01.10.232 Lospa Tibial Insert (PS) - Size 3 / 10mm 0108806373822353172605281070AFDL30 70AFDL30 5/28/2026 Lospa Knee System 01.10.232 Lospa Tibial Insert (PS) - Size 3 / 10mm 0108806373822353172701281070AGA137 70AGA137 1/28/2027 Lospa Knee System 01.10.232 Lospa Tibial Insert (PS) - Size 3 / 10mm 0108806373822353172812281070AJA334 70AJA334 12/28/2028 Lospa Knee System 01.10.232 Lospa Tibial Insert (PS) - Size 3 / 10mm 0108806373822353172903281070AJD632 70AJD632 3/28/2029 Lospa Knee System 01.10.234 Lospa Tibial Insert (PS) - Size 3 / 12mm 0108806373822360172503281070AEB930 70AEB930 3/28/2025 Lospa Knee System 01.10.234 Lospa Tibial Insert (PS) - Size 3 / 12mm 0108806373822360172506281070AEDF31 70AEDF31 6/28/2025 Lospa Knee System 01.10.234 Lospa Tibial Insert (PS) - Size 3 / 12mm 0108806373822360172603281070AFD534 70AFD534 3/28/2026 Lospa Knee System 01.10.234 Lospa Tibial Insert (PS) - Size 3 / 12mm 0108806373822360172611281070AFG431 70AFG431 11/28/2026 Lospa Knee System 01.10.234 Lospa Tibial Insert (PS) - Size 3 / 12mm 0108806373822360172710281070AGL433 70AGL433 10/28/2027 Lospa Knee System 01.10.236 Lospa Tibial Insert (PS) - Size 3 / 14mm 0108806373822377172407281070ADGA32 70ADGA32 7/28/2024 Lospa Knee System 01.10.236 Lospa Tibial Insert (PS) - Size 3 / 14mm 0108806373822377172412281070AEAC32 70AEAC32 12/28/2024 Lospa Knee System 01.10.236 Lospa Tibial Insert (PS) - Size 3 / 14mm 0108806373822377172512281070AEMM32 70AEMM32 12/28/2025 Lospa Knee System 01.10.236 Lospa Tibial Insert (PS) - Size 3 / 14mm 0108806373822377172710281070AGL434 70AGL434 10/28/2027 Lospa Knee System 01.10.236 Lospa Tibial Insert (PS) - Size 3 / 14mm 0108806373822377172812281070AHBR37 70AHBR37 12/28/2028 Lospa Knee System 01.10.238 Lospa Tibial Insert (PS) - Size 3 / 16mm 0108806373822384172601281070AFAR32 70AFAR32 1/28/2026 Lospa Knee System 01.10.238 Lospa Tibial Insert (PS) - Size 3 / 16mm 0108806373822384172711281070AGLR30 70AGLR30 11/28/2027 Lospa Knee System 01.10.23A Lospa Tibial Insert (PS) - Size 3 / 18mm 0108806373822391172608281070AFH131 70AFH131 8/28/2026 Lospa Knee System 01.10.23C Lospa Tibial Insert (PS) - Size 3 / 20mm 0108806373822407172807281070AFH132 70AFH132 7/28/2028 Lospa Knee System 01.10.242 Lospa Tibial Insert (PS) - Size 4 / 10mm 0108806373822551172503281070AEB933 70AEB933 3/28/2025 Lospa Knee System 01.10.244 Lospa Tibial Insert (PS) - Size 4 / 12mm 0108806373822568172603281070AFCR38 70AFCR38 3/28/2026 Lospa Knee System 01.10.246 Lospa Tibial Insert (PS) - Size 4 / 14mm 0108806373822575172503281070AEB935 70AEB935 3/28/2025 Lospa Knee System 01.10.246 Lospa Tibial Insert (PS) - Size 4 / 14mm 0108806373822575172603281070AFCR40 70AFCR40 3/28/2026 Lospa Knee System 01.10.246 Lospa Tibial Insert (PS) - Size 4 / 14mm 0108806373822575172710281070AGEY32 70AGEY32 10/28/2027 Lospa Knee System 01.10.246 Lospa Tibial Insert (PS) - Size 4 / 14mm 0108806373822575172808281070AHBR35 70AHBR35 8/28/2028 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172502281070AEB936 70AEB936 2/28/2025 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172703281070AGCL37 70AGCL37 3/28/2027 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172710281070AGL437 70AGL437 10/28/2027 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172811281070AHK731 70AHK731 11/28/2028 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172902281070AJAK35 70AJAK35 2/28/2029 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172903281070AJBD37 70AJBD37 3/28/2029 Lospa Knee System 01.10.248 Lospa Tibial Insert (PS) - Size 4 / 16mm 0108806373822582172907281070AJDT32 70AJDT32 7/28/2029 Lospa Knee System 01.10.24A Lospa Tibial Insert (PS) - Size 4 / 18mm 0108806373822599172603281070AFD540 70AFD540 3/28/2026 Lospa Knee System 01.10.24A Lospa Tibial Insert (PS) - Size 4 / 18mm 0108806373822599172701281070AFLJ32 70AFLJ32 1/28/2027 Lospa Knee System 01.10.24A Lospa Tibial Insert (PS) - Size 4 / 18mm 0108806373822599172702281070AFLJ32 70AFLJ32 2/28/2027 Lospa Knee System 01.10.24A Lospa Tibial Insert (PS) - Size 4 / 18mm 0108806373822599172710281070AGL438 70AGL438 10/28/2027 Lospa Knee System 01.10.24A Lospa Tibial Insert (PS) - Size 4 / 18mm 0108806373822599172903281070AJBD38 70AJBD38 3/28/2029 Lospa Knee System 01.10.24C Lospa Tibial Insert (PS) - Size 4 / 20mm 0108806373822605172601281070AFH135 70AFH135 1/28/2026 Lospa Knee System 01.10.24C Lospa Tibial Insert (PS) - Size 4 / 20mm 0108806373822605172609281070AFH135 70AFH135 9/28/2026 Lospa Knee System 01.10.24C Lospa Tibial Insert (PS) - Size 4 / 20mm 0108806373822605172807281070AHGK30 70AHGK30 7/28/2028 Lospa Knee System 01.10.24C Lospa Tibial Insert (PS) - Size 4 / 20mm 0108806373822605172906281070AJFN30 70AJFN30 6/28/2029 Lospa Knee System 01.10.254 Lospa Tibial Insert (PS) - Size 5 / 12mm 0108806373822667172904281060AJBS31 60AJBS31 4/28/2029 Lospa Knee System 01.10.256 Lospa Tibial Insert (PS) - Size 5 / 14mm 0108806373822674172708281060AGGE37 60AGGE37 8/28/2027 Lospa Knee System 01.10.256 Lospa Tibial Insert (PS) - Size 5 / 14mm 0108806373822674172812281060AHMH30 60AHMH30 12/28/2028 Lospa Knee System 01.10.256 Lospa Tibial Insert (PS) - Size 5 / 14mm 0108806373822674172906281060AJBS32 60AJBS32 6/28/2029 Lospa Knee System 01.10.256 Lospa Tibial Insert (PS) - Size 5 / 14mm 0108806373822674172908281060AJG137 60AJG137 8/28/2029 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172604281060AFDL40 60AFDL40 4/28/2026 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172702281060AGB533 60AGB533 2/28/2027 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172709281060AGGE38 60AGGE38 9/28/2027 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172710281060AGJG36 60AGJG36 10/28/2027 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172907281060AJEY35 60AJEY35 7/28/2029 Lospa Knee System 01.10.258 Lospa Tibial Insert (PS) - Size 5 / 16mm 0108806373822681172909281060AJEY35 60AJEY35 9/28/2029 Lospa Knee System 01.10.25A Lospa Tibial Insert (PS) - Size 5 / 18mm 0108806373822698172512281060AEK535 60AEK535 12/28/2025 Lospa Knee System 01.10.25A Lospa Tibial Insert (PS) - Size 5 / 18mm 0108806373822698172701281060AFLJ34 60AFLJ34 1/28/2027 Lospa Knee System 01.10.25A Lospa Tibial Insert (PS) - Size 5 / 18mm 0108806373822698172710281060AGL440 60AGL440 10/28/2027 Lospa Knee System 01.10.25A Lospa Tibial Insert (PS) - Size 5 / 18mm 0108806373822698172909281060AJKN31 60AJKN31 9/28/2029 Lospa Knee System 01.10.25C Lospa Tibial Insert (PS) - Size 5 / 20mm 0108806373822704172808281060AHHT30 60AHHT30 8/28/2028 Lospa Knee System 01.10.264 Lospa Tibial Insert (PS) - Size 6 / 12mm 0108806373822766172701281000AGB331 00AGB331 1/28/2027 Lospa Knee System 01.10.266 Lospa Tibial Insert (PS) - Size 6 / 14mm 0108806373822773172701281000AGB332 00AGB332 1/28/2027 Lospa Knee System 01.10.266 Lospa Tibial Insert (PS) - Size 6 / 14mm 0108806373822773172801281000AGB332 00AGB332 1/28/2028 Lospa Knee System 01.10.268 Lospa Tibial Insert (PS) - Size 6 / 16mm 0108806373822780172806281000AFJ137 00AFJ137 6/28/2028 Lospa Knee System 01.10.274 Lospa Tibial Insert (PS) - Size 7 / 12mm 0108806373822865172503281060AEBA31 60AEBA31 3/28/2025 Lospa Knee System 01.10.274 Lospa Tibial Insert (PS) - Size 7 / 12mm 0108806373822865172902281060AJAK43 60AJAK43 2/28/2029 Lospa Knee System 01.10.276 Lospa Tibial Insert (PS) - Size 7 / 14mm 0108806373822872172501281060AEAD32 60AEAD32 1/28/2025 Lospa Knee System 01.10.276 Lospa Tibial Insert (PS) - Size 7 / 14mm 0108806373822872172701281060AGB133 60AGB133 1/28/2027 Lospa Knee System 01.10.276 Lospa Tibial Insert (PS) - Size 7 / 14mm 0108806373822872172808281060AHBR31 60AHBR31 8/28/2028 Lospa Knee System 01.10.276 Lospa Tibial Insert (PS) - Size 7 / 14mm 0108806373822872172812281060AJA744 60AJA744 12/28/2028 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172603281060AFCM34 60AFCM34 3/28/2026 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172605281060AFEN43 60AFEN43 5/28/2026 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172701281060AFLJ36 60AFLJ36 1/28/2027 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172705281060AGDG39 60AGDG39 5/28/2027 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172710281060AGKD37 60AGKD37 10/28/2027 Lospa Knee System 01.10.278 Lospa Tibial Insert (PS) - Size 7 / 16mm 0108806373822889172807281060AHBR32 60AHBR32 7/28/2028 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172512281060AEK544 60AEK544 12/28/2025 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172608281060AFH149 60AFH149 8/28/2026 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172701281060AFLJ35 60AFLJ35 1/28/2027 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172708281060AGGE41 60AGGE41 8/28/2027 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172710281060AGL431 60AGL431 10/28/2027 Lospa Knee System 01.10.27A Lospa Tibial Insert (PS) - Size 7 / 18mm 0108806373822896172711281060AGLR43 60AGLR43 11/28/2027 Lospa Knee System 01.10.27C Lospa Tibial Insert (PS) - Size 7 / 20mm 0108806373822902172507281060AEG932 60AEG932 7/28/2025 Lospa Knee System 01.10.27C Lospa Tibial Insert (PS) - Size 7 / 20mm 0108806373822902172807281060AHDB30 60AHDB30 7/28/2028 Lospa Knee System 01.10.27C Lospa Tibial Insert (PS) - Size 7 / 20mm 0108806373822902172906281060AJFN32 60AJFN32 6/28/2029 Lospa Knee System 01.10.27C Lospa Tibial Insert (PS) - Size 7 / 20mm 0108806373822902172908281060AJFN32 60AJFN32 8/28/2029 Lospa Knee System 01.10.288 Lospa Tibial Insert (PS) - Size 8 / 16mm 0108806373822988172701281070AGB337 70AGB337 1/28/2027 Lospa Knee System 01.10.288 Lospa Tibial Insert (PS) - Size 8 / 16mm 0108806373822988172908281070AGB337 70AGB337 8/28/2029 Lospa Knee System 01.10.28A Lospa Tibial Insert (PS) - Size 8 / 18mm 0108806373822995172701281070AGB338 70AGB338 1/28/2027 Lospa Knee System 01.10.292 Lospa Tibial Insert (PS) - Size 9 / 10mm 0108806373823053172404281080ADCH30 80ADCH30 4/28/2024 Lospa Knee System 01.10.292 Lospa Tibial Insert (PS) - Size 9 / 10mm 0108806373823053172710281080AGKD38 80AGKD38 10/28/2027 Lospa Knee System 01.10.292 Lospa Tibial Insert (PS) - Size 9 / 10mm 0108806373823053172712281080AGLS41 80AGLS41 12/28/2027 Lospa Knee System 01.10.292 Lospa Tibial Insert (PS) - Size 9 / 10mm 0108806373823053172906281080AJFK35 80AJFK35 6/28/2029 Lospa Knee System 01.10.294 Lospa Tibial Insert (PS) - Size 9 / 12mm 0108806373823060172507281080AEET35 80AEET35 7/28/2025 Lospa Knee System 01.10.294 Lospa Tibial Insert (PS) - Size 9 / 12mm 0108806373823060172701281080AGB135 80AGB135 1/28/2027 Lospa Knee System 01.10.294 Lospa Tibial Insert (PS) - Size 9 / 12mm 0108806373823060172703281080AGC933 80AGC933 3/28/2027 Lospa Knee System 01.10.294 Lospa Tibial Insert (PS) - Size 9 / 12mm 0108806373823060172810281080AHLL30 80AHLL30 10/28/2028 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172405281080ADCH34 80ADCH34 5/28/2024 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172504281080AEBA36 80AEBA36 4/28/2025 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172512281080AEMM42 80AEMM42 12/28/2025 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172604281080AFDL48 80AFDL48 4/28/2026 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172703281080AGCL32 80AGCL32 3/28/2027 Lospa Knee System 01.10.296 Lospa Tibial Insert (PS) - Size 9 / 14mm 0108806373823077172810281080AHLL31 80AHLL31 10/28/2028 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172403281080ADBJ30 80ADBJ30 3/28/2024 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172502281080AEBA37 80AEBA37 2/28/2025 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172609281080AFH230 80AFH230 9/28/2026 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172711281080AGLR48 80AGLR48 11/28/2027 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172712281080AGLR48 80AGLR48 12/28/2027 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172811281080AHHM31 80AHHM31 11/28/2028 Lospa Knee System 01.10.298 Lospa Tibial Insert (PS) - Size 9 / 16mm 0108806373823084172902281080AJA748 80AJA748 2/28/2029 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172406281080ADCH35 80ADCH35 6/28/2024 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172603281080AFD548 80AFD548 3/28/2026 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172610281080AFH231 80AFH231 10/28/2026 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172611281080AFH231 80AFH231 11/28/2026 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172611281080AFLJ38 80AFLJ38 11/28/2026 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172707281080AFLJ38 80AFLJ38 7/28/2027 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172710281080AGL442 80AGL442 10/28/2027 Lospa Knee System 01.10.29A Lospa Tibial Insert (PS) - Size 9 / 18mm 0108806373823091172812281080AGLR49 80AGLR49 12/28/2028 Lospa Knee System 01.10.2B2 Lospa Tibial Insert (PS) - Size 11 / 10mm 0108806373823251172610281030AFG739 30AFG739 10/28/2026 Lospa Knee System 01.10.2B2 Lospa Tibial Insert (PS) - Size 11 / 10mm 0108806373823251172711281030AGKD40 30AGKD40 11/28/2027 Lospa Knee System 01.10.2B2 Lospa Tibial Insert (PS) - Size 11 / 10mm 0108806373823251172801281030AGLS43 30AGLS43 1/28/2028 Lospa Knee System 01.10.2B4 Lospa Tibial Insert (PS) - Size 11 / 12mm 0108806373823268172502281030AEBA40 30AEBA40 2/28/2025 Lospa Knee System 01.10.2B4 Lospa Tibial Insert (PS) - Size 11 / 12mm 0108806373823268172509281030AEHF43 30AEHF43 9/28/2025 Lospa Knee System 01.10.2B4 Lospa Tibial Insert (PS) - Size 11 / 12mm 0108806373823268172711281030AGKD41 30AGKD41 11/28/2027 Lospa Knee System 01.10.2B4 Lospa Tibial Insert (PS) - Size 11 / 12mm 0108806373823268172905281030AJCK32 30AJCK32 5/28/2029 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172411281030ADM439 30ADM439 11/28/2024 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172501281030AEAD40 30AEAD40 1/28/2025 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172507281030AEBA41 30AEBA41 7/28/2025 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172609281030AFG741 30AFG741 9/28/2026 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172610281030AFG741 30AFG741 10/28/2026 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172704281030AGDH36 30AGDH36 4/28/2027 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172710281030AGL443 30AGL443 10/28/2027 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172808281030AGLS45 30AGLS45 8/28/2028 Lospa Knee System 01.10.2B6 Lospa Tibial Insert (PS) - Size 11 / 14mm 0108806373823275172901281030AJAJ31 30AJAJ31 1/28/2029 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172501281030AEAP36 30AEAP36 1/28/2025 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172603281030AFD631 30AFD631 3/28/2026 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172609281030AFH232 30AFH232 9/28/2026 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172705281030AGB835 30AGB835 5/28/2027 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172710281030AGL444 30AGL444 10/28/2027 Lospa Knee System 01.10.2B8 Lospa Tibial Insert (PS) - Size 11 / 16mm 0108806373823282172810281030AHFM31 30AHFM31 10/28/2028 Lospa Knee System 01.10.2BA Lospa Tibial Insert (PS) - Size 11 / 18mm 0108806373823299172406281030ADCH38 30ADCH38 6/28/2024 Lospa Knee System 01.10.2BC Lospa Tibial Insert (PS) - Size 11 / 20mm 0108806373823305172412281020ACC543 20ACC543 12/28/2024 Lospa Knee System 01.10.2BC Lospa Tibial Insert (PS) - Size 11 / 20mm 0108806373823305172608281030AFJ133 30AFJ133 8/28/2026 Lospa Knee System 01.10.2BC Lospa Tibial Insert (PS) - Size 11 / 20mm 0108806373823305172609281030AFJ133 30AFJ133 9/28/2026 Lospa Knee System 01.10.2C2 Lospa Tibial Insert (PS) - Size 12 / 10mm 0108806373823350172507281060AEHF30 60AEHF30 7/28/2025 Lospa Knee System 01.10.2C2 Lospa Tibial Insert (PS) - Size 12 / 10mm 0108806373823350172909281060AJJG30 60AJJG30 9/28/2029 Lospa Knee System 01.10.2C4 Lospa Tibial Insert (PS) - Size 12 / 12mm 0108806373823367172407281060ADF236 60ADF236 7/28/2024 Lospa Knee System 01.10.2C4 Lospa Tibial Insert (PS) - Size 12 / 12mm 0108806373823367172505281060AEDG36 60AEDG36 5/28/2025 Lospa Knee System 01.10.2C4 Lospa Tibial Insert (PS) - Size 12 / 12mm 0108806373823367172602281060AFBP37 60AFBP37 2/28/2026 Lospa Knee System 01.10.2C4 Lospa Tibial Insert (PS) - Size 12 / 12mm 0108806373823367172906281060AJEY38 60AJEY38 6/28/2029 Lospa Knee System 01.10.2C6 Lospa Tibial Insert (PS) - Size 12 / 14mm 0108806373823374172407281060ADF237 60ADF237 7/28/2024 Lospa Knee System 01.10.2C6 Lospa Tibial Insert (PS) - Size 12 / 14mm 0108806373823374172505281060AEDG37 60AEDG37 5/28/2025 Lospa Knee System 01.10.2C8 Lospa Tibial Insert (PS) - Size 12 / 16mm 0108806373823381172505281060AEDG38 60AEDG38 5/28/2025 Lospa Knee System 01.10.2C8 Lospa Tibial Insert (PS) - Size 12 / 16mm 0108806373823381172508281060AEJE42 60AEJE42 8/28/2025 Lospa Knee System 01.10.2C8 Lospa Tibial Insert (PS) - Size 12 / 16mm 0108806373823381172602281060AFBP38 60AFBP38 2/28/2026 Lospa Knee System 01.10.2C8 Lospa Tibial Insert (PS) - Size 12 / 16mm 0108806373823381172604281060AFCM32 60AFCM32 4/28/2026 Lospa Knee System 01.10.2C8 Lospa Tibial Insert (PS) - Size 12 / 16mm 0108806373823381172812281060AHF135 60AHF135 12/28/2028 Lospa Knee System 01.10.2CA Lospa Tibial Insert (PS) - Size 12 / 18mm 0108806373823398172505281060AEDG39 60AEDG39 5/28/2025 Lospa Knee System 01.10.2CA Lospa Tibial Insert (PS) - Size 12 / 18mm 0108806373823398172611281060AFM830 60AFM830 11/28/2026 Lospa Knee System 01.10.2CA Lospa Tibial Insert (PS) - Size 12 / 18mm 0108806373823398172906281060AJEY41 60AJEY41 6/28/2029 Lospa Knee System 01.10.2CC Lospa Tibial Insert (PS) - Size 12 / 20mm 0108806373823404172411281060ACCR35 60ACCR35 11/28/2024 Lospa Knee System 01.10.1E4 Lospa Tibial Insert (CR) - Size 14 / 12mm 0108806373852794172502281010aebg33 10aebg33 2/28/2025 Lospa Knee System 01.10.1E4 Lospa Tibial Insert (CR) - Size 14 / 12mm 0108806373852794172505281010AEDH40 10AEDH40 5/28/2025 Lospa Knee System 01.10.1E6 Lospa Tibial Insert (CR) - Size 14 / 14mm 0108806373852800172606281010AFBP44 10AFBP44 6/28/2026 Lospa Knee System 01.10.1EC Lospa Tibial Insert (CR) - Size 14 / 20mm 0108806373852831172804281010AHDB44 10AHDB44 4/28/2028 Lospa Knee System 01.10.2E2 Lospa Tibial Insert (PS) - Size 14 / 10mm 0108806373852886172510281010AEJE43 10AEJE43 10/28/2025 Lospa Knee System 01.10.2E2 Lospa Tibial Insert (PS) - Size 14 / 10mm 0108806373852886172606281010AFCS30 10AFCS30 6/28/2026 Lospa Knee System 01.10.2E2 Lospa Tibial Insert (PS) - Size 14 / 10mm 0108806373852886172710281010AGBE30 10AGBE30 10/28/2027 Lospa Knee System 01.10.2E4 Lospa Tibial Insert (PS) - Size 14 / 12mm 0108806373852893172410281010ADL931 10ADL931 10/28/2024 Lospa Knee System 01.10.2E4 Lospa Tibial Insert (PS) - Size 14 / 12mm 0108806373852893172606281010AFCS31 10AFCS31 6/28/2026 Lospa Knee System 01.10.2E4 Lospa Tibial Insert (PS) - Size 14 / 12mm 0108806373852893172710281010AGB836 10AGB836 10/28/2027 Lospa Knee System 01.10.2E6 Lospa Tibial Insert (PS) - Size 14 / 14mm 0108806373852909172411281010ADL932 10ADL932 11/28/2024 Lospa Knee System 01.10.2E8 Lospa Tibial Insert (PS) - Size 14 / 16mm 0108806373852916172507281010AEHF45 10AEHF45 7/28/2025 Lospa Knee System 01.10.2E8 Lospa Tibial Insert (PS) - Size 14 / 16mm 0108806373852916172702281010AGB837 10AGB837 2/28/2027 Lospa Knee System 01.10.2E8 Lospa Tibial Insert (PS) - Size 14 / 16mm 0108806373852916172711281010AGB837 10AGB837 11/28/2027 Lospa Knee System 01.10.2EA Lospa Tibial Insert (PS) - Size 14 / 18mm 0108806373852923172410281010ADL934 10ADL934 10/28/2024 Lospa Knee System 01.10.2EA Lospa Tibial Insert (PS) - Size 14 / 18mm 0108806373852923172511281010AEJE47 10AEJE47 11/28/2025 Lospa Knee System 01.10.2EA Lospa Tibial Insert (PS) - Size 14 / 18mm 0108806373852923172702281010AGB838 10AGB838 2/28/2027 Lospa Knee System 01.10.2EC Lospa Tibial Insert (PS) - Size 14 / 20mm 0108806373852930172804281010AHDB32 10AHDB32 4/28/2028 Lospa Knee System 01.10.9EC LOSPA Patella Component 38mm 0108806373853692172503281000AED827 00AED827 3/28/2025 Lospa Knee System 01.10.9EC LOSPA Patella Component 38mm 0108806373853692172603281000AFAM25 00AFAM25 3/28/2026 Lospa Modular Knee System K2.IBN.4061-10A TIBIAL INSERT-FIXED-CONSTRAINED-03-10T 0108806373861161172706281010AGCN30 10AGCN30 6/28/2027 Lospa Modular Knee System K2.IBN.4061-10A TIBIAL INSERT-FIXED-CONSTRAINED-03-10T 0108806373861161172708281010AGHQ33 10AGHQ33 8/28/2027 Lospa Modular Knee System K2.IBN.4061-10A TIBIAL INSERT-FIXED-CONSTRAINED-03-10T 0108806373861161172712281010AGL830 10AGL830 12/28/2027 Lospa Modular Knee System K2.IBN.4061-10A TIBIAL INSERT-FIXED-CONSTRAINED-03-10T 0108806373861161172802281010AHAC38 10AHAC38 2/28/2028 Lospa Modular Knee System K2.IBN.4061-12A TIBIAL INSERT-FIXED-CONSTRAINED-03-12T 0108806373861185172708281010AGHQ34 10AGHQ34 8/28/2027 Lospa Modular Knee System K2.IBN.4061-12A TIBIAL INSERT-FIXED-CONSTRAINED-03-12T 0108806373861185172712281010AGL831 10AGL831 12/28/2027 Lospa Modular Knee System K2.IBN.4061-12A TIBIAL INSERT-FIXED-CONSTRAINED-03-12T 0108806373861185172802281010AHAC39 10AHAC39 2/28/2028 Lospa Modular Knee System K2.IBN.4061-14A TIBIAL INSERT-FIXED-CONSTRAINED-03-14T 0108806373861208172706281010AGCP31 10AGCP31 6/28/2027 Lospa Modular Knee System K2.IBN.4061-14A TIBIAL INSERT-FIXED-CONSTRAINED-03-14T 0108806373861208172712281010AGLA30 10AGLA30 12/28/2027 Lospa Modular Knee System K2.IBN.4061-14A TIBIAL INSERT-FIXED-CONSTRAINED-03-14T 0108806373861208172802281010AHAC40 10AHAC40 2/28/2028 Lospa Modular Knee System K2.IBN.4061-16A TIBIAL INSERT-FIXED-CONSTRAINED-03-16T 0108806373861222172706281010AGCP32 10AGCP32 6/28/2027 Lospa Modular Knee System K2.IBN.4061-16A TIBIAL INSERT-FIXED-CONSTRAINED-03-16T 0108806373861222172708281010AGHQ36 10AGHQ36 8/28/2027 Lospa Modular Knee System K2.IBN.4061-16A TIBIAL INSERT-FIXED-CONSTRAINED-03-16T 0108806373861222172712281010AGLA31 10AGLA31 12/28/2027 Lospa Modular Knee System K2.IBN.4061-16A TIBIAL INSERT-FIXED-CONSTRAINED-03-16T 0108806373861222172801281010AHAC41 10AHAC41 1/28/2028 Lospa Modular Knee System K2.IBN.4061-18A TIBIAL INSERT-FIXED-CONSTRAINED-03-18T 0108806373861246172706281010AGCP33 10AGCP33 6/28/2027 Lospa Modular Knee System K2.IBN.4061-18A TIBIAL INSERT-FIXED-CONSTRAINED-03-18T 0108806373861246172707281010AGHQ30 10AGHQ30 7/28/2027 Lospa Modular Knee System K2.IBN.4061-18A TIBIAL INSERT-FIXED-CONSTRAINED-03-18T 0108806373861246172710281010AGL930 10AGL930 10/28/2027 Lospa Modular Knee System K2.IBN.4061-18A TIBIAL INSERT-FIXED-CONSTRAINED-03-18T 0108806373861246172711281010AGL930 10AGL930 11/28/2027 Lospa Modular Knee System K2.IBN.4061-18A TIBIAL INSERT-FIXED-CONSTRAINED-03-18T 0108806373861246172802281010AHAC42 10AHAC42 2/28/2028 Lospa Modular Knee System K2.IBN.4061-20A TIBIAL INSERT-FIXED-CONSTRAINED-03-20T 0108806373861260172706281010AGG430 10AGG430 6/28/2027 Lospa Modular Knee System K2.IBN.4061-20A TIBIAL INSERT-FIXED-CONSTRAINED-03-20T 0108806373861260172708281010AGJ130 10AGJ130 8/28/2027 Lospa Modular Knee System K2.IBN.4061-20A TIBIAL INSERT-FIXED-CONSTRAINED-03-20T 0108806373861260172712281010AGLA32 10AGLA32 12/28/2027 Lospa Modular Knee System K2.IBN.4061-23A TIBIAL INSERT-FIXED-CONSTRAINED-03-23T 0108806373861291172809281010AHHN30 10AHHN30 9/28/2028 Lospa Modular Knee System K2.IBN.4061-26A TIBIAL INSERT-FIXED-CONSTRAINED-03-26T 0108806373861321172809281010AHHN31 10AHHN31 9/28/2028 Lospa Modular Knee System K2.IBN.4264-10A TIBIAL INSERT-FIXED-CONSTRAINED-04-10T 0108806373861352172708281010AGHQ37 10AGHQ37 8/28/2027 Lospa Modular Knee System K2.IBN.4264-10A TIBIAL INSERT-FIXED-CONSTRAINED-04-10T 0108806373861352172712281010AGL832 10AGL832 12/28/2027 Lospa Modular Knee System K2.IBN.4264-10A TIBIAL INSERT-FIXED-CONSTRAINED-04-10T 0108806373861352172803281010AHAC44 10AHAC44 3/28/2028 Lospa Modular Knee System K2.IBN.4264-12A TIBIAL INSERT-FIXED-CONSTRAINED-04-12T 0108806373861376172706281010AGCP35 10AGCP35 6/28/2027 Lospa Modular Knee System K2.IBN.4264-12A TIBIAL INSERT-FIXED-CONSTRAINED-04-12T 0108806373861376172708281010AGHQ38 10AGHQ38 8/28/2027 Lospa Modular Knee System K2.IBN.4264-12A TIBIAL INSERT-FIXED-CONSTRAINED-04-12T 0108806373861376172712281010AGL833 10AGL833 12/28/2027 Lospa Modular Knee System K2.IBN.4264-14A TIBIAL INSERT-FIXED-CONSTRAINED-04-14T 0108806373861390172708281010AGHQ39 10AGHQ39 8/28/2027 Lospa Modular Knee System K2.IBN.4264-14A TIBIAL INSERT-FIXED-CONSTRAINED-04-14T 0108806373861390172712281010AGL834 10AGL834 12/28/2027 Lospa Modular Knee System K2.IBN.4264-16A TIBIAL INSERT-FIXED-CONSTRAINED-04-16T 0108806373861413172708281010AGHQ40 10AGHQ40 8/28/2027 Lospa Modular Knee System K2.IBN.4264-16A TIBIAL INSERT-FIXED-CONSTRAINED-04-16T 0108806373861413172712281010AGL835 10AGL835 12/28/2027 Lospa Modular Knee System K2.IBN.4264-18A TIBIAL INSERT-FIXED-CONSTRAINED-04-18T 0108806373861437172706281010AGCP38 10AGCP38 6/28/2027 Lospa Modular Knee System K2.IBN.4264-18A TIBIAL INSERT-FIXED-CONSTRAINED-04-18T 0108806373861437172708281010AGHQ41 10AGHQ41 8/28/2027 Lospa Modular Knee System K2.IBN.4264-18A TIBIAL INSERT-FIXED-CONSTRAINED-04-18T 0108806373861437172712281010AGL836 10AGL836 12/28/2027 Lospa Modular Knee System K2.IBN.4264-18A TIBIAL INSERT-FIXED-CONSTRAINED-04-18T 0108806373861437172804281010AHAD30 10AHAD30 4/28/2028 Lospa Modular Knee System K2.IBN.4264-20A TIBIAL INSERT-FIXED-CONSTRAINED-04-20T 0108806373861451172706281010AGCP39 10AGCP39 6/28/2027 Lospa Modular Knee System K2.IBN.4264-20A TIBIAL INSERT-FIXED-CONSTRAINED-04-20T 0108806373861451172708281010AGJ131 10AGJ131 8/28/2027 Lospa Modular Knee System K2.IBN.4264-20A TIBIAL INSERT-FIXED-CONSTRAINED-04-20T 0108806373861451172712281010AGLA33 10AGLA33 12/28/2027 Lospa Modular Knee System K2.IBN.4264-23A TIBIAL INSERT-FIXED-CONSTRAINED-04-23T 0108806373861482172809281010AHHN32 10AHHN32 9/28/2028 Lospa Modular Knee System K2.IBN.4264-26A TIBIAL INSERT-FIXED-CONSTRAINED-04-26T 0108806373861512172708281010AGJF42 10AGJF42 8/28/2027 Lospa Modular Knee System K2.IBN.4264-26A TIBIAL INSERT-FIXED-CONSTRAINED-04-26T 0108806373861512172809281010AHHN33 10AHHN33 9/28/2028 Lospa Modular Knee System K2.IBN.4467-10A TIBIAL INSERT-FIXED-CONSTRAINED-05-10T 0108806373861543172706281010AGCN32 10AGCN32 6/28/2027 Lospa Modular Knee System K2.IBN.4467-10A TIBIAL INSERT-FIXED-CONSTRAINED-05-10T 0108806373861543172711281010AGL837 10AGL837 11/28/2027 Lospa Modular Knee System K2.IBN.4467-12A TIBIAL INSERT-FIXED-CONSTRAINED-05-12T 0108806373861567172706281010AGCP40 10AGCP40 6/28/2027 Lospa Modular Knee System K2.IBN.4467-12A TIBIAL INSERT-FIXED-CONSTRAINED-05-12T 0108806373861567172708281010AGHQ43 10AGHQ43 8/28/2027 Lospa Modular Knee System K2.IBN.4467-14A TIBIAL INSERT-FIXED-CONSTRAINED-05-14T 0108806373861581172706281010AGCP41 10AGCP41 6/28/2027 Lospa Modular Knee System K2.IBN.4467-14A TIBIAL INSERT-FIXED-CONSTRAINED-05-14T 0108806373861581172711281010AGL839 10AGL839 11/28/2027 Lospa Modular Knee System K2.IBN.4467-16A TIBIAL INSERT-FIXED-CONSTRAINED-05-16T 0108806373861604172711281010AGL840 10AGL840 11/28/2027 Lospa Modular Knee System K2.IBN.4467-18A TIBIAL INSERT-FIXED-CONSTRAINED-05-18T 0108806373861628172706281010AGCP43 10AGCP43 6/28/2027 Lospa Modular Knee System K2.IBN.4467-18A TIBIAL INSERT-FIXED-CONSTRAINED-05-18T 0108806373861628172711281010AGLA34 10AGLA34 11/28/2027 Lospa Modular Knee System K2.IBN.4467-18A TIBIAL INSERT-FIXED-CONSTRAINED-05-18T 0108806373861628172801281010AHAD36 10AHAD36 1/28/2028 Lospa Modular Knee System K2.IBN.4467-20A TIBIAL INSERT-FIXED-CONSTRAINED-05-20T 0108806373861642172706281010AGCP44 10AGCP44 6/28/2027 Lospa Modular Knee System K2.IBN.4467-20A TIBIAL INSERT-FIXED-CONSTRAINED-05-20T 0108806373861642172709281010AGJ132 10AGJ132 9/28/2027 Lospa Modular Knee System K2.IBN.4467-20A TIBIAL INSERT-FIXED-CONSTRAINED-05-20T 0108806373861642172711281010AGL931 10AGL931 11/28/2027 Lospa Modular Knee System K2.IBN.4467-23A TIBIAL INSERT-FIXED-CONSTRAINED-05-23T 0108806373861673172809281050AHHN34 50AHHN34 9/28/2028 Lospa Modular Knee System K2.IBN.4467-26A TIBIAL INSERT-FIXED-CONSTRAINED-05-26T 0108806373861703172708281010AGJF45 10AGJF45 8/28/2027 Lospa Modular Knee System K2.IBN.4467-26A TIBIAL INSERT-FIXED-CONSTRAINED-05-26T 0108806373861703172809281050AGJF45 50AGJF45 9/28/2028 Lospa Modular Knee System K2.IBN.4467-26A TIBIAL INSERT-FIXED-CONSTRAINED-05-26T 0108806373861703172809281050AHHN35 50AHHN35 9/28/2028 Lospa Modular Knee System K2.IBN.4672-10A TIBIAL INSERT-FIXED-CONSTRAINED-07-10T 0108806373861925172711281010AGL841 10AGL841 11/28/2027 Lospa Modular Knee System K2.IBN.4672-12A TIBIAL INSERT-FIXED-CONSTRAINED-07-12T 0108806373861949172711281010AGLY31 10AGLY31 11/28/2027 Lospa Modular Knee System K2.IBN.4672-12A TIBIAL INSERT-FIXED-CONSTRAINED-07-12T 0108806373861949172804281010AHAD39 10AHAD39 4/28/2028 Lospa Modular Knee System K2.IBN.4672-14A TIBIAL INSERT-FIXED-CONSTRAINED-07-14T 0108806373861963172708281010AGHR37 10AGHR37 8/28/2027 Lospa Modular Knee System K2.IBN.4672-16A TIBIAL INSERT-FIXED-CONSTRAINED-07-16T 0108806373861987172706281010AGCP47 10AGCP47 6/28/2027 Lospa Modular Knee System K2.IBN.4672-16A TIBIAL INSERT-FIXED-CONSTRAINED-07-16T 0108806373861987172708281010AGHR38 10AGHR38 8/28/2027 Lospa Modular Knee System K2.IBN.4672-16A TIBIAL INSERT-FIXED-CONSTRAINED-07-16T 0108806373861987172712281010AGL844 10AGL844 12/28/2027 Lospa Modular Knee System K2.IBN.4672-16A TIBIAL INSERT-FIXED-CONSTRAINED-07-16T 0108806373861987172808281010AHAD41 10AHAD41 8/28/2028 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172706281010AGCP48 10AGCP48 6/28/2027 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172709281010AGJF30 10AGJF30 9/28/2027 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172712281010AGLA35 10AGLA35 12/28/2027 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172806281010AHAD42 10AHAD42 6/28/2028 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172901281010AHAD42 10AHAD42 1/28/2029 Lospa Modular Knee System K2.IBN.4672-18A TIBIAL INSERT-FIXED-CONSTRAINED-07-18T 0108806373862007172906281010AHAD42 10AHAD42 6/28/2029 Lospa Modular Knee System K2.IBN.4672-20A TIBIAL INSERT-FIXED-CONSTRAINED-07-20T 0108806373862021172706281010AGCP49 10AGCP49 6/28/2027 Lospa Modular Knee System K2.IBN.4672-20A TIBIAL INSERT-FIXED-CONSTRAINED-07-20T 0108806373862021172708281010AGJ133 10AGJ133 8/28/2027 Lospa Modular Knee System K2.IBN.4672-20A TIBIAL INSERT-FIXED-CONSTRAINED-07-20T 0108806373862021172712281010AGLA36 10AGLA36 12/28/2027 Lospa Modular Knee System K2.IBN.4672-23A TIBIAL INSERT-FIXED-CONSTRAINED-07-23T 0108806373862052172809281010AHHN36 10AHHN36 9/28/2028 Lospa Modular Knee System K2.IBN.4672-26A TIBIAL INSERT-FIXED-CONSTRAINED-07-26T 0108806373862083172809281010AHHN37 10AHHN37 9/28/2028 Lospa Modular Knee System K2.IBN.4874-10A TIBIAL INSERT-FIXED-CONSTRAINED-09-10T 0108806373862304172708281010AGHS30 10AGHS30 8/28/2027 Lospa Modular Knee System K2.IBN.4874-10A TIBIAL INSERT-FIXED-CONSTRAINED-09-10T 0108806373862304172712281010AGL845 10AGL845 12/28/2027 Lospa Modular Knee System K2.IBN.4874-10A TIBIAL INSERT-FIXED-CONSTRAINED-09-10T 0108806373862304172806281010AHAD44 10AHAD44 6/28/2028 Lospa Modular Knee System K2.IBN.4874-12A TIBIAL INSERT-FIXED-CONSTRAINED-09-12T 0108806373862328172708281010AGHS31 10AGHS31 8/28/2027 Lospa Modular Knee System K2.IBN.4874-12A TIBIAL INSERT-FIXED-CONSTRAINED-09-12T 0108806373862328172711281010AGLA37 10AGLA37 11/28/2027 Lospa Modular Knee System K2.IBN.4874-14A TIBIAL INSERT-FIXED-CONSTRAINED-09-14T 0108806373862342172706281010AGCQ31 10AGCQ31 6/28/2027 Lospa Modular Knee System K2.IBN.4874-14A TIBIAL INSERT-FIXED-CONSTRAINED-09-14T 0108806373862342172708281010AGHS32 10AGHS32 8/28/2027 Lospa Modular Knee System K2.IBN.4874-14A TIBIAL INSERT-FIXED-CONSTRAINED-09-14T 0108806373862342172711281010AGLA38 10AGLA38 11/28/2027 Lospa Modular Knee System K2.IBN.4874-14A TIBIAL INSERT-FIXED-CONSTRAINED-09-14T 0108806373862342172801281010AHAD46 10AHAD46 1/28/2028 Lospa Modular Knee System K2.IBN.4874-16A TIBIAL INSERT-FIXED-CONSTRAINED-09-16T 0108806373862366172708281010AGHS33 10AGHS33 8/28/2027 Lospa Modular Knee System K2.IBN.4874-16A TIBIAL INSERT-FIXED-CONSTRAINED-09-16T 0108806373862366172711281010AGLA39 10AGLA39 11/28/2027 Lospa Modular Knee System K2.IBN.4874-18A TIBIAL INSERT-FIXED-CONSTRAINED-09-18T 0108806373862380172707281010AGHQ31 10AGHQ31 7/28/2027 Lospa Modular Knee System K2.IBN.4874-18A TIBIAL INSERT-FIXED-CONSTRAINED-09-18T 0108806373862380172711281010AGLA40 10AGLA40 11/28/2027 Lospa Modular Knee System K2.IBN.4874-18A TIBIAL INSERT-FIXED-CONSTRAINED-09-18T 0108806373862380172901281010AHAE30 10AHAE30 1/28/2029 Lospa Modular Knee System K2.IBN.4874-18A TIBIAL INSERT-FIXED-CONSTRAINED-09-18T 0108806373862380172906281010AHAE30 10AHAE30 6/28/2029 Lospa Modular Knee System K2.IBN.4874-20A TIBIAL INSERT-FIXED-CONSTRAINED-09-20T 0108806373862403172706281010AGCQ34 10AGCQ34 6/28/2027 Lospa Modular Knee System K2.IBN.4874-20A TIBIAL INSERT-FIXED-CONSTRAINED-09-20T 0108806373862403172708281010AGJ134 10AGJ134 8/28/2027 Lospa Modular Knee System K2.IBN.4874-20A TIBIAL INSERT-FIXED-CONSTRAINED-09-20T 0108806373862403172712281010AGLA41 10AGLA41 12/28/2027 Lospa Modular Knee System K2.IBN.4874-23A TIBIAL INSERT-FIXED-CONSTRAINED-09-23T 0108806373862434172809281010AHHN38 10AHHN38 9/28/2028 Lospa Modular Knee System K2.IBN.4874-26A TIBIAL INSERT-FIXED-CONSTRAINED-09-26T 0108806373862465172809281010AHHN39 10AHHN39 9/28/2028 Lospa Modular Knee System K2.IBN.5076-10A TIBIAL INSERT-FIXED-CONSTRAINED-11-10T 0108806373862687172708281010AGHQ82 10AGHQ82 8/28/2027 Lospa Modular Knee System K2.IBN.5076-10A TIBIAL INSERT-FIXED-CONSTRAINED-11-10T 0108806373862687172710281010AGL932 10AGL932 10/28/2027 Lospa Modular Knee System K2.IBN.5076-10A TIBIAL INSERT-FIXED-CONSTRAINED-11-10T 0108806373862687172802281010AHAE32 10AHAE32 2/28/2028 Lospa Modular Knee System K2.IBN.5076-12A TIBIAL INSERT-FIXED-CONSTRAINED-11-12T 0108806373862700172706281010AGCQ35 10AGCQ35 6/28/2027 Lospa Modular Knee System K2.IBN.5076-12A TIBIAL INSERT-FIXED-CONSTRAINED-11-12T 0108806373862700172708281010AGHS34 10AGHS34 8/28/2027 Lospa Modular Knee System K2.IBN.5076-12A TIBIAL INSERT-FIXED-CONSTRAINED-11-12T 0108806373862700172711281010AGLA42 10AGLA42 11/28/2027 Lospa Modular Knee System K2.IBN.5076-12A TIBIAL INSERT-FIXED-CONSTRAINED-11-12T 0108806373862700172802281010AHAE33 10AHAE33 2/28/2028 Lospa Modular Knee System K2.IBN.5076-14A TIBIAL INSERT-FIXED-CONSTRAINED-11-14T 0108806373862724172706281010AGCQ36 10AGCQ36 6/28/2027 Lospa Modular Knee System K2.IBN.5076-14A TIBIAL INSERT-FIXED-CONSTRAINED-11-14T 0108806373862724172708281010AGHS35 10AGHS35 8/28/2027 Lospa Modular Knee System K2.IBN.5076-14A TIBIAL INSERT-FIXED-CONSTRAINED-11-14T 0108806373862724172712281010AGLA43 10AGLA43 12/28/2027 Lospa Modular Knee System K2.IBN.5076-14A TIBIAL INSERT-FIXED-CONSTRAINED-11-14T 0108806373862724172803281010AHAE34 10AHAE34 3/28/2028 Lospa Modular Knee System K2.IBN.5076-16A TIBIAL INSERT-FIXED-CONSTRAINED-11-16T 0108806373862748172706281010AGCQ37 10AGCQ37 6/28/2027 Lospa Modular Knee System K2.IBN.5076-16A TIBIAL INSERT-FIXED-CONSTRAINED-11-16T 0108806373862748172708281010AGHS36 10AGHS36 8/28/2027 Lospa Modular Knee System K2.IBN.5076-16A TIBIAL INSERT-FIXED-CONSTRAINED-11-16T 0108806373862748172711281010AGLA44 10AGLA44 11/28/2027 Lospa Modular Knee System K2.IBN.5076-16A TIBIAL INSERT-FIXED-CONSTRAINED-11-16T 0108806373862748172803281010AHAE35 10AHAE35 3/28/2028 Lospa Modular Knee System K2.IBN.5076-18A TIBIAL INSERT-FIXED-CONSTRAINED-11-18T 0108806373862762172706281010AGCQ38 10AGCQ38 6/28/2027 Lospa Modular Knee System K2.IBN.5076-18A TIBIAL INSERT-FIXED-CONSTRAINED-11-18T 0108806373862762172708281010AGHS37 10AGHS37 8/28/2027 Lospa Modular Knee System K2.IBN.5076-18A TIBIAL INSERT-FIXED-CONSTRAINED-11-18T 0108806373862762172710281010AGL933 10AGL933 10/28/2027 Lospa Modular Knee System K2.IBN.5076-18A TIBIAL INSERT-FIXED-CONSTRAINED-11-18T 0108806373862762172803281010AGL933 10AGL933 3/28/2028 Lospa Modular Knee System K2.IBN.5076-18A TIBIAL INSERT-FIXED-CONSTRAINED-11-18T 0108806373862762172803281010AHAE36 10AHAE36 3/28/2028 Lospa Modular Knee System K2.IBN.5076-20A TIBIAL INSERT-FIXED-CONSTRAINED-11-20T 0108806373862786172709281010AGJ135 10AGJ135 9/28/2027 Lospa Modular Knee System K2.IBN.5076-20A TIBIAL INSERT-FIXED-CONSTRAINED-11-20T 0108806373862786172711281010AGLA45 10AGLA45 11/28/2027 Lospa Modular Knee System K2.IBN.5076-20A TIBIAL INSERT-FIXED-CONSTRAINED-11-20T 0108806373862786172803281010AHAE37 10AHAE37 3/28/2028 Lospa Modular Knee System K2.IBN.5076-23A TIBIAL INSERT-FIXED-CONSTRAINED-11-23T 0108806373862816172809281010AHHN40 10AHHN40 9/28/2028 Lospa Modular Knee System K2.IBN.5076-26A TIBIAL INSERT-FIXED-CONSTRAINED-11-26T 0108806373862847172809281010AHHN41 10AHHN41 9/28/2028 Lospa Modular Knee System K2.IBN.5281-10A TIBIAL INSERT-FIXED-CONSTRAINED-12-10T 0108806373862878172711281010AGLP30 10AGLP30 11/28/2027 Lospa Modular Knee System K2.IBN.5281-10A TIBIAL INSERT-FIXED-CONSTRAINED-12-10T 0108806373862878172806281010AHAE38 10AHAE38 6/28/2028 Lospa Modular Knee System K2.IBN.5281-12A TIBIAL INSERT-FIXED-CONSTRAINED-12-12T 0108806373862892172711281010AGLP31 10AGLP31 11/28/2027 Lospa Modular Knee System K2.IBN.5281-12A TIBIAL INSERT-FIXED-CONSTRAINED-12-12T 0108806373862892172806281010AHAE39 10AHAE39 6/28/2028 Lospa Modular Knee System K2.IBN.5281-14A TIBIAL INSERT-FIXED-CONSTRAINED-12-14T 0108806373862915172711281010AGLP32 10AGLP32 11/28/2027 Lospa Modular Knee System K2.IBN.5281-14A TIBIAL INSERT-FIXED-CONSTRAINED-12-14T 0108806373862915172806281010AHAE40 10AHAE40 6/28/2028 Lospa Modular Knee System K2.IBN.5281-16A TIBIAL INSERT-FIXED-CONSTRAINED-12-16T 0108806373862939172711281010AGLP33 10AGLP33 11/28/2027 Lospa Modular Knee System K2.IBN.5281-16A TIBIAL INSERT-FIXED-CONSTRAINED-12-16T 0108806373862939172806281010AHAE41 10AHAE41 6/28/2028 Lospa Modular Knee System K2.IBN.5281-18A TIBIAL INSERT-FIXED-CONSTRAINED-12-18T 0108806373862953172711281010AGLP34 10AGLP34 11/28/2027 Lospa Modular Knee System K2.IBN.5281-20A TIBIAL INSERT-FIXED-CONSTRAINED-12-20T 0108806373862977172711281010AGLP35 10AGLP35 11/28/2027 Lospa Modular Knee System K2.IBN.5281-20A TIBIAL INSERT-FIXED-CONSTRAINED-12-20T 0108806373862977172806281010AHAE43 10AHAE43 6/28/2028 Lospa Modular Knee System K2.IBN.5281-23A TIBIAL INSERT-FIXED-CONSTRAINED-12-23T 0108806373863004172809281010AHHN42 10AHHN42 9/28/2028 Lospa Modular Knee System K2.IBN.5281-26A TIBIAL INSERT-FIXED-CONSTRAINED-12-26T 0108806373863035172809281010AHHN43 10AHHN43 9/28/2028 Lospa Modular Knee System K2.IBN.5684-10A TIBIAL INSERT-FIXED-CONSTRAINED-14-10T 0108806373863257172711281010AGLP36 10AGLP36 11/28/2027 Lospa Modular Knee System K2.IBN.5684-12A TIBIAL INSERT-FIXED-CONSTRAINED-14-12T 0108806373863271172711281010AGLP37 10AGLP37 11/28/2027 Lospa Modular Knee System K2.IBN.5684-14A TIBIAL INSERT-FIXED-CONSTRAINED-14-14T 0108806373863295172711281010AGLP38 10AGLP38 11/28/2027 Lospa Modular Knee System K2.IBN.5684-16A TIBIAL INSERT-FIXED-CONSTRAINED-14-16T 0108806373863318172711281010AGLP39 10AGLP39 11/28/2027 Lospa Modular Knee System K2.IBN.5684-18A TIBIAL INSERT-FIXED-CONSTRAINED-14-18T 0108806373863332172711281010AGLP40 10AGLP40 11/28/2027 Lospa Modular Knee System K2.IBN.5684-20A TIBIAL INSERT-FIXED-CONSTRAINED-14-20T 0108806373863356172711281010AGLP41 10AGLP41 11/28/2027 Lospa Modular Knee System K2.IBN.5684-23A TIBIAL INSERT-FIXED-CONSTRAINED-14-23T 0108806373863387172809281010AHHN44 10AHHN44 9/28/2028 Lospa Modular Knee System K2.IBN.5684-26A TIBIAL INSERT-FIXED-CONSTRAINED-14-26T 0108806373863417172809281010AHHN45 10AHHN45 9/28/2028 Lospa Knee System 01.10.1G2 Lospa Tibial Insert (CR) - Size 16 / 10mm 0108806373868580172706281010AFDF35 10AFDF35 6/28/2027 Lospa Knee System 01.10.1G4 Lospa Tibial Insert (CR) - Size 16 / 12mm 0108806373868597172706281010AFDF36 10AFDF36 6/28/2027 Lospa Knee System 01.10.1G6 Lospa Tibial Insert (CR) - Size 16 / 14mm 0108806373868603172706281010AFDF37 10AFDF37 6/28/2027 Lospa Knee System 01.10.1G8 Lospa Tibial Insert (CR) - Size 16 / 16mm 0108806373868610172706281010AFDF38 10AFDF38 6/28/2027 Lospa Knee System 01.10.1GA Lospa Tibial Insert (CR) - Size 16 / 18mm 0108806373868627172706281010AFDF39 10AFDF39 6/28/2027 Lospa Knee System 01.10.2G2 Lospa Tibial Insert (PS) - Size 16 / 10mm 0108806373868689172706281010AFDF30 10AFDF30 6/28/2027 Lospa Knee System 01.10.2G4 Lospa Tibial Insert (PS) - Size 16 / 12mm 0108806373868696172706281010AFDF31 10AFDF31 6/28/2027 Lospa Knee System 01.10.2G6 Lospa Tibial Insert (PS) - Size 16 / 14mm 0108806373868702172706281010AFDF32 10AFDF32 6/28/2027 Lospa Knee System 01.10.2G8 Lospa Tibial Insert (PS) - Size 16 / 16mm 0108806373868719172706281010AFDF33 10AFDF33 6/28/2027 Lospa Knee System 01.10.2GA Lospa Tibial Insert (PS) - Size 16 / 18mm 0108806373868726172706281010AFDF34 10AFDF34 6/28/2027
Other devices under the same FDA submission
2 devices cleared under K110404, K130673, K160157 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LOSPA Total Knee System 01.10.007Corentec Co., Ltd · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 report
- LOSPA Total Knee System 01.10.064Corentec Co., Ltd · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 94380).
- Z-1935-2024 · Open, ClassifiedLOSPA Patella Component Model/Catalog Number: 01.10.9XX Software Version: N/A Product Description: Patella
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page