RayStation 11.0.0.951, 11.0.1.29, 11.0.3.116 & 11.0.4.15.… recallZ-1748-2024 · Raysearch Laboratories AB · initiated Mar 28, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for reported SSD to be too high.
FDA root cause category: Software design
What the company is doing
A Field Safety Notice for Medical Device Correction #133261, dated 3/28/24, was sent to consignees via email. Consignees were provided the instructions to use the Measure Tool to measure the distance in the patient view to make sure the reported SSD value is correct if it is to be used for patient setup or other critical tasks. If the SSD value is found to be incorrect, small changes to the isocenter, beam angles or the affected ROI's geometry can be made until the correct value is achieved. Consignees are to inform planning staff and all users of the workarounds provided in the Field Safety Notice. Consignees are to inspect their product and identify if installed units possess the affected software versions. Consignees are to reply to the recall notification confirming that they read and understood the Field Safety Notificaiton. RaySearch is to deploy a new version of RayStation/RayPlan to correct the issue, to be released in April 2024. The provided Field Safety Notice is to be maintained with affected units. Consignees with any questions are to contact [email protected].
Recall details
- Recall number
- Z-1748-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 28, 2024
- Posted by the FDA
- May 1, 2024
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 10 units
- Product code
- MUJ
- FDA submissions
- K211867
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK.
Product description
RayStation 11.0.0.951, 11.0.1.29, 11.0.3.116 & 11.0.4.15. Radiation Therapy Treatment Planning System.
Code information (models, lots, serial numbers)
UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916 and 0735000201063120220616; GTIN: 07350002010389, 07350002010433, 07350002010440 and 07350002010631; Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15; Expiration Date: 2027-06-22.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- RayStation 11.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
- RayStation 11.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment7 recalls0 reports
- RayStation 11A SP2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment5 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94388).
- Z-1741-2024 · Open, ClassifiedRayStation 7.0.0.19. Radiation Therapy Treatment Planning System.
- Z-1742-2024 · Open, ClassifiedRayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.
- Z-1743-2024 · Open, ClassifiedRayStation 8.1.0.47, 8.1.1.8 and 8.1.2.5. Radiation Therapy Treatment Planning System.
- Z-1744-2024 · Open, ClassifiedRayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.
- Z-1745-2024 · Open, ClassifiedRayStation 9.1.0.933 and 9.2.0.483. Radiation Therapy Treatment Planning System.
- Z-1746-2024 · Open, ClassifiedRayStation 10.0.0.1154, 10.0.1.52 & 10.0.2.10. Radiation Therapy Treatment Planning System.
- Z-1747-2024 · Open, ClassifiedRayStation 10.1.0.613 and 10.1.1.54. Radiation Therapy Treatment Planning System.
- Z-1749-2024 · Open, ClassifiedRayStation 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 &12.3.0.119. Radiation Therapy Treatment Planning System.
- Z-1750-2024 · Open, ClassifiedRayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
- Z-1751-2024 · Open, ClassifiedRayStation 14.0.0.3338. Radiation Therapy Treatment Planning System.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page