Novalung sensor box is a component of the Novalung Console… recallZ-1822-2024 · Fresenius Medical Care Holdings, Inc. · initiated Apr 8, 2024
Class II Open, Classified All recalls posted in May 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.
FDA root cause category: Device Design
What the company is doing
Fresenius Medical issued Urgent Medical Device Correction Letter (FA-2021-33-C_REV B) on 4/8/24. Letter states reason for recall, health risk and action to take: If you experience either failure codes (#206/#208) your machine will continue to operate. The essential performance is not affected. It is therefore important to note that the adjusted pump speed will be maintained by the system. The intended use of the Novalung system is to provide respiratory and cardiopulmonary support by extracorporeal circulation and physiologic gas exchange is not impacted by the flow measurement error messages To correct this issue, FMCRTG will update the sensor box on your Novalung console. An FMCRTG Field Service Technician will be contacting you to schedule this correction. In the event 206/208 alarms occur, you may silence the alarms intermittently using the audio pause button. If you have continued alarms, it is recommended to replace either the sensor box or the entire console if a secondary or backup console is available. Reference section 6.2 of the Instructions for Use (Part Number US-003-2019-08). For questions or concerns, please contact your local sales representative or FMCRTG Technical Support Services at 1-800-405-1321.
Recall details
- Recall number
- Z-1822-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 8, 2024
- Posted by the FDA
- May 15, 2024
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 88 units
- Product code
- QJZ
- FDA submissions
- K191407
- Distribution
- US Nationwide distribution.
Product description
Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation and CO2 removal) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure Product Code: F30000163
Code information (models, lots, serial numbers)
UDI-DI: 4057224 (+ serial number) Serial Numbers:XCONUS0001 to XCONUS0124
Other devices under the same FDA submission
2 devices cleared under K191407 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Novalung System XLung kitXenios AG · Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure6 deaths150 reports
- Novalung System Novalung consoleXenios AG · Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure7 deaths66 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page