IPS e.max ZirCAD recallZ-1706-2024 · Ivoclar Vivadent AG · initiated Apr 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ivoclar Vivadent AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Defective blocks may show cracks after sintering If cracks are unnoticed, restorations may fracture in the patient's mouth and require restoration
FDA root cause category: Under Investigation by firm
What the company is doing
Ivokar issued Urgent Use and Safety Notice (03-2024-003-R) Letter to Distributors/End users via Fed'X or Email on 4/05/24. Letter states reason for recall, heath risk and action to take: 1. Identification of the IPS e.max ZirCAD Prime CEREC/inLab, C17 Block batches affected by the recall. 2. Separate the blocks so that they can no longer be used for the fabrication of dental restorations. 3. Contact the Customer Service of the distributor from whom you purchased the affected IPS e.max ZirCAD Prime CEREC/inLab, C17 Blocks to announce the return of the devices. Please note on the delivery bill Recall Ivoclar Vivadent AG (03-2024-003-R) . Your distributor will collect the devices affected by the recall and return them to Ivoclar. 4. In addition, please complete the Customer Reply Form in the attachment and send the completed form to [email protected]. 1. Queries regarding the Risk Minimising Safety Measures described in this Field Safety Notice should be addressed to: Responsible Department at Manufacturer of the Devices: Global Vigilance Email [email protected] Postal Address Ivoclar Vivadent AG Customer Care Global Vigilance Bendererstrasse 2 FL-9494 Schaan Liechtenstein All queries by email will be answered within one working day (24 hours). If you have given the products to a third party, please forward a copy of this information or inform Global Vigilance via [email protected]. 2. Queries regarding replacement for the IPS e.max ZirCAD Prime CEREC/inLab, C17 Block affected by this field safety notice please contact the customer service of your distributor of the Ivoclar Vivadent AG device. We will arrange all refunds with your responsible distributor. Your distributor will check the availability of the replacement material with Ivoclar Partner Support and let you know when the replacement material will be shipped to you.
Recall details
- Recall number
- Z-1706-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 5, 2024
- Posted by the FDA
- Apr 30, 2024
- Recalling firm
- Ivoclar Vivadent AG
- Quantity
- 12525 units
- Product code
- EIH
- FDA submissions
- K051705
- Distribution
- US Nationwide distribution.
Product description
IPS e.max ZirCAD CER/in. Prime A3 C17/5 -A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth Article Number: 758439
Code information (models, lots, serial numbers)
UDI-DI: (01)07615208438659 (11)231117 (10)Z063MT (241)758439; (01)07615208438659 (11)231214 (10)Z067JF (241)758439; (01)07615208438659 (11)231214 (10)Z067JC (241)758439; (01)07615208438659 (11)231214 (10)Z067JG (241)758439; (01)07615208438659 (11)231214 (10)Z067JB (241)758439; (01)07615208438659 (11)240105 (10)Z06B6Z (241)758439; (01)07615208438659 (11)240112 (10)Z06CBG (241)758439 Batch No. Z063MT, Z067JF, Z067JC, Z067JG, Z067JB, Z06B6Z, Z06CBG
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IPS e.max ZirCAD 758439Ivoclar Vivadent Aktiengesellschaft · POWDER, PORCELAIN1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94401).
- Z-1703-2024 · Open, ClassifiedIPS e.max ZirCAD CER/in. Prime BL1 C17/3- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758434
- Z-1704-2024 · Open, ClassifiedIPS e.max ZirCAD CER/in. Prime A1 C17/5 - A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758437
- Z-1705-2024 · Open, ClassifiedIPS e.max ZirCAD CER/in. Prime A2 C17/5 -A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758438
- Z-1707-2024 · Open, ClassifiedIPS e.max ZirCAD CER/in. Prime B1 C17/5 -A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758441
- Z-1708-2024 · Open, ClassifiedIPS e.max ZirCAD CER/in. Prime C2 C17/5- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758443
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page