Skip to content

Medline medical procedure kits, labeled as follows recallZ-1918-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Mar 1, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Trumpet Needle Guide ring can detach when excessive pressure is applied.

FDA root cause category: Process control

What the company is doing

Medline issued a correction notice to its consignees on 03/01/2024 via email and USPS first class mail. The notice explained the issue and requested the consignee identify and quarantine all affected products, and respond to the notice. Once the response was received Medline would provide labels to place over the affected product in stock. If the product was further distributed the party which distributed the product was directed to notify their customers.

Recall details

Recall number
Z-1918-2024
Status
Open, Classified
Initiated by the company
Mar 1, 2024
Posted by the FDA
May 29, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
12994 units
Product code
OJH
Distribution
US Nationwide distribution.

Product description

Medline medical procedure kits, labeled as follows: a) PARACERVICAL PUD TRAY, REF DYNDH1832; b) AHS PARACERVICAL/PUDENDAL KIT, REF DYNJRA2210; c) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046; d) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046H; Medical procedure convenience tray

Code information (models, lots, serial numbers)

a) DYNDH1832: UDI/DI 40195327330713 (case), 10195327330712 (unit), Lot Numbers: 23KBI682, 23KBT359, 23KBX629; b) DYNJRA2210: UDI/DI 40195327424047 (case), 10195327424046 (unit), Lot Numbers: 23JBX245, 23LBN930; c) DYNJRA9046: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23JBX368, 23JBX369, 23JBX370, 23LBA896, 23LBH751, 23LBL612, 23LBV325, 23LBV326; d) DYNJRA9046H: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23LBL612, 23LBV325;

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page