Reagent Test Strips for Blood, Urobilinogen, Bilirubin… recallZ-1834-2024 · Macherey Nagel Gmbh & Co. Kg · initiated Apr 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Macherey Nagel Gmbh & Co. Kg or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec 11-Way, #102-2285, LOT 65402.
FDA root cause category: Under Investigation by firm
What the company is doing
On April4, 2024 URGENT FIELD SAFETY NOTICE letters were sent to direct consignees instructing them to do the following: For distributors: 1. The product named in this safety information must not longer be sold. 2. Confirm the receipt of the information with the attached confirmation of receipt form by April 19, 2024. 3. Pass this safety information on to customers of the affected products. 4. Have the receipt of the information confirmed with the attached confirmation of receipt form. Forward the confirmation forms to our attention using the contact information following. For users: 1. If you have a product Urispec 11-Way, #102-2285, LOT 65402, please check if the leucocytes test pad is present before use of the test strips. 2. Pass the information on to other people and departments who may be working with the product Urispec 11-Way, #102-2285, LOT 65402. 3. Confirm the receipt of the information with the attached confirmation of receipt form by April 19, 2024. 4. If your product is affected, please send the product to your local distributor and arrange for replacement or contact MACHEREY-NAGEL directly. Please make sure in your organization that all users of the above products and other persons to be informed become aware of this Urgent Field Safety Notice. Please keep this information at least until all measures have been completed. Contact persons: Contact for sales / marketing: Dr. Sarah Geenen Product Management Rapid Tests Email: [email protected] Tel.: +49-(0)2421 969 340, Regulatory contact: Dr. Markus Meusel QAM, Manager Regulatory Affairs Email: [email protected] Tel. +49 (0) 2421 969 210.
Recall details
- Recall number
- Z-1834-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 4, 2024
- Posted by the FDA
- May 16, 2024
- Recalling firm
- Macherey Nagel Gmbh & Co. Kg
- Quantity
- 5,200 packages
- Product code
- JIL
- FDA submissions
- K991927
- Distribution
- US Nationwide distribution in the states of CO, FL, IN, NV, TX.
Product description
Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.
Code information (models, lots, serial numbers)
Product #102-2285, UDI/DI: 00304040001479, LOT 65402.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Henry Schein 1022285Henry Schein, Inc. · BLOOD, OCCULT, ENZYMATIC METHOD, IN URINE1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page