Life Scope VS recallZ-2025-2024 · Nihon Kohden America Inc · initiated Apr 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nihon Kohden America Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to software issue the device may give false "SpO2 Probe Failure" alarm

FDA root cause category: Software design

What the company is doing

On 4/29-30/2024, the firm sent "URGENT: MEDICAL DEVICE CORRECTION" Letters via FedEx to customer informing them that due to a software issue, the SpO2 probe failure message may occur when the value of the resistor built into the SpO2 relay code is incorrectly determined to be outside the normal range. This issue applies to all Life Scope BSM-3572A bedside monitors with software version 08-94. Customer are instructed to: 1. Please notify all individuals that need to be made aware of this issue/correction, and that this Notification should be posted with or near the affected products. Please notify any facilities if any affected product has been transferred to another facility. 2. Complete the attached form and return to Nihon Kohden. 3. If the SpO2 probe failure message is received and the probe appears undamaged, the bedside monitor may be providing an incorrect alarm. This issue may be temporarily resolved by turning off and restarting the BSM-3572A. 4. If adverse reactions or quality problems are experienced with the use of this product, please report to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. 5. If adverse reactions or quality problems are experienced with the use of this product, or if there are any questions or need for further assistance, please contact Nihon Kohden Technical Support at: 1-800-325-0283 or email mailto:[email protected].

Recall details

Recall number
Z-2025-2024
Status
Open, Classified
Initiated by the company
Apr 29, 2024
Posted by the FDA
Jun 6, 2024
Recalling firm
Nihon Kohden America Inc
Quantity
16 systems
Product code
MHX
Distribution
U.S. Nationwide distribution in the state of IA.

Product description

BSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility.

Code information (models, lots, serial numbers)

Model/Catalog Number: BSM-3572A UDI-DI code: 04931921113691 Serial Numbers: 1613, 1614, 1616, 1617, 1618, 1619, 1621, 1622, 1624, 1625, 1627, 1631, 1638, 1639, 1640, 1651

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.