Skip to content

Tri-Staple 2.0 recallZ-1891-2024 · Covidien, LP · initiated Apr 15, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery

FDA root cause category: Under Investigation by firm

What the company is doing

Medtronic issued Urgent Medical Device Recall letter on 4/15/24 via UPS 2-day. Letter states reason for recall, health risk and action to take: Customer Actions: " Immediately identify and quarantine all unused Tri-Staple" 2.0 Black Reinforced Intelligent Reload, Model SIGTRSB60AXT, from the specified lot(s) mentioned above. " Return all unused product in your inventory, corresponding to the specified lot(s) mentioned above, to Medtronic as indicated in the enclosed Customer Confirmation Form. Please complete the enclosed Customer Confirmation Form, even if you have no product to return. Pass on this information to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please keep a copy of this letter in your file. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.

Recall details

Recall number
Z-1891-2024
Status
Open, Classified
Initiated by the company
Apr 15, 2024
Posted by the FDA
May 24, 2024
Recalling firm
Covidien, LP
Quantity
1681 units
Product code
GDW
FDA submissions
K173270
Distribution
Worldwide - US Nationwide distribution in the state of NJ and the countries of Australia, Japan, New Zealand, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand.

Product description

Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis Model: SIGTRSB60AXT

Code information (models, lots, serial numbers)

GTIN: 10884521717633 20884521717630 Lot #: N3G1372Y

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K173270 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page