Ingenia recallZ-1830-2024 · Philips North America Llc · initiated Apr 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
FDA root cause category: Software design
What the company is doing
Philips issued URGENT Medical Device Correction letter on 4/22/24 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: a. Customers can continue to use the identified system(s) in accordance with the intended use pertaining to setting patient ventilation. (Reference IFU, including below information). {If your Ingenia 1.5T or Ingenia 3.0T system(s) has been upgraded to Ingenia Evolution configuration and running R5.7 software, it is impacted by this issue. (Note: customer will not see SW version detail of R5.7.1). o As stated in both IFU sections Adjust Ventilation in Bore and Adjusting the Bore Ventilation, The MR system has a recommended ventilation level. This is a static level based on average scans and system specific conditions. The system does not change the recommended ventilation level for e.g. high SAR scans, examination room temperature and patient weights, clothing and conditions. It is the responsibility of the operator to determine the level of ventilation for different scans and conditions together with the patient. Note: The recommended level is identified as level 3 in the IFU and in the images shown in Figure 1. o When setting ventilation at the VitalScreen, (IFU section Adjusting the Bore Ventilation) and ventilation is set to a value marked with a caution sign, you are warned that the patient might get too warm and that you need to pay attention. o When setting ventilation at the Operator Console (IFU Section Adjusting Ventilation in Bore): Current patient ventilation level, Level 3 is recommended. Use the <=> and <-> buttons above to modify the patient ventilation level. o Note: Should Patient Ventilation level be below level 3, operator is to pay additional attention to patient temperature. b. Refer to Appendix A for specific details regarding the software issue descriptions and advice to customers. c. Please co
Recall details
- Recall number
- Z-1830-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 22, 2024
- Posted by the FDA
- May 15, 2024
- Recalling firm
- Philips North America Llc
- Quantity
- 19 units
- Product code
- LNH
- FDA submissions
- K193215
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.
Product description
Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117
Code information (models, lots, serial numbers)
UDI-DI: (01)00884838099739(21) Serial Numbers: 42029 42217 42246 71044 71774 85146 85645 85663 85664 85668 85677 85680 85681 85685 85700 85704 85705 85706 85713
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Evolution Upgrade 3.0TPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging6 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94493).
- Z-1829-2024 · Open, ClassifiedIngenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page