Skip to content

ClearCut recallZ-1923-2024 · Alcon Research LLC · initiated Apr 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Alcon Research LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

FDA root cause category: Process control

What the company is doing

On April 22, 2024 VOLUNTARY MEDICAL DEVICE FIELD CORRECTION letters to customers. Actions to be taken by the Customer / User: If a knife does not perform as expected, or if you feel it does not meet your expectation for sharpness, discontinue use of that specific knife and replace the knife with a new one to avoid potential complications. To acknowledge your receipt of this Voluntary Medical Device Correction notification, please take the following steps: 1. Forward this notification to all departments or organizations using Alcon Ophthalmic Knives. 2. Follow the risk mitigation precautions provided in this notice when using identified catalogue numbers of ophthalmic knives. 3. Please complete the attached Response Form indicating your understanding of the included instructions and return the attached Response Form via email or fax to Alcon. Email: [email protected] Fax: 817-302-4337 Alcon has reported this issue to Health Authorities in accordance with applicable regulations. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email ([email protected]) or by phone (1-800-757-9780). Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program either online (FDA EMAIL), by regular mail or by fax: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm. Should you have any questions or concerns about this matter or need help finding a replacement or substitute sterile standalone ophthalmic surgical knife, please call Alcon Customer Service or contact your Alcon Sales Representative.

Recall details

Recall number
Z-1923-2024
Status
Open, Classified
Initiated by the company
Apr 22, 2024
Posted by the FDA
May 30, 2024
Recalling firm
Alcon Research LLC
Quantity
427,764 units
Product code
HNN
FDA submissions
K110166
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.

Product description

ClearCut Sideport Knife Dual Bevel 1.2mm Angled Model/Catalog Number: 8065921541 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A

Code information (models, lots, serial numbers)

Model/Catalog Number: 8065921541; UDI/DI: 00380659215414; Lot/Batch: 15W7XT, 15W7Y8, 15W7Y9, 15W9E6, 15W9E8, 15X5JJ, 15X56A, 15X56C, 15X5KV, 15X5KX, 1618UH, 1618UJ, 161A1K, 1618V4, 1618V5, 161A21, 161A22, 162WY5, 162WY6, 162WYL, 162WYN, 1636Y8, 1636Y9, 1636YV, 164RXL, 164RY8, 164RYA, 164WV3, 164WW9, 164WWC, 1668MR, 1668MT, 1668NH, 1668NJ, 1668NK, 166E89, 166E90, 166E91, 167PVK, 167R34, 167PW7, 167PW8, 167PW9, 167R3K, 167R3L, 168TE8, 168UEX, 168TEN, 168UFJ, 168UFL, 169L38, 169L3M, 169P3C, 169P3D, 16ALXE, 16ALXF, 16AM34, 16ALY0, 16AM36, 16AV03, 16AV1T, 16DFKA, 16DFXT, 16DFL0, 16DFY7, 16DFY8, 16DNJ4, 16DNJN, 16DNJP, 16DNP3, 16DNP4, 16EM6V, 16EM6W, 16JAVH, 16JAVJ, 16JAVK, 16JAVL, 16JAVM;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94501).

Sources and updates