Skip to content

Cascade IOMAX Cortical Module, REF: 190296-200 recallZ-1882-2024 · Cadwell Industries Inc · initiated Apr 23, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cadwell Industries Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.

FDA root cause category: Process design

What the company is doing

On 4/22/24, Cadwell Industries, Inc. issued a "Urgent: Medical Device Recall" notification to affected consignees via Email: 1. Cease use of the device immediately and return all identified devices to Cadwell Industries, Inc. 2. Please complete the attached response form and return via email to [email protected]. 3. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred the IOMAX Cortical Module to. 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Recall details

Recall number
Z-1882-2024
Status
Open, Classified
Initiated by the company
Apr 23, 2024
Posted by the FDA
May 22, 2024
Recalling firm
Cadwell Industries Inc
Quantity
54 units
Product code
GWF
FDA submissions
K162199
Distribution
Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.

Product description

Cascade IOMAX Cortical Module, REF: 190296-200

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI: 00840067100653/ Serial numbers: 19029603AA1223002 19029603AA1223003 19029603AA1223004 19029603AA1223005 19029603AA1223006 19029603AA1223007 19029603AA1223011 19029603AA1223012 19029603AA1223013 19029603AA1223014 19029603AA1223015 19029603AA1223016 19029603AA1223017 19029603AA1223018 19029603AA1223019 19029603AA1223020 19029603AA1223024 19029603AB1223021 19029603AB1223023 19029603AA1223025 19029603AA0124001 19029603AA0124002 19029603AA0224008 19029603AA0124007 19029603AA0124012 19029603AA0124017 19029603AA0124004 19029603AA0124010 19029603AA0124016 19029603AA0224010 19029603AA0124015 19029603AA1223026 19029603AA0124008 19029603AA0124009 19029603AA1223027 19029603AA1223028 19029603AA1223029 19029603AA0124003 19029603AA0224001 19029603AA1223030 19029603AA1223008 19029603AA1223009 19029603AA0124013 19029603AA1223031 19029603AA1223001 19029603AA0224002 19029603AA0224003 19029603AA0224004 19029603AA0224005 19029603AA0224006 19029603AA0224007 19029603AA0124014 19029603AA0224009 19029603AA0124011

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page