Ingenia recallZ-1852-2024 · Philips North America Llc · initiated May 3, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
FDA root cause category: Under Investigation by firm
What the company is doing
Philips issued Urgent Medical Device Correction notifications to customers on 22-Apr-2024 via FedEx . Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: If a smoke/fire alarm is detected: a. Immediately stop scanning and evacuate the patient and staff from the examination room. b. If a developing fire is detected, adhere to established hospital fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button. c. Do not attempt to continue scanning. d. Immediately contact Philips Service. " Ensure all users are aware of facility specific Emergency Procedures as outlined in Chapter 2: Safety in the Instructions for Use Emergency procedures The User is required to establish emergency procedures for the following situations: " A medical emergency " A fire " An emergency that requires immediate removal of the magnetic field " The release of helium gas into the examination room Philips MRI systems have an Emergency Table Stop button in case there is an emergency during tabletop movement. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. " Post this notice near the affected MR system(s) for ease of reference. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: [email protected]. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 5. The actions planned by Philips to correct the problem Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the g-MDU connections in the technical room and apply the proper torque to the connection if necessary.(reference FCO78100582). Philips plans t
Recall details
- Recall number
- Z-1852-2024
- Status
- Open, Classified
- Initiated by the company
- May 3, 2024
- Posted by the FDA
- May 17, 2024
- Recalling firm
- Philips North America Llc
- Quantity
- 1 unit
- Product code
- LNH
- FDA submissions
- K193215
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Product description
Ingenia Elition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781358
Code information (models, lots, serial numbers)
UDI-DI: (01)00884838088115(21) Serial Number: 45030
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94508).
- Z-1845-2024 · Open, ClassifiedIngenia 1.5T Model Number (REF): (1) 782140; (2) 782115; (3) 782101; (4) 781396; (5) 781341; (6) 781315
- Z-1846-2024 · Open, ClassifiedIngenia 1.5T CX - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 781262 (2) 781261
- Z-1847-2024 · Open, ClassifiedIngenia 1.5T S Model Number (REF): 781347
- Z-1848-2024 · Open, ClassifiedIngenia 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782103 (2) 781377 (3) 781342
- Z-1849-2024 · Open, ClassifiedIngenia 3.0T CX Model Number (REF): (1) 782105 (2) 781271
- Z-1850-2024 · Open, ClassifiedIngenia Ambition S Model Number (REF): (1) 782139; (2) 782133; (3) 782108; (4) 781359
- Z-1851-2024 · Open, ClassifiedIngenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
- Z-1854-2024 · Open, ClassifiedMR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782152 (2) 782110
- Z-1855-2024 · Open, ClassifiedPanorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
- Z-1862-2024 · Open, ClassifiedSmartPath to dStream for XR and 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782129; (…
- Z-1863-2024 · Open, ClassifiedSmartPath to Ingenia Elition X - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782118
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page