Skip to content

Medline Luer Slip Disposable Syringe Without Needle, Ref… recallZ-1938-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Apr 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries, LP, issued a recall notice to its consignees on 04/22/2024 via USPS first class mail. The notice explained the problem with the device, potential risk, and requested that the affected product be destroyed.

Recall details

Recall number
Z-1938-2024
Status
Open, Classified
Initiated by the company
Apr 22, 2024
Posted by the FDA
May 31, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
5,075,234 syringes
Product code
FMF
FDA submissions
K061275
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.

Product description

MEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020

Code information (models, lots, serial numbers)

UDI/DI 10080196306183, All lots in the format: 897YYMMXXXX

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 94537).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page