MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES… recallZ-2040-2024 · Ge Medical Systems Israel, Functional Imaging · initiated Apr 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ge Medical Systems Israel, Functional Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.
FDA root cause category: Device Design
What the company is doing
The recalling firm issued letters on and dated 4/15/2024 via FedEx informing the consignee of the safety issue and actions to be taken by the customer/user. The actions state the consignee can continue to use the system but until GE HealthCare implements the correction for this potential issue on the affected device and before conducting any servicing activities that involve movement of the covers, GE HealthCare Service should be contacted for guidance. An acknowledgement form was enclosed for completion and return. The letter contained an embedded photograph of the device showing where to find the manufacturing date to determine if it is an affected device. The letter also informs the consignee they will be contacted by a GE HealthCare representative to arrange for the correction.
Recall details
- Recall number
- Z-2040-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2024
- Posted by the FDA
- Jun 7, 2024
- Recalling firm
- Ge Medical Systems Israel, Functional Imaging
- Quantity
- 133 devices
- Product code
- KPS
- FDA submissions
- K212004
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.
Product description
MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
MyoSPECT manufactured on or before 2024-02-13: Model H3912AA, GTIN 00195278421586 - Serial numbers NGCC80094, NGCB80033, NGCB80044, NGCB80045, NGCA80017, NGCB80039, NGCA80006, NGCA80011, NGCA80003, NGCA80012, NGCB80052, NGCB80060, NGCC80074, NGCC80081, NGCB80038, NGCD80107, NGCB80070, NGCC80093, NGCD80106, NGCB80026, NGCB80050, NGCB80058, NGCC80084, NGCB80027, NGCB80072, NGCB80032, NGCB80031, NGCC80080, NGCB80067, NGCA80021, NGCA80010, NGCA80015, NGCB80043, NGCC80090, NGCC80091, NGCD80104, NGCB80028, NGCB80022, NGCB80023, NGCB80048, NGCB80035, NGCD80103, NGCC80100, NGCC80083, NGCB80042, NGCC80075, NGCB80068, NGCB80040, NGCB80051, NGCB80064, NGCC80078, NGCC80079, NGCC80087, NGCC80092, NGCD80105, NGCD80108, NGCB80059, NGCA80009, NGCC80086, NGCC80095, NGCA80020, NGCC80088, NGCB80055, NGCB80029, NGCB80041, NGCC80076, NGCA80016, NGCB80036, NGCC80098, NGCB80037, NGCB80046, NGCA80007, NGCC80082, NGCB80061, NGCA80002, NGCC80102, NGCB80057, NGCC80089, NGCB80030, NGCC80097, NGCA80005, NGCB80049, NGCC80101, NGCB80071, NGCA80001, NGCB80066, NGCB80065, NGCC80085, NGCC80077, NGCB80069, NGCA80014, MYGA10001, NGCC80096, NGCB80024, NGCB80047, NGCC80099, NGCB80062, NGCB80034, NGCA80008, NGCA80013, NGCB80025, NGCA80004, NGCA80019, NGCB80056, NGCB80063, NGCB80054, and NGCB80053. Model H3912BC, GTIN 00195278488619 - Serial number MYOC93001. MyoSPECT ES manufactured on or before 2024-02-15: Model H3912AB, GTIN 00195278421579 - Serial numbers NGEB81002, NGED81025, NGEB81007, NGEB81006, NGEC81019, NGEB81012, NGEB81008, NGEC81014, NGEB81005, NGEC81017, NGEC81022, NGEC81021, NGEC81020, NGEB81011, NGEB81010, NGEC81023, NGEC81024, NGEC81013, NGEB81009, NGEB81003, NGEB81004, NGEC81018, NGEC81015, and NGEC81016. Model H3912BD, GTIN 00195278488626 - Serial number MYEB94001.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- MyoSPECT ES H3912ABG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- MyoSPECT H3912AAG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- MyoSPECT H3912BCG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- MyoSPECT ES H3912BDG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page