2nd Generation CentriMag Primary Console, REF: 201-90401… recallZ-2215-2024 · Thoratec Switzerland GMBH · initiated May 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Thoratec Switzerland GMBH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
FDA root cause category: Process control
What the company is doing
On 5/22/24, correction notice distribution started. Customers were asked to do the following: If the console shuts down without alarms or alerts present, follow Instruction for Use (IFU) for Console and Motor malfunction as described in the Section Emergency / Troubleshooting , and switch to a backup console, which must be prepared as required backup components in the immediate vicinity. As stated in the IFU these backup systems should have a battery charge sufficient for at least one hour of operation, be connected to wall power, and be powered off. Note: When the backup console is put into use, the previously used console can be restarted and set up as a backup console. Per the IFU and current practice, a full console backup system is always required to be available in the immediate vicinity of each patient whenever the system is used. Firm is in the process of implementing an additional in-process test by Q3 of 2024, which will ensure that all systems which will be distributed are confirmed to withstand a surge up to 2.0 kV. Distribute this notice to those who need to be aware within your institution and to any organization where potentially affected devices may have been transferred in the event that devices were transferred elsewhere. Complete and return the acknowledgement form via email to [email protected] Should you have any questions about this communication, please contact firm Customer Service at 1 (800) 456-1477
Recall details
- Recall number
- Z-2215-2024
- Status
- Open, Classified
- Initiated by the company
- May 22, 2024
- Posted by the FDA
- Jun 27, 2024
- Recalling firm
- Thoratec Switzerland GMBH
- Quantity
- 3,425
- Product code
- DSQ
- FDA submissions
- P170038
- Distribution
- Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRAN, IRAQ, ISRAEL, ITALY, KENYA, KOREA, KUWAIT, LITHUANIA, MACEDONIA MALAYSIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NORWAY, OMAN, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
Product description
2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System
Code information (models, lots, serial numbers)
REF/Serial Number or Range: 201-30300/ L00610-0004-L06786-0002 201-90401/ 19234129-20119751 and L00610-0011-L08005-0001 201-90411/19234128-20190463 and L03849-0004-L08002-0001 201-90421/19474336-19606873 and L04780-0005-L07601-0001 201-90701/ 19415572 and L03774-0007-L08007-0001 L201-90401/ L00694-0006-L06579-0004 and L201-90401 L201-90411/ 19258710-19605455 and L01627-0007-L07980-0001 L201-90421/ L01716-0010-L06482-0005 UDI-DI: 07640135140689, 07640135140726, 07640135140702, 07640135140719, 07640135140856, 05415067037282, 07640135140894, 07640135149019, 07640135149002, 07640135140702, 07640135140719
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 2nd Gen CentriMag Primary Console 201-90411Thoratec Switzerland GmbH · Control, Pump Speed, Cardiopulmonary Bypass1 recall459 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page